Peter Huber Kältemaschinenbau Unimotive GL 07w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unimotive GL 07w

Peter Huber Kältemaschinenbau Unimotive GL 07w — Temperature control units.

The Unimotive GL 07w is a water-cooled temperature control unit with a 12 kW heating capacity and a -45 to 95 °C working range, using R-744 refrigerant in the first stage. Its magnetically coupled stainless steel pump and G3/4 cooling-water connections distinguish it within the Unimotive series, where sibling models 10w/20w and XT variants cover other capacities. It is intended for automotive temperature simulation and material testing with water-ethylene glycol mixtures.

Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection

Model Display resolutionHeating capacity, WPower supplyPump connectionRefrigerantSafety class
Unimotive GL 07w Current 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive 10w 0.0112000400V 3~ 50HzM38x1,5 AG maleR-452A (A1, H280)III / FL
Unimotive 10w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 20w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 20w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 26w 0, 01 K24000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 26w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 27w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive GL 07w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 10w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 10w-XT 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 20w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 20w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 30w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 60w 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 60w-XT 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 12 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable no
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -45...95 °C
Dimensions, Weight & Mechanical Design
Weight 511 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-744 (A1, H280) (1st stage)
Maximum viscosity 200 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control, and operation

Integration & Operating Notes

Mounting
Install drip tray and respect space requirements

The responsible body must install a condensation water / heat transfer fluid drip tray below the unit. Overall dimensions are 730x804x1738 mm and net weight is 511 kg, so adequate floor space and structural support are required. Minimum wall clearances depend on air- or water-cooling and on whether the unit runs with or without a bath.

Operation Manual, Section on Installation; Technical Data Sheet, page 2
Electrical
Provide 400 V 3~ 50 Hz supply with PE and plug

The unit requires a 400 V 3~ 50 Hz power supply with protective earth. Three-phase devices under 63 A are supplied with cable but without plug, so a suitable plug must be provided on site. Leakage current exceeds 3.5 mA, requiring appropriate protective earthing and RCD selection. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Technical Data Sheet, page 2; Operation Manual, Section on Electrical Connection
Process Limits
Cooling water must be 5–20 °C and 1–6 bar

Cooling water inlet temperature must be between 5 °C and 20 °C, with a differential pressure of at least 1 bar and a maximum pressure of 6 bar. Only drinking water or filtered river water may be used; sea water is prohibited. A backflow preventer on the building side is required. If cooling water temperature falls below the condensation point, internal pipes must be insulated.

Technical Data Sheet, page 2; Operation Manual, Section on Cooling Water Connection
Control & I/O
Pilot ONE controller with Ethernet, USB, RS232

The Pilot ONE controller provides a display resolution of 0.01 K and interfaces including Ethernet, USB (Host and Device), and RS232. An external Pt100 sensor can be connected. The optional Flow Control Cube enables precise flow rate measurement and control, which may be needed for accurate testing.

Technical Data Sheet, page 1; Operation Manual, Section on Interfaces
Maintenance
Shut off cooling water during downtimes

Cooling water supply should be shut off even for shorter downtimes (e.g., overnight) because flow can occur even when the unit is switched off. Before switching off, the heat transfer fluid must be brought to room temperature to avoid overpressure in the system.

Operation Manual, Section on Normal Operation

Selection & Integration Q&A

When should I choose the Unimotive GL 07w over the 10w or 20w models?

Choose the GL 07w when a 12 kW heating capacity and -45 to 95 °C range with water cooling suit your process. The 10w and 20w offer different capacities; the XT variants may provide extended features. Select based on required heating/cooling power and temperature range.

What are the requirements for the cooling water supply and adjacent piping?

Cooling water must be 5–20 °C, 1–6 bar differential pressure, max 6 bar, with G3/4 male connections. Use only drinking or filtered river water; sea water is prohibited. Install a backflow preventer and insulate pipes if below condensation point.

What maintenance and accessories are required that are not included in base delivery?

A suitable plug for the 400 V 3~ power cable must be provided on site. A backflow preventer and drip tray are required. The optional Flow Control Cube is not included. Regular maintenance includes shutting off cooling water during downtimes and bringing fluid to room temperature before shutdown.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unimotive GL 07w — Technical Data Sheet
Datasheet ·96,7 КБ ·2 pp.
View PDF
PDF
Unimotive Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF