No photo available yet

Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat TR401 Dynamisches

Peter Huber Kältemaschinenbau Unistat TR401 Dynamisches — Temperature control units.

The Unistat TR401 Dynamisches is a water-cooled, pressure-pump temperature control unit with a split Pilot ONE controller, covering 50 to 400 °C. It is the TR401 controller variant of the Unistat TR series, distinguished from TR400/TR402 siblings by its separate HT-Circulator and Controller connection. It serves high-temperature process loops requiring dynamic, variable-speed circulation.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat TR401 Dynamisches Current 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305 HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305w HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T320w HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330 HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 Dynamisches 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T340 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T350 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T402 Dynamisches 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Unistat TR402 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE (potential-free contact)
Expansion volume 5 l
Pump connection M24x1, 5 (Flow) M24x1, 5 (Return)
Pump speed adjustable Yes
Pump type Pressure pump, variable speed
External sensor connection Pt100
Temperature range 50...400 °C
Dimensions, Weight & Mechanical Design
Weight 55 kg (net)
Electrical, Power & Drive Specifications
Current consumption 14 A (max)
Fuse rating 16 A
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and behavior

Installation, connection and control

Integration & Operating Notes

Electrical
Supply tolerance must not combine voltage and frequency drift

The ±10 % voltage tolerance applies only as long as the frequency tolerance does not run in the opposite direction. Verify site supply quality before connection, since a combined deviation outside this envelope can damage the unit.

Unistat TR401 Technical Data, page 2
Electrical
External UL 489 branch protection is mandatory

The equipment requires external branch circuit protection according to UL 489. Size the upstream breaker for the 14 A maximum current and 16 A fuse rating, and have hard-wired connection performed by an electrician.

Unistat TR401 Technical Data, page 2; Unistat Operation Manual, page 37
Process Limits
Expansion vessel lock state depends on open or closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications it must be locked to prevent fluid loss. Confirm the application type before commissioning.

Unistat Operation Manual, page 33
Process Limits
Air in the circuit trips low-level protection

Low-level protection relies on a pressure sensor in the fluid circuit. Insufficient fill level or inadequate venting prevents the specified pressure from being reached, interrupting temperature control and circulation. Fill and vent thoroughly before starting.

Unistat Operation Manual, page 19
Mounting
Acclimation and upright transport before power-on

Transport and store the unit upright only; horizontal transport damages the compressor. Allow about 2 hours acclimation when moving between cold and warm environments before switching on. Activate leveling feet and parking brakes before commissioning.

Unistat Operation Manual, pages 22 and 29
Cooling & Media
Shut off cooling water during short downtimes

Close the cooling water supply even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Periodically check hoses and connections for leaks and quality and replace as required.

Unistat Operation Manual, page 29
Control & I/O
Process Safety needs compressor set to Permanently ON

Process Safety mode keeps cooling active at the overtemperature trigger only if the compressor is configured to Permanently ON via System Settings > Power/ECO Settings. Configure this before relying on the mode for exothermic process protection.

Unistat Operation Manual, page 19
Control & I/O
Controller cable length constrains cabinet layout

The Pilot ONE controller connects to the machine by a cable of approximately 5 m. Plan the control cabinet or operator station position within this distance, and note that controller connection applies specifically to TR401 and TR402 variants.

Unistat TR401 Technical Data, page 1; Unistat TR Operation Manual, page 32

Selection & Integration Q&A

When should I choose the TR401 over the TR400 or TR402 siblings?

Choose TR401 when a split HT-Circulator and Controller arrangement with a roughly 5 m controller cable is required. TR400 does not use this controller connection; TR402 is a separate controller variant. Confirm the controller variant against the process layout before ordering.

What supply and protection must be provided for this unit?

Provide 200–240 V, 1~ or 2~, 50/60 Hz with max 14 A and a 16 A fuse, plus external branch circuit protection to UL 489. At 230 V the current is just below 16 A and can be limited to 13 A or 10 A at first start-up. Hard-wiring must be done by an electrician.

What site conditions and duty limits apply to installation?

Install indoors at 5...40 °C ambient with IP20 protection, on a level surface with leveling feet and parking brakes engaged. Do not install in an ATEX zone without an Ex px cabinet. Avoid recurring very dynamic high-temperature changes in short intervals; the guarantee expires under such cyclic loading.

What maintenance and accessories are needed beyond base delivery?

Base delivery includes a power cable with country-specific plug (specify at order). Plan for M24x1.5 process hoses and G1/2 cooling water fittings, a drip tray below the unit, periodic hose inspection and replacement, and cleaning of the hat-type strainer on water-cooled units.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat TR401 Technical Data
Datasheet ·54,7 КБ ·2 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF
PDF
Unistat® TR 7 Temperature Control Unit — Operation Manual
Manual ·3,3 МБ ·92 pp.
View PDF