Peter Huber Kältemaschinenbau Unistat T320 Dynamisches
Peter Huber Kältemaschinenbau Unistat T320 Dynamisches — Temperature control units.
The Unistat T320 Dynamisches is a 12 kW heating, 65–300 °C dynamic temperature control unit with a magnetically coupled stainless-steel variable-speed pump (up to 96 l/min, 3.5 bar) and 24 l expansion tank. It has no active refrigeration, so the 65 °C lower limit depends on application heat loss or water cooling. It serves closed process loops requiring precise, pressure-controlled high-temperature tempering.
Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection
| Model | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C | Expansion volume, m³ |
|---|---|---|---|---|---|---|
| Unistat T320 Dynamisches Current | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Chili Dynamisches | 0.01 | 200-240V 1~/2~ 50/60Hz | M24x1,5 AG male | Pressure pump, variable | 65–300 | 0.002 |
| Unistat T305 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305 HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305w HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T320w HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330w HT Dynamisches | 0, 01 K | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T340 Dynamisches | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T340 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T350 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T402 Dynamisches | 0, 01 | 400V 3~N 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 80–425 | — |
| Unistat TR401 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 50–400 | 0.005 |
| Unistat TR402 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 80–425 | — |
Full specifications
| Control output | POKO ONE (potential-free contact) |
| Expansion volume | 24 l |
| Filling volume | 3 l (min.) |
| Heating capacity | 12 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | 65...300 °C |
| Weight | 168 kg (net) |
| Current consumption | 17 A (max.) |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s (max. permissible) |
| Ambient temperature limit | 5...40 °C (permissible) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The T320 Dynamisches is the 12 kW heating variant of the Unistat dynamic series, distinguished from the T305 siblings by its higher heating capacity and 65–300 °C range. It uses a magnetically coupled stainless-steel pump with soft start and integrated pressure control, optionally with an external pressure sensor.
- The expansion tank (24 l, not thermoregulated) is lockable for open baths and must remain unlocked for externally closed applications. Minimum filling volume is 3 l.
- The unit is rated IP20, safety class III / FL per DIN 12876 with adjustable overtemperature protection, and falls under PED Article 4.3.
Behavior under load and media
- No active refrigeration: reaching 65 °C requires application heat loss between 0.1 and 2.8 kW (depending on unit) or suitable water cooling; ambient must be well below 60 °C.
- Maximum permissible kinematic viscosity of the circulating fluid is 50 mm²/s, constraining thermofluid choice at low temperatures.
- Heating power 12 kW and pump delivery up to 96 l/min at 3.5 bar define heat-up rate and loop pressure budget.
Installation, control and operation
- Supply is 460 V 3~ 60 Hz, 17 A max, 3x20 A fusing; three-phase devices under 63 A come with cable but no plug. Voltage/frequency tolerances are coupled: ±5 % voltage only with ±2 % frequency; -5 % voltage with +2 % frequency is not allowed.
- Pump connections are M30x1.5 male, bore shape Y (60°) per DIN 3863, mating ball socket per DIN 3870.
- Control via Pilot ONE with POKO ONE potential-free contact, Ethernet/USB/RS232, Pt100 external sensor, 0.01 K display resolution. Process Safety mode requires compressor set to Permanently ON; Power Failure Auto resumes control after outage.
- Transport and store upright only; empty before moving; allow ~2 h acclimation after temperature change; leveling feet/parking brakes before commissioning.
Integration & Operating Notes
The unit cannot cool below ambient on its own. To reach the 65 °C lower limit, the application heat loss must fall between 0.1 and 2.8 kW depending on the unit, or water cooling must be provided; ambient must be well below 60 °C. Specifying the T320 for lower temperatures without an external heat sink will fail.
A ±5 % voltage deviation is only permissible with ±2 % frequency deviation; the combination of -5 % voltage and +2 % frequency is explicitly not allowed. Verify site power quality before commissioning to avoid invalid supply conditions and equipment stress.
Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so the site must provide a suitable plug and 3x20 A fusing. Hard-wired connection must be carried out by an electrician and the supply must match the rating plate.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications it must be locked to prevent fluid loss. Set and verify this before commissioning, as the wrong state damages the application or loses fluid.
Only transport and store the unit upright and emptied; horizontal transport damages the compressor and filled transport risks overflow. After moving between cold and warm environments, allow about 2 hours acclimation before power-on to prevent condensation-induced electrical faults. Casters are not for transport and carry only 25 % of total mass symmetrically.
The maximum permissible kinematic viscosity of the circulating fluid is 50 mm²/s, which limits thermofluid selection at low temperatures and keeps pump load and flow within design limits. Verify the chosen fluid's viscosity at the lowest operating temperature.
When Process Safety mode is used to protect against exothermic reactions, the compressor must be set to Permanently ON so cooling remains active at the overtemperature trigger. Configure this in the Pilot ONE before relying on the safety function.
Shut off the cooling water supply to the unit even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Periodically check hoses and connections for leaks and quality and replace as needed.
Selection & Integration Q&A
Pick the T320 when you need 12 kW heating and a 65–300 °C range for a closed loop; the T305 variants offer lower heating capacity. The T305 HT is for higher-temperature duty. All lack active refrigeration, so the 65 °C minimum still depends on heat loss or water cooling.
Pump connections are M30x1.5 male with bore shape Y (60°) per DIN 3863; hoses need ball sockets per DIN 3870. Supply is 460 V 3~ 60 Hz, 17 A max, 3x20 A fusing; the unit ships with cable but no plug, and voltage/frequency tolerances are coupled (±5 % only with ±2 %).
Operate between 5 and 40 °C ambient at IP20 in a dry, protected area; never inside an ATEX zone. Transport and store upright and empty, allow ~2 h acclimation after temperature change, and activate leveling feet/parking brakes before commissioning. Casters are not for transport.
The site must provide the 460 V plug and 3x20 A fusing; tempering hoses and DIN 3870 ball sockets are not included. Periodically check hoses for leaks and replace as needed, and shut off cooling water even overnight. An external pressure sensor for the pump is optional, not standard.