Peter Huber Kältemaschinenbau Unistat T402 Dynamisches

Peter Huber Kältemaschinenbau Unistat T402 Dynamisches — Temperature control units.

The Unistat T402 Dynamisches is a pressure-pump temperature control unit with a variable-speed pump, 6 kW heating, and a working range of 80 to 425 °C, positioned above the T305 siblings in the Unistat series. It requires a 400 V three-phase supply at up to 17 A and is intended for closed-loop heating of external circuits with high-temperature heat transfer fluids.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat T402 Dynamisches Current 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305 HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305w HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T320w HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330 HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 Dynamisches 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T340 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T350 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat TR401 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005
Unistat TR402 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE (potential-free contact)
Heating capacity 6 kW
Pump connection M24x1, 5 AG male
Pump speed adjustable Yes
Pump type Pressure pump, variable speed
External sensor connection Pt100
Temperature range 80...425 °C
Dimensions, Weight & Mechanical Design
Weight 62 kg (net)
Electrical, Power & Drive Specifications
Current consumption 17 A (max)
Power supply 400V 3~N 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Cooling water connection G1/2 male
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation and operation

Integration & Operating Notes

Electrical
Three-phase supply must be sized for 17 A per phase

The unit draws up to 17 A per phase at 400 V 3~N 50/60 Hz. The site electrical infrastructure, including breaker and cable sizing, must be rated accordingly. Three-phase devices under 63 A are delivered with cable but without a plug, so the customer must provide and install a suitable plug or hard-wire the connection.

Datasheet, Specifications; Operation Manual, Commissioning
Process Limits
Expansion vessel must be locked for open applications

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications the expansion vessel must be locked to prevent fluid loss. This setting must be verified before commissioning.

Operation Manual, Commissioning
Process Limits
Nitrogen blanket pressure must not exceed 0.2 bar

The expansion tank lid can be fixed and has connections for a nitrogen blanket. Exceeding 0.2 bar could damage the tank or create a safety hazard. Ensure the nitrogen supply is regulated to this maximum.

Datasheet, Specifications
Maintenance
Shut off cooling water during downtimes

Shut off the cooling water supply even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Periodically check hoses and connections for leaks and quality, and replace as needed.

Operation Manual, Commissioning; Service and Maintenance
Control & I/O
Process Safety mode requires compressor permanently ON

When using Process Safety mode for exothermic process protection, the compressor must be set to Permanently ON to keep cooling active at the overtemperature trigger. This ensures cooling remains available if the overtemperature protection activates.

Operation Manual, Normal Operation

Selection & Integration Q&A

When should I choose the Unistat T402 Dynamisches over the T305 or T305 HT siblings?

Choose the T402 when your process requires temperatures up to 425 °C and 6 kW heating. The T305 and T305 HT cover lower temperature ranges; select the T402 for high-temperature closed-loop applications where the T305's range is insufficient.

What are the requirements for connecting the external circuit and cooling water?

The pump connection is M24x1.5 male, so hoses or adapters must match this thread. The cooling water connection is G1/2 male, requiring a G1/2 female coupling on site. Ensure the external circuit's pressure drop does not exceed the pump's 0.9 bar maximum.

What are the ambient and installation limits for this unit?

The ambient temperature must stay between 5 °C and 40 °C. The unit must be installed upright, on a level surface with leveling feet or casters locked. It must not be installed in an ATEX zone. Allow 2 hours acclimation after temperature changes before power-on.

What maintenance is required and what accessories are not included?

Periodically check hoses and connections for leaks and replace as needed. Shut off cooling water during downtimes. The unit is delivered with a cable but without a plug; the customer must provide a suitable plug. No other accessories are included in base delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat T402 Temperature Control System Datasheet
Datasheet ·17,8 КБ ·2 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF