Peter Huber Kältemaschinenbau Unistat T345 Dynamisches

Peter Huber Kältemaschinenbau Unistat T345 Dynamisches — Temperature control units.

The Unistat T345 Dynamisches is a dynamically controlled, magnetically coupled stainless-steel pump temperature control unit with 72 kW heating and a 65–300 °C working range, positioned above the T305 siblings in heating capacity. It serves high-temperature process loops where the 56 l unregulated expansion tank and 200 l/min pump flow support closed or open external circuits.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat T345 Dynamisches Current 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305 HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305w HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T320w HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330 HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 Dynamisches 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T340 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T350 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T402 Dynamisches 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Unistat TR401 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005
Unistat TR402 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Expansion volume 56 l
Heating capacity 72 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range 65...300 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior and process limits

Installation and operation

Integration & Operating Notes

Process Limits
No active refrigeration: 65 °C minimum is application-dependent

The unit cannot reach its 65 °C lower limit on its own; the minimum depends on heat loss in the application or on suitable water cooling. Specifying the T345 for cooling duties below ambient without an external heat sink or water cooling will not meet the process requirement.

Unistat T345 Technical Datasheet, page 2
Electrical
Leakage current above 3.5 mA requires RCD coordination

Protective earthing and residual-current device selection must account for a leakage current exceeding 3.5 mA; type-B or higher RCD coordination may be required. The supply is 400 V 3~ 50 Hz, and voltage/frequency tolerances must be respected simultaneously, with combinations such as -5% voltage and +2% frequency not allowed.

Unistat T345 Technical Datasheet, page 2
Electrical
Three-phase units under 63 A arrive with cable but no plug

The site must provide a suitable plug and connection for the supplied cable. Hard-wired power connection must be carried out by an electrician, and supply voltage and frequency must match the rating plate before commissioning.

Unistat T345 Technical Datasheet, page 2; Unistat Operation Manual, page 37
Mounting
Expansion vessel lock state depends on open vs closed application

For externally closed applications the expansion vessel must remain unlocked; locking it will damage the application due to overpressure. For externally open applications the expansion vessel must be locked to prevent fluid loss. The 56 l tank is not thermoregulated, which affects fluid volume and expansion allowance in closed systems.

Unistat Operation Manual, page 33; Unistat T345 Technical Datasheet, page 1
Mounting
Leveling feet and parking brakes must be set before commissioning

Casters are not suitable for transport and carry only 25% of total mass symmetrically; use an industrial truck for relocation. Before commissioning, activate parking brakes and unscrew or activate leveling feet to ensure stable, level installation and prevent creeping during operation.

Unistat Operation Manual, page 22
Process Limits
Thermofluid viscosity must stay below 50 mm²/s

The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s, which limits fluid choice at low operating temperatures. Select a fluid whose viscosity at the lowest expected temperature remains within this limit to avoid pump overload.

Unistat T345 Technical Datasheet, page 1
Control & I/O
Process Safety mode requires compressor set to Permanently ON

When Process Safety mode is used for exothermic process protection, the compressor must be configured to Permanently ON so that cooling remains active at the overtemperature trigger. Power Failure Auto resumes temperature control automatically after a power outage if it was active, which affects unattended operation risk assessment.

Unistat Operation Manual, pages 19–20
Maintenance
Shut off cooling water even for overnight downtimes

The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.

Unistat Operation Manual, page 29

Selection & Integration Q&A

When should I choose the Unistat T345 Dynamisches rather than a T305 sibling?

Pick the T345 when the process needs the higher 72 kW heating capacity and a 65–300 °C range; the T305, T305 HT and T305w HT differ in heating capacity and temperature-range configuration. The T345 has no active refrigeration, so sub-ambient duties require a chiller or water cooling.

What must the site provide for the power connection and protective coordination?

A 400 V 3~ 50 Hz supply with voltage and frequency tolerances respected simultaneously; three-phase units under 63 A come with cable but no plug, so the site supplies the plug. Leakage current exceeds 3.5 mA, so earthing and RCD selection must be coordinated, and connection must be done by an electrician.

What mounting and environment limits apply from the manual?

Install indoors and dry (IP20) with ambient 5–40 °C; transport and store upright only and move only when emptied. Allow about 2 hours acclimation when changing between cold and warm environments before power-on, and set leveling feet and parking brakes before commissioning.

What maintenance and accessories are needed beyond the base delivery?

The base delivery excludes the power plug for three-phase units under 63 A and the mating M38x1.5 hoses/fittings. Periodically check hoses and connections for leaks and quality and replace as needed; shut off cooling water even overnight. Set the adjustable DIN 12876 overtemperature protection during commissioning.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat T345 Technical Datasheet
Datasheet ·17,8 КБ ·2 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF