Peter Huber Kältemaschinenbau CC-106A
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-106A

Peter Huber Kältemaschinenbau CC-106A — Thermostatic baths.

The CC-106A is a 4.4-litre immersion circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated for 25–100 °C (20–100 °C with water cooling, 15–100 °C with refrigerator). It differs from the CC-104A/108A/110A/112A siblings by its 4.4 l bath volume and 1.5–2.1 kW heating capacity, serving as a temperature-control source for external circuits or open baths in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
CC-106A Current Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-104A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-108A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-110A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-112A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
CC-118A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
KISS 104A KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
KISS 106A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
KISS 108A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 110A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 112A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
KISS 118A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Dimensions (w × D × H) 147 x 307 x 330 mm
Maximum delivery flow rate 27 l/min (pressure pump max.; 0, 7 bar)
suction pump max.; 0, 4 bar: 22 l/min
Pump connection M16x1 AG male (Option, #19606)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 1, 5 - 2, 1 kW
Bath opening WxD/BD mfr 130 x 110 / 150 mm
Bath volume mfr 4, 4 l
Filling volume min. mfr 2, 5 l (min.)
Dimensions, Weight & Mechanical Design
Weight 5 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal and hydraulic behaviour

Control, connection and operation

Integration & Operating Notes

Electrical
Provide a 10–16 A branch circuit with UL 489 protection

The unit draws up to 10 A on a 200–240 V 1~/2~ 50/60 Hz supply and requires a minimum 10 A fuse (maximum 16 A). External branch circuit protection according to UL 489 must be installed; voltage and frequency tolerances are interdependent, so a −10% voltage combined with +3% frequency is not permitted.

CC-106A Immersion Circulator Datasheet, page 1–2
Process Limits
Match cooling method to the required low temperature

The achievable lower limit depends on the cooling configuration: 25 °C without cooling, 20 °C with water cooling, and 15 °C with a refrigerator. Specify the cooling option at ordering if the process requires temperatures below 25 °C.

CC-106A Immersion Circulator Datasheet, page 1
Process Limits
Do not use polyglycol with the Huber pump adapter

When the optional Huber pump adapter (M16x1 AG male, #19606) is fitted, polyglycol is not permissible as heat transfer fluid. Select an alternative fluid compatible with the pump adapter and the wetted materials before commissioning.

CC-106A Immersion Circulator Datasheet, page 2
Process Limits
Charge at least 2.5 l of heat transfer fluid

The bath must be filled with a minimum of 2.5 l (4.4 l total volume) to ensure correct pump and heating behaviour. Operating with insufficient fluid risks pump cavitation and heater over-temperature.

CC-106A Immersion Circulator Datasheet, page 1
Mounting
Keep ambient temperature between 5 °C and 40 °C

The circulator is rated for an ambient temperature of 5–40 °C. Installation outside this window affects temperature stability and may trip the overtemperature protection; the IP20 enclosure also requires a dry, dust-controlled location.

CC-106A Immersion Circulator Datasheet, page 1
Electrical
Specify the country-specific mains plug at ordering

Single- and two-phase devices (100–240 V) are delivered with a power cable and a country-specific plug that must be stated on the order. Missing specification delays commissioning because the plug cannot be assumed.

CC-106A Immersion Circulator Datasheet, page 2
Control & I/O
Set the adjustable overtemperature cut-out for the application

The unit provides adjustable overtemperature protection per DIN 12876 (safety class III / FL). The cut-out threshold must be configured during commissioning to match the heat transfer fluid and process, otherwise the safety function does not protect the application.

CC-106A Immersion Circulator Datasheet, page 1
Maintenance
Register the unit to activate the 4-year warranty

The 4-year warranty requires registration. Complete the registration at commissioning so that warranty coverage is secured before any service claim.

CC-106A Immersion Circulator Datasheet, page 1

Selection & Integration Q&A

When should I choose the CC-106A rather than the CC-104A or CC-108A sibling?

Pick the CC-106A when the process needs a 4.4 l bath with 1.5–2.1 kW heating and ±0.02 K stability. The CC-104A is smaller; the CC-108A/110A/112A step up in bath size. Match the bath volume to the external loop and immersion depth (150 mm) before selecting.

What supply and protection must be provided for the CC-106A, and what plug comes with it?

Provide 200–240 V 1~/2~ 50/60 Hz with a 10 A minimum fuse (16 A maximum) and external UL 489 branch protection. The unit ships with a country-specific plug that must be specified at ordering; voltage/frequency tolerance combinations such as −10% V with +3% Hz are not allowed.

What maintenance and accessories are needed beyond the base delivery?

The base unit requires registration to activate the 4-year warranty. The M16x1 AG male pump connection (#19606) is an option, and polyglycol is not permitted with the Huber pump adapter. External Pt100 sensors and the cooling source (water or refrigerator) are not part of the base delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-106A Immersion Circulator Datasheet
Datasheet ·128,0 КБ ·2 pp.
View PDF