Peter Huber Kältemaschinenbau Unimotive GL 07w
Peter Huber Kältemaschinenbau Unimotive GL 07w — Temperature control units.
The Unimotive GL 07w is a water-cooled temperature control unit with a 12 kW heating capacity and a -45 to 95 °C working range, using R-744 refrigerant in the first stage. Its magnetically coupled stainless steel pump and G3/4 cooling-water connections distinguish it within the Unimotive series, where sibling models 10w/20w and XT variants cover other capacities. It is intended for automotive temperature simulation and material testing with water-ethylene glycol mixtures.
Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection
| Model | Display resolution | Heating capacity, W | Power supply | Pump connection | Refrigerant | Safety class |
|---|---|---|---|---|---|---|
| Unimotive GL 07w Current | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive 10w | 0.01 | 12000 | 400V 3~ 50Hz | M38x1,5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 10w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 20w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 20w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 26w | 0, 01 K | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 26w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 27w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive GL 07w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 10w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 10w-XT | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 20w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 20w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 30w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 60w | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 60w-XT | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 12 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | no |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -45...95 °C |
| Weight | 511 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-744 (A1, H280) (1st stage) |
| Maximum viscosity | 200 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The GL 07w is a water-cooled unit with a 12 kW heating capacity and a temperature range of -45 to 95 °C, using R-744 (A1, H280) refrigerant in the first stage.
- It is equipped with a magnetically coupled stainless steel pump with M38x1.5 AG male connections and a non-adjustable pump speed.
- The cooling-water circuit uses G3/4 male connections and requires a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.
- Sibling models in the Unimotive series include the 10w, 10w-XT, 20w, and 20w-XT, which differ in capacity and features; the GL 07w is the base water-cooled variant with 12 kW heating.
Behavior under load and media
- The unit is designed for operation with water-ethylene glycol mixtures with corrosion protection (e.g., Glysantin®) down to -45 °C, making it suitable for automotive temperature simulation and material testing.
- Cooling power decreases as cooling water temperature rises, and the cooling water flow rate may need to be increased accordingly.
- The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only compatible cooling water may be used.
- The unit has adjustable overtemperature protection according to DIN 12876, which must be set for safe operation.
Installation, control, and operation
- The Pilot ONE controller with POKO ONE control output provides display resolution of 0.01 K and interfaces including Ethernet, USB (Host and Device), and RS232.
- External sensor connection is via Pt100.
- The unit requires a 400 V 3~ 50 Hz power supply and has a protection rating of IP20.
- Cooling water must be supplied at 5–20 °C inlet temperature; if the cooling water temperature falls below the condensation point, the cooling water pipes inside the unit must be insulated.
- The unit must not be installed in an ATEX zone, and a drip tray for condensate/heat transfer fluid must be installed below the unit.
- A backflow preventer on the building side is required for the cooling water connection.
- The unit is a Group 1 Class A device according to EN55011, suitable for industrial EMC environments.
- Leakage current exceeds 3.5 mA, requiring appropriate protective earthing and RCD selection.
- Three-phase devices under 63 A are supplied with cable but without plug; a suitable plug must be provided on site.
- Replacement power cables are limited to 4 m for 3-phase connections.
- The heat transfer fluid must be brought to room temperature before switching off to avoid overpressure.
- Cooling water supply should be shut off during downtimes.
- The optional Flow Control Cube enables precise flow rate measurement and control.
Integration & Operating Notes
The responsible body must install a condensation water / heat transfer fluid drip tray below the unit. Overall dimensions are 730x804x1738 mm and net weight is 511 kg, so adequate floor space and structural support are required. Minimum wall clearances depend on air- or water-cooling and on whether the unit runs with or without a bath.
The unit requires a 400 V 3~ 50 Hz power supply with protective earth. Three-phase devices under 63 A are supplied with cable but without plug, so a suitable plug must be provided on site. Leakage current exceeds 3.5 mA, requiring appropriate protective earthing and RCD selection. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Cooling water inlet temperature must be between 5 °C and 20 °C, with a differential pressure of at least 1 bar and a maximum pressure of 6 bar. Only drinking water or filtered river water may be used; sea water is prohibited. A backflow preventer on the building side is required. If cooling water temperature falls below the condensation point, internal pipes must be insulated.
The Pilot ONE controller provides a display resolution of 0.01 K and interfaces including Ethernet, USB (Host and Device), and RS232. An external Pt100 sensor can be connected. The optional Flow Control Cube enables precise flow rate measurement and control, which may be needed for accurate testing.
Cooling water supply should be shut off even for shorter downtimes (e.g., overnight) because flow can occur even when the unit is switched off. Before switching off, the heat transfer fluid must be brought to room temperature to avoid overpressure in the system.
Selection & Integration Q&A
Choose the GL 07w when a 12 kW heating capacity and -45 to 95 °C range with water cooling suit your process. The 10w and 20w offer different capacities; the XT variants may provide extended features. Select based on required heating/cooling power and temperature range.
Cooling water must be 5–20 °C, 1–6 bar differential pressure, max 6 bar, with G3/4 male connections. Use only drinking or filtered river water; sea water is prohibited. Install a backflow preventer and insulate pipes if below condensation point.
A suitable plug for the 400 V 3~ power cable must be provided on site. A backflow preventer and drip tray are required. The optional Flow Control Cube is not included. Regular maintenance includes shutting off cooling water during downtimes and bringing fluid to room temperature before shutdown.