Peter Huber Kältemaschinenbau CC-106A
Peter Huber Kältemaschinenbau CC-106A — Thermostatic baths.
The CC-106A is a 4.4-litre immersion circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated for 25–100 °C (20–100 °C with water cooling, 15–100 °C with refrigerator). It differs from the CC-104A/108A/110A/112A siblings by its 4.4 l bath volume and 1.5–2.1 kW heating capacity, serving as a temperature-control source for external circuits or open baths in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection
| Model | Controller | Interface | Max delivery flow rate, liter/minute | Pump connection | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| CC-106A Current | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 5 | 5–40 |
| CC-104A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-108A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 6 | 5–40 |
| CC-110A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-112A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-118A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 104A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male | 5 | 5–40 |
| KISS 106A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 5 | 5–40 |
| KISS 108A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 110A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 112A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 118A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
Full specifications
| Dimensions (w × D × H) | 147 x 307 x 330 mm |
| Maximum delivery flow rate | 27 l/min (pressure pump max.; 0, 7 bar) suction pump max.; 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male (Option, #19606) |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | 25...100 °C with cooling unit: 15...100 °C · with water cooling: 20...100 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Bath opening WxD/BD mfr | 130 x 110 / 150 mm |
| Bath volume mfr | 4, 4 l |
| Filling volume min. mfr | 2, 5 l (min.) |
| Weight | 5 kg |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Immersion circulator with 4.4 l bath volume, 130 x 110 mm bath opening (150 mm depth), and 2.5 l minimum filling capacity.
- Safety class III / FL per DIN 12876 with adjustable overtemperature protection; IP20 protection; 5 kg dry weight.
- Sibling models CC-104A, CC-108A, CC-110A and CC-112A share the series architecture; the CC-106A sits at the 4.4 l / 1.5–2.1 kW point of that size ladder.
Thermal and hydraulic behaviour
- Temperature stability ±0.02 K; heating power scales with supply voltage: 1.5 kW at 200 V, 1.6 kW at 208 V, 1.8 kW at 220 V, 2.0 kW at 230 V, 2.1 kW at 240 V.
- Variable-speed pressure and suction pump: pressure side up to 27 l/min at 0.7 bar; suction side up to 22 l/min at 0.4 bar.
- Operating range is cooling-dependent: 25–100 °C without cooling, 20–100 °C with water cooling, 15–100 °C with refrigerator.
Control, connection and operation
- Pilot ONE controller with Ethernet, USB host/device and RS232 interfaces; external Pt100 sensor connection.
- M16x1 AG male pump connection is an option (#19606); polyglycol must not be used as heat transfer fluid when the Huber pump adapter is fitted.
- Single- and two-phase units are delivered with a country-specific plug that must be specified at ordering; external branch circuit protection to UL 489 is required.
Integration & Operating Notes
The unit draws up to 10 A on a 200–240 V 1~/2~ 50/60 Hz supply and requires a minimum 10 A fuse (maximum 16 A). External branch circuit protection according to UL 489 must be installed; voltage and frequency tolerances are interdependent, so a −10% voltage combined with +3% frequency is not permitted.
The achievable lower limit depends on the cooling configuration: 25 °C without cooling, 20 °C with water cooling, and 15 °C with a refrigerator. Specify the cooling option at ordering if the process requires temperatures below 25 °C.
When the optional Huber pump adapter (M16x1 AG male, #19606) is fitted, polyglycol is not permissible as heat transfer fluid. Select an alternative fluid compatible with the pump adapter and the wetted materials before commissioning.
The bath must be filled with a minimum of 2.5 l (4.4 l total volume) to ensure correct pump and heating behaviour. Operating with insufficient fluid risks pump cavitation and heater over-temperature.
The circulator is rated for an ambient temperature of 5–40 °C. Installation outside this window affects temperature stability and may trip the overtemperature protection; the IP20 enclosure also requires a dry, dust-controlled location.
Single- and two-phase devices (100–240 V) are delivered with a power cable and a country-specific plug that must be stated on the order. Missing specification delays commissioning because the plug cannot be assumed.
The unit provides adjustable overtemperature protection per DIN 12876 (safety class III / FL). The cut-out threshold must be configured during commissioning to match the heat transfer fluid and process, otherwise the safety function does not protect the application.
The 4-year warranty requires registration. Complete the registration at commissioning so that warranty coverage is secured before any service claim.
Selection & Integration Q&A
Pick the CC-106A when the process needs a 4.4 l bath with 1.5–2.1 kW heating and ±0.02 K stability. The CC-104A is smaller; the CC-108A/110A/112A step up in bath size. Match the bath volume to the external loop and immersion depth (150 mm) before selecting.
Provide 200–240 V 1~/2~ 50/60 Hz with a 10 A minimum fuse (16 A maximum) and external UL 489 branch protection. The unit ships with a country-specific plug that must be specified at ordering; voltage/frequency tolerance combinations such as −10% V with +3% Hz are not allowed.
The base unit requires registration to activate the 4-year warranty. The M16x1 AG male pump connection (#19606) is an option, and polyglycol is not permitted with the Huber pump adapter. External Pt100 sensors and the cooling source (water or refrigerator) are not part of the base delivery.