Peter Huber Kältemaschinenbau Unimotive GL 20w
Peter Huber Kältemaschinenbau Unimotive GL 20w — Temperature control units.
The Unimotive GL 20w is a water-cooled temperature control unit with a magnetically coupled stainless-steel pump, covering -45 to 95 °C with 24 kW heating. It is the GL modification of the 20w series, intended for water-ethylene glycol mixtures with corrosion protection down to -45 °C. It serves automotive test benches and processes requiring precise temperature control with plant cooling water.
Peter Huber Kältemaschinenbau Unimotive Series — Model Comparison & Selection
| Model | Display resolution | Heating capacity, W | Power supply | Pump connection | Refrigerant | Safety class |
|---|---|---|---|---|---|---|
| Unimotive GL 20w Current | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive 10w | 0.01 | 12000 | 400V 3~ 50Hz | M38x1,5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 10w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 20w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 20w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 26w | 0, 01 K | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive 26w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | I / NFL |
| Unimotive 27w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-452A (A1, H280) | III / FL |
| Unimotive GL 07w | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 07w-XT | 0, 01 | 12000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 10w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 10w-XT | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 20w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 30w | 0, 01 | 24000 | 460V 3~ 60Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 50w-XT | 0, 01 | 24000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
| Unimotive GL 60w | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | III / FL |
| Unimotive GL 60w-XT | 0, 01 | 48000 | 400V 3~ 50Hz | M38x1, 5 AG male | R-744 (A1, H280) | I / NFL |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 24 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | no |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -45...95 °C |
| Weight | 612 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-744 (A1, H280) (1st stage) |
| Maximum viscosity | 200 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 2100 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The Unimotive GL 20w is a water-cooled temperature control unit in the Unimotive series, distinguished from sibling models (10w, 10w-XT, 20w, 20w-XT) by its GL designation and 20w size class. It uses a magnetically coupled stainless-steel pump with fixed speed (not adjustable) and M38x1.5 male pump connections. The refrigerant is R-744 (CO2) in the first stage, with a charge of 3.5 kg and GWP 1. The unit weighs 612 kg and has IP20 protection.
Working range and media
- Temperature range: -45 to 95 °C, with a display resolution of 0.01 K.
- Heating capacity: 24 kW.
- Intended for water-ethylene glycol mixtures with corrosion protection (e.g., Glysantin®) down to -45 °C.
- Cooling water: connection G3/4 male, consumption 2100 l/h at 15 °C cooling water and 0 °C flow, minimum differential pressure 1 bar, maximum pressure 6 bar.
- Cooling water intake temperature must be between 5 °C and 20 °C; higher temperatures reduce cooling capacity and may require increased flow.
Control and operation
The unit is equipped with the Pilot ONE controller and POKO ONE control output. Interfaces include Ethernet, USB (Host and Device), and RS232. An external Pt100 sensor can be connected. The pump speed is not adjustable. The unit requires a 400 V 3~ 50 Hz supply with maximum current 55.7 A and 3x63 A fusing. Three-phase devices below 63 A are supplied with cable but without a plug. Leakage current exceeds 3.5 mA, requiring appropriate protective earthing. The unit is a Group 1 Class A device per EN55011, suitable for industrial EMC environments only.
Installation and site requirements
Ambient temperature must be 5–40 °C. The unit must not be installed in an ATEX zone. A drip tray for condensate/heat transfer fluid must be installed below the unit. Minimum wall clearances depend on cooling type and operation with or without a bath. Cooling water must be drinking water or filtered river water (sea water prohibited), with a backflow preventer on the building side. Cooling water supply should be shut off during downtimes. Outdoor installation requires frost-protected lines and cooling water temperature not below 5 °C. Power cables are length-limited to 4 m for 3-phase connections. Pressure-sensitive applications (e.g., glass reactors) must account for maximum flow pressure and may need an external bypass. Heat transfer fluid must be brought to room temperature before switching off to avoid overpressure.
Integration & Operating Notes
Exceeding 20 °C reduces cooling capacity and may require higher flow rates; below 5 °C risks frost damage. The datasheet specifies this intake range, and the manual prohibits cooling water below 5 °C for outdoor installation. Plan the plant cooling water supply accordingly.
The unit's leakage current is above 3.5 mA, which affects protective earthing design and residual current device selection. Ensure the installation provides a reliable protective earth connection and that any RCD is rated for high leakage currents to avoid nuisance tripping.
The unit draws up to 55.7 A at 400 V 3~ 50 Hz and must be protected with 3x63 A fuses. Three-phase devices below 63 A are delivered with a cable but without a plug, so site wiring must include a suitable connector or hardwired connection by a certified electrician.
The responsible body must install a condensation water / heat transfer fluid drip tray below the temperature control unit to contain leaks or condensate. This is mandatory to prevent damage to the floor and surroundings, especially when operating at low temperatures where condensation occurs.
The unit is constructed for operation with water-ethylene glycol mixture including corrosion protection (e.g., Glysantin®) down to -45 °C. Using other fluids may cause corrosion, fouling, or inadequate performance. Minimum filling capacity is 8.9 l, and expansion tank capacity is 19.9 l.
Cooling water can flow through the unit even when switched off, so shut off the supply during shorter downtimes (e.g., overnight) to prevent unnecessary water consumption and potential water damage. Also, bring heat transfer fluid to room temperature before switching off to avoid overpressure.
The Pilot ONE controller provides Ethernet, USB (Host and Device), and RS232 interfaces, allowing integration into automation systems. An external Pt100 sensor can be connected for precise temperature measurement at the application. Ensure cabling and network configuration meet industrial EMC requirements (Group 1 Class A).
Selection & Integration Q&A
Choose the GL 20w when you need water-ethylene glycol operation down to -45 °C with 24 kW heating and water cooling. The 20w is air-cooled; the 10w has lower capacity. The GL 20w is specifically built for glycol mixtures with corrosion protection, making it suitable for automotive test fluids.
Cooling water connection is G3/4 male. Supply must be 5–20 °C, differential pressure 1–4 bar, max 6 bar, flow 2100 l/h at 15 °C. Use drinking or filtered river water (no sea water), install a backflow preventer, and provide a drip tray. Shut off water during downtimes.
The unit is delivered with cable but without a plug for 3-phase. A drip tray and backflow preventer are site-provided. Heat transfer fluid (water-glycol mixture) must be supplied separately; min filling 8.9 l. Regular maintenance includes checking cooling water quality and shutting off water during downtimes.