Peter Huber Kältemaschinenbau Unimotive GL 20w-XT
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unimotive GL 20w-XT

Peter Huber Kältemaschinenbau Unimotive GL 20w-XT — Temperature control units.

The Unimotive GL 20w-XT is a water-cooled temperature control unit with a 24 kW heater, a magnetically coupled stainless-steel pump, and a -45…150 °C working range using water-ethylene glycol. It is the extended-temperature (XT) variant of the 20w series, distinguished by its low-end capability and R-744 first-stage refrigerant. It serves automotive temperature simulation and material testing where precise, wide-range thermal conditioning is required.

Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection

Model Display resolutionHeating capacity, WPower supplyPump connectionRefrigerantSafety class
Unimotive GL 20w-XT Current 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive 10w 0.0112000400V 3~ 50HzM38x1,5 AG maleR-452A (A1, H280)III / FL
Unimotive 10w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 20w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 20w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 26w 0, 01 K24000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive 26w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)I / NFL
Unimotive 27w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-452A (A1, H280)III / FL
Unimotive GL 07w 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 07w-XT 0, 0112000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 10w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 10w-XT 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 20w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 30w 0, 0124000460V 3~ 60HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 50w-XT 0, 0124000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL
Unimotive GL 60w 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)III / FL
Unimotive GL 60w-XT 0, 0148000400V 3~ 50HzM38x1, 5 AG maleR-744 (A1, H280)I / NFL

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 24 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable no
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -45...150 °C
Dimensions, Weight & Mechanical Design
Weight 700 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-744 (A1, H280) (1st stage)
Maximum viscosity 200 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 2100 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The GL 20w-XT is a water-cooled, IP20-rated unit built for indoor installation. Its pump is magnetically coupled and stainless steel, with a fixed (non-adjustable) speed and an M38x1.5 male connection. The XT suffix denotes the extended low-temperature modification within the Unimotive 20w family; the sibling models 10w, 10w-XT, 20w, and 20w-XT differ in capacity and temperature span. The unit uses R-744 (CO2) as first-stage refrigerant, with a 3.3 kg charge and GWP 1.

Thermal behavior and media

Cooling capacity is strongly temperature-dependent: 28 kW at 20 °C falling to 6.8 kW at -40 °C, so the duty point must be checked against the actual process temperature. Published capacities are valid only at 20 °C room temperature, 15 °C cooling water inlet, and 1 bar differential pressure; higher cooling water temperature reduces capacity. The unit is designed for water-ethylene glycol mixtures with corrosion protection down to -45 °C, and the maximum permissible kinematic viscosity of the heat transfer fluid is 200 mm²/s. The pump delivers 201 l/min at zero head but only 42 l/min at 5.0 bar, so the external circuit resistance must be matched to the pump curve.

Installation, control, and operation

The Pilot ONE controller with POKO ONE control output provides a display resolution of 0.01 K and interfaces including Ethernet, USB (host and device), and RS232. An external Pt100 sensor connection is available. The unit requires a 400 V 3~ 50 Hz supply; because its current consumption exceeds 63 A, it is delivered without cable and without plug. Voltage and frequency tolerances are ±5% and ±2%, but the combination of -5% voltage with +2% frequency is not allowed. Cooling water must be supplied at 5–20 °C with a differential pressure of 1–4 bar and a maximum static pressure of 6 bar; the connection is G3/4 male. A backflow preventer on the building side is mandatory, and cooling water must be shut off during downtimes. The unit must not be installed in an ATEX zone, and a drip tray for condensate/heat transfer fluid must be provided below the unit.

Integration & Operating Notes

Mounting
Indoor, frost-protected location with drip tray required

The unit is IP20 and must operate between 5 °C and 40 °C ambient. Outdoor installation requires frost-protected cooling water lines and a minimum cooling water temperature of 5 °C. The responsible body must install a condensation water / heat transfer fluid drip tray below the unit.

Operation Manual, Section Commissioning; Datasheet
Electrical
400 V 3~ 50 Hz supply, no cable or plug supplied

Because current consumption exceeds 63 A, the unit is delivered without cable and without plug. Hardwiring must be done by a certified electrician, and the unit must always be connected to a socket with protective earth. Replacement power cables are limited to 4 m for 3-phase connections.

Datasheet; Operation Manual, Section Electrical Connection
Process Limits
Cooling water 5–20 °C, 1–4 bar differential, max 6 bar

Cooling water must be supplied between 5 °C and 20 °C with a differential pressure of 1–4 bar and must not exceed 6 bar static pressure. Only drinking water or filtered river/sea water may be used; sea water is prohibited. A backflow preventer on the building side is required.

Datasheet; Operation Manual, Section Installation
Maintenance
Shut off cooling water during downtimes

Cooling water supply should be shut off even for shorter downtimes (e.g. overnight) because flow can occur even when the unit is switched off. Before switching off, the heat transfer fluid must be brought to room temperature to avoid overpressure in the system.

Operation Manual, Section Normal Operation
Process Limits
Check pump curve against external circuit resistance

The circulation pump delivers 201 l/min at zero head but only 42 l/min at 5.0 bar. The operating point must be matched to the pressure drop of the external application. For pressure-sensitive applications such as glass reactors, the maximum flow pressure must be considered and an external bypass may be required.

Datasheet; Operation Manual, Section Performance

Selection & Integration Q&A

When should I choose the Unimotive GL 20w-XT over the 20w or the 10w-XT?

Choose the GL 20w-XT when you need the extended -45 °C low end and the higher heating capacity of the 20w frame. The 20w without XT has a narrower low-temperature range; the 10w-XT has lower capacity. Select based on required cooling capacity at your lowest process temperature.

What are the requirements for the cooling water circuit and adjacent plumbing?

Supply cooling water at 5–20 °C with a differential pressure of 1–4 bar and a maximum static pressure of 6 bar. Use the G3/4 male connection. Install a backflow preventer on the building side and provide a drip tray below the unit. Only drinking water or filtered river/sea water is permitted; sea water is prohibited.

What are the ambient and duty-cycle limits for installation?

The unit is IP20 and must be installed indoors between 5 °C and 40 °C ambient. It must not be placed in an ATEX zone. For outdoor installation, cooling water lines must be frost-protected and cooling water must not fall below 5 °C. Shut off cooling water during downtimes.

What maintenance and accessories are required that are not included in the base delivery?

The unit is delivered without a power cable and plug due to >63 A current. A drip tray and a building-side backflow preventer are required but not supplied. Use water-ethylene glycol mixture with corrosion protection. Replacement power cables are limited to 4 m for 3-phase connections.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unimotive GL 20w-XT
Datasheet ·96,9 КБ ·2 pp.
View PDF
PDF
Unimotive Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF