Peter Huber Kältemaschinenbau KISS 112A
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS 112A

Peter Huber Kältemaschinenbau KISS 112A — Thermostatic baths.

The KISS 112A is a 12-litre immersion circulator with a 275 x 161 mm bath opening and a non-variable pressure/suction pump, operating from 25 to 100 °C without cooling, 20 to 100 °C with water cooling, and 15 to 100 °C with a refrigerator. Its KISS controller, OLED display and ±0.05 K stability distinguish it within the series, while the 1.5–2.1 kW heater and 200–240 V single- or two-phase supply define its electrical envelope. It serves as a temperature-control source for external circuits or open baths in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
KISS 112A Current KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
CC-104A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-106A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-108A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-110A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-112A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
CC-118A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
KISS 104A KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
KISS 106A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
KISS 108A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 110A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 118A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic
Bath volume 12 l
Dimensions (w × D × H) 333 x 360 x 335 mm
Maximum delivery flow rate 14 l/min (pressure pump max.; 0, 25 bar)
Pump connection M16x1 AG male (Option, #19606) (Option, #19606)
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability mfr ±0, 05 K
Heating capacity mfr 1, 5 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min (suction pump max.; 0, 17 bar)
Bath opening WxD/BD mfr 275 x 161 / 150 mm
Filling volume min. mfr 7, 3 l (min.)
Dimensions, Weight & Mechanical Design
Weight 8 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz V, Hz
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour and media

Installation, control and operation

Integration & Operating Notes

Electrical
Branch circuit must be sized for 10 A at 200–240 V

The unit draws up to 10 A from a 200–240 V single- or two-phase 50/60 Hz supply. An external branch circuit protection device according to UL 489 is mandatory, and the supply must not combine opposing voltage and frequency tolerances (e.g. -10% voltage with +3% frequency is not allowed).

KISS 112A Immersion Circulator Datasheet, page 2
Cooling & Media
Low-temperature range requires water cooling or a refrigerator

Without cooling the unit operates from 25 to 100 °C; with water cooling the lower limit extends to 20 °C, and with a refrigerator to 15 °C. The cooling method must be selected at the system design stage to match the required process temperature.

KISS 112A Immersion Circulator Datasheet, page 1
Process Limits
Avoid acetone and polyglycols with the standard plastic pump

The plastic pump is not resistant to acetone or polyglycols (depending on the manufacturer). These fluids must be avoided or a more resistant plastic option ordered. For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.

KISS 112A Immersion Circulator Datasheet, page 2
Mounting
Install only in a dry location; IP20 enclosure

The enclosure is rated IP20, so it must be installed in a dry location protected from water ingress. Ambient temperature must stay between 5 °C and 40 °C for specified performance.

KISS 112A Immersion Circulator Datasheet, page 1
Control & I/O
Autostart restarts after power failure — assess in risk analysis

The autostart function restarts the unit automatically after a power failure. This behaviour must be considered in the risk assessment, especially for unattended operation or processes where automatic restart is hazardous.

KISS 112A Immersion Circulator Datasheet, page 1
Electrical
Specify country-specific plug at ordering

Single- and two-phase devices (100 V to 240 V) are delivered with a power cable and country-specific plug, which must be specified when ordering. The pump connection M16x1 AG male is an option (#19606).

KISS 112A Immersion Circulator Datasheet, page 2

Selection & Integration Q&A

When should I choose the KISS 112A over the CC-104A through CC-110A siblings?

The KISS 112A is the 12-litre variant with a 275 x 161 mm bath opening and 150 mm depth. The CC-104A to CC-110A models are listed in the same series but their capacities are not specified in the provided facts; select the KISS 112A when a 12-litre bath and the KISS controller with OLED display and ±0.05 K stability are required.

What power supply and branch protection does the KISS 112A require?

It requires 200–240 V, 1~ or 2~, 50/60 Hz, with a maximum current of 10 A. An external branch circuit protection device according to UL 489 is mandatory. Voltage and frequency tolerances are ±10% only if they do not oppose each other.

What are the ambient and installation limits for the KISS 112A?

Ambient temperature must be between 5 °C and 40 °C. The enclosure is IP20, so it must be installed in a dry location protected from water ingress. The unit is classified EN 55011 Class A Group 1 for industrial environments, not residential.

What maintenance and accessory considerations apply to the KISS 112A?

The plastic pump is not resistant to acetone or polyglycols; for freeze protection, mix water with glysantine or ethylene glycol. The M16x1 AG male pump connection is an option (#19606). A country-specific plug must be specified at ordering; the power cable is included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS 112A Immersion Circulator Datasheet
Datasheet ·107,6 КБ ·2 pp.
View PDF