Peter Huber Kältemaschinenbau KISS 112A
Peter Huber Kältemaschinenbau KISS 112A — Thermostatic baths.
The KISS 112A is a 12-litre immersion circulator with a 275 x 161 mm bath opening and a non-variable pressure/suction pump, operating from 25 to 100 °C without cooling, 20 to 100 °C with water cooling, and 15 to 100 °C with a refrigerator. Its KISS controller, OLED display and ±0.05 K stability distinguish it within the series, while the 1.5–2.1 kW heater and 200–240 V single- or two-phase supply define its electrical envelope. It serves as a temperature-control source for external circuits or open baths in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection
| Model | Controller | Interface | Max delivery flow rate, liter/minute | Pump connection | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| KISS 112A Current | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-104A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-106A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 5 | 5–40 |
| CC-108A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 6 | 5–40 |
| CC-110A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-112A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-118A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 104A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male | 5 | 5–40 |
| KISS 106A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 5 | 5–40 |
| KISS 108A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 110A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 118A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
Full specifications
| Alarm output | optic, acoustic |
| Bath volume | 12 l |
| Dimensions (w × D × H) | 333 x 360 x 335 mm |
| Maximum delivery flow rate | 14 l/min (pressure pump max.; 0, 25 bar) |
| Pump connection | M16x1 AG male (Option, #19606) (Option, #19606) |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | 25...100 °C with cooling unit: 15...100 °C · with water cooling: 20...100 °C |
| Temperature stability mfr | ±0, 05 K |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min (suction pump max.; 0, 17 bar) |
| Bath opening WxD/BD mfr | 275 x 161 / 150 mm |
| Filling volume min. mfr | 7, 3 l (min.) |
| Weight | 8 kg |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz V, Hz |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The KISS 112A is a 12-litre immersion circulator with a 275 x 161 mm bath opening and 150 mm depth; minimum filling is 7.3 l.
- It uses a pressure and suction pump with fixed speed (not adjustable), delivering up to 14 l/min at 0.25 bar pressure and 10.5 l/min at 0.17 bar suction.
- The KISS controller with OLED display provides ±0.05 K temperature stability; heating capacity is 1.5–2.1 kW.
- Sibling models CC-104A through CC-110A are listed in the same series; the KISS 112A is the 12-litre variant, while the CC models are not further specified in the provided facts.
Thermal behaviour and media
- The usable temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator.
- The plastic pump is not resistant to acetone or polyglycols (depending on manufacturer); these fluids must be avoided or a more resistant plastic option ordered.
- For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.
- Safety class III / FL with adjustable overtemperature protection according to DIN 12876.
Installation, control and operation
- Power supply: 200–240 V, 1~ or 2~, 50/60 Hz, max. current 10 A; external branch circuit protection according to UL 489 is required.
- Voltage and frequency tolerances are ±10% only if they do not oppose each other (e.g. -10% voltage with +3% frequency is not allowed).
- Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.
- Enclosure is IP20, so installation must be in a dry location protected from water ingress; ambient temperature must stay between 5 °C and 40 °C.
- Interfaces: USB (Device) and RS232; alarm output is optic and acoustic.
- The autostart function restarts the unit automatically after a power failure, which must be considered in the risk assessment.
- EMC classification: EN 55011 Class A Group 1, intended for industrial environments, not residential.
Integration & Operating Notes
The unit draws up to 10 A from a 200–240 V single- or two-phase 50/60 Hz supply. An external branch circuit protection device according to UL 489 is mandatory, and the supply must not combine opposing voltage and frequency tolerances (e.g. -10% voltage with +3% frequency is not allowed).
Without cooling the unit operates from 25 to 100 °C; with water cooling the lower limit extends to 20 °C, and with a refrigerator to 15 °C. The cooling method must be selected at the system design stage to match the required process temperature.
The plastic pump is not resistant to acetone or polyglycols (depending on the manufacturer). These fluids must be avoided or a more resistant plastic option ordered. For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.
The enclosure is rated IP20, so it must be installed in a dry location protected from water ingress. Ambient temperature must stay between 5 °C and 40 °C for specified performance.
The autostart function restarts the unit automatically after a power failure. This behaviour must be considered in the risk assessment, especially for unattended operation or processes where automatic restart is hazardous.
Single- and two-phase devices (100 V to 240 V) are delivered with a power cable and country-specific plug, which must be specified when ordering. The pump connection M16x1 AG male is an option (#19606).
Selection & Integration Q&A
The KISS 112A is the 12-litre variant with a 275 x 161 mm bath opening and 150 mm depth. The CC-104A to CC-110A models are listed in the same series but their capacities are not specified in the provided facts; select the KISS 112A when a 12-litre bath and the KISS controller with OLED display and ±0.05 K stability are required.
It requires 200–240 V, 1~ or 2~, 50/60 Hz, with a maximum current of 10 A. An external branch circuit protection device according to UL 489 is mandatory. Voltage and frequency tolerances are ±10% only if they do not oppose each other.
Ambient temperature must be between 5 °C and 40 °C. The enclosure is IP20, so it must be installed in a dry location protected from water ingress. The unit is classified EN 55011 Class A Group 1 for industrial environments, not residential.
The plastic pump is not resistant to acetone or polyglycols; for freeze protection, mix water with glysantine or ethylene glycol. The M16x1 AG male pump connection is an option (#19606). A country-specific plug must be specified at ordering; the power cable is included.