Peter Huber Kältemaschinenbau Unimotive GL 60w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unimotive GL 60w

Peter Huber Kältemaschinenbau Unimotive GL 60w — Temperature control units.

The Unimotive GL 60w is a water-cooled temperature control unit with a magnetically coupled stainless-steel pump and 48 kW heating capacity, covering -45 to 95 °C for water-glycol heat transfer fluids. It is the largest modification in the Unimotive w-series, distinguished by its 19.9 l expansion tank and 8.9 l minimum fill. It serves automotive temperature simulation and material testing where precise, high-capacity heating and cooling are required.

Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection

Model Display resolutionHeating capacity, WPower supplyPump connectionRefrigerantSafety class
Unimotive GL 60w Current 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive 10w 0.0112000400V 3~ 50HzM38x1,5 AG maleR-452A (A1, H280)III / FL
Unimotive 10w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 20w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 20w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 26w 0, 01 K24000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 26w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 27w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive GL 07w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 07w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 10w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 10w-XT 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 20w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 20w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 30w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 60w-XT 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 48 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable no
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -45...95 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-744 (A1, H280) (1st stage)
Maximum viscosity 200 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The GL 60w is a water-cooled (w) temperature control unit in the Unimotive series, using R-744 (CO2) refrigerant in the first stage and a magnetically coupled stainless-steel pump with M38x1.5 male connections. Its 19.9 l expansion tank and 8.9 l minimum filling capacity set it apart from smaller siblings such as the Unimotive 10w and 20w, which have lower fluid volumes and heating capacities. The unit is designed for water-ethylene glycol mixtures with corrosion protection down to -45 °C, making it suitable for automotive temperature simulation rather than general laboratory use.

Behavior under load and media

Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure; higher cooling water temperatures reduce cooling power and may increase flow rate. The pump is not speed-adjustable, so flow pressure must be considered for pressure-sensitive applications such as glass reactors. Heating capacity is 48 kW, and the temperature range spans -45 to 95 °C with 0.01 K display resolution. The cooling water circuit materials (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM) require compatible site water; sea water is prohibited.

Installation, control and operation

The unit requires a 400 V 3~ 50 Hz supply with ±5% voltage and ±2% frequency tolerance (opposite-sign deviations not allowed), and three-phase devices below 63 A are supplied with cable but without plug. Cooling water connections are G3/4 male, with differential pressure between 1 and 4 bar and maximum pressure 6 bar. The Pilot ONE controller with POKO ONE control output provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 external sensor connection. IP20 protection and 5–40 °C ambient limit installation to dry indoor environments; a drip tray and backflow preventer are required on site.

Integration & Operating Notes

Mounting
Install only in dry indoor areas within 5–40 °C ambient

The IP20 rating and 5–40 °C ambient limit exclude outdoor or wet locations. Minimum wall clearances depend on air- or water-cooling and on whether the unit runs with or without a bath, so the installation layout must reserve the specified clearance to avoid overheating.

Datasheet, Specifications; Operation Manual, Section on Setup Mode
Electrical
Provide a 400 V 3~ 50 Hz supply with PE and a site plug

Three-phase devices below 63 A are supplied with cable but without plug, so the site must provide a suitable power connection. Voltage and frequency tolerances are ±5% and ±2%, but opposite-sign deviations such as -5% voltage with +2% frequency are not allowed. Hardwiring must be done by a certified electrician.

Datasheet, Specifications; Operation Manual, Section on Electrical Connection
Process Limits
Cooling water must be 5–20 °C with 1–4 bar differential

The unit requires cooling water inlet between 5 °C and 20 °C, differential pressure between 1 and 4 bar, and maximum pressure 6 bar. Higher cooling water temperature reduces cooling power and may increase flow rate, so capacity margins must account for warmer site water. Only drinking water or filtered river water may be used; sea water is prohibited.

Datasheet, Specifications; Operation Manual, Section on Cooling Water Connection
Maintenance
Shut off cooling water during downtimes and protect from frost

Cooling water supply should be shut off even for shorter downtimes (e.g. overnight) because flow can occur when the unit is switched off. Outdoor installation requires frost-protected cooling water lines and a minimum cooling water temperature of 5 °C. A backflow preventer on the building side is mandatory to protect the drinking water supply.

Operation Manual, Section on Cooling Water Connection
Control & I/O
Pilot ONE controller offers Ethernet, USB and RS232 interfaces

The Pilot ONE controller with POKO ONE control output provides Ethernet, USB (host and device) and RS232 for integration into automation systems. An external Pt100 sensor connection is available for monitoring the application temperature. The pump speed is not adjustable, so flow pressure must be considered for pressure-sensitive applications.

Datasheet, Specifications

Selection & Integration Q&A

When should I choose the Unimotive GL 60w over the smaller Unimotive 10w or 20w?

Choose the GL 60w when you need 48 kW heating capacity, a 19.9 l expansion tank and 8.9 l minimum fill for larger fluid volumes or higher heat loads. The 10w and 20w have lower capacities and smaller tanks, so they suit smaller systems with lower heat demand.

What are the requirements for the cooling water supply and adjacent piping?

Provide cooling water at 5–20 °C with differential pressure between 1 and 4 bar and maximum pressure 6 bar. Connections are G3/4 male. Use only drinking water or filtered river water; sea water is prohibited. Install a backflow preventer on the building side.

What mounting and environmental limits apply from the manual?

Install indoors only, within 5–40 °C ambient, with IP20 protection. Do not install in an ATEX zone. Provide a drip tray below the unit and respect minimum wall clearances. For outdoor siting, frost-protect cooling water lines and keep water above 5 °C.

What maintenance and accessories are required that are not included in base delivery?

The base unit is supplied without a power plug for three-phase connections below 63 A, so a site plug is needed. A drip tray and backflow preventer are site responsibilities. Shut off cooling water during downtimes. Heat transfer fluid (water-glycol with corrosion protection) must be supplied separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unimotive GL 60w
Datasheet ·31,5 КБ ·2 pp.
View PDF
PDF
Unimotive Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF