Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat T340 Dynamisches

Peter Huber Kältemaschinenbau Unistat T340 Dynamisches — Temperature control units.

The Unistat T340 Dynamisches is a closed, magnetically coupled stainless-steel circulation system with 48 kW heating and a 65–300 °C working range, distinguished within the Unistat series by its high heating capacity and 56 l expansion volume. It has no active refrigeration, so the 65 °C lower limit depends on application heat loss or water cooling. It serves high-temperature process thermostating in closed loops.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat T340 Dynamisches Current 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305 HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305w HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T320w HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330 HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T350 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T402 Dynamisches 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Unistat TR401 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005
Unistat TR402 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—

Full specifications

Core Performance & Functional Specifications
Alarm output optic, acoustic, relay
Control output POKO ONE (potential-free contact)
Expansion volume 56 l
Heating capacity 48 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range 65...300 °C
Dimensions, Weight & Mechanical Design
Weight 373 kg (net)
Electrical, Power & Drive Specifications
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s (max. permissible)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Control, connection and operation

Integration & Operating Notes

Process Limits
65 °C lower limit needs external heat loss or water cooling

Because the unit has no active refrigeration, the minimum setpoint of 65 °C is only reachable when the application dissipates heat or suitable water cooling is provided. Size the cooling water circuit and plan heat dissipation before installation; for air-cooled operation the heat loss must lie between 0.1 kW and 2.8 kW and ambient must be well below 60 °C.

Unistat T340 Temperature Control System Technical Data, page 2
Electrical
Provide a plug on site and design earthing for >3.5 mA leakage

Three-phase devices drawing less than 63 A are delivered with cable but without a plug, so the installer must fit a suitable plug. Leakage current exceeds 3.5 mA, which governs protective earthing and residual current device selection. Supply tolerance is ±5% voltage with ±2% frequency, and the combination of -5% voltage with +2% frequency is not allowed.

Unistat T340 Temperature Control System Technical Data, page 2
Process Limits
Expansion vessel lock state depends on open or closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For externally open applications the expansion vessel must be locked to prevent fluid loss. Verify the setting during commissioning according to the actual circuit type.

Unistat Operation Manual, page 33
Control & I/O
Process Safety mode requires compressor set to Permanently ON

When Process Safety mode is used to protect against exothermic reactions, the compressor must be configured to Permanently ON so cooling remains active at the overtemperature trigger. Power Failure Auto resumes temperature control automatically after an outage if it was active, which affects unattended operation risk assessment.

Unistat Operation Manual, pages 19 and 20

Selection & Integration Q&A

When should I choose the Unistat T340 Dynamisches instead of a smaller sibling such as the T305 Dynamisches or T305 HT Dynamisches?

Pick the T340 when the process needs the 48 kW heating capacity and 56 l expansion volume for large or high-load closed circuits in the 65–300 °C range. Smaller siblings suit lower heat demand; the Chili Dynamisches is a different line.

What must I provide on the utility and piping side to connect this unit to my process loop?

Match pipework or flexible hoses to the M38x1.5 male pump connection with bore shape Y (60°) per DIN 3863. Provide a plug on site for the 460 V 3~ 60 Hz supply, and plan water cooling or sufficient application heat loss to reach 65 °C.

What site and duty-cycle conditions does the manual impose on installation and unattended operation?

Install level with parking brakes or leveling feet set, outside any ATEX zone, and allow about 2 hours acclimation after a temperature change before power-on. Power Failure Auto can resume control after an outage, so assess unattended risk accordingly.

What maintenance and consumables are required, and what is not included in base delivery?

Periodically inspect and replace hoses and hose connections, and shut off cooling water during downtimes. The unit ships with cable but no plug for three-phase supply below 63 A; the plug, cooling water hookup and thermal fluid are not part of base delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat T340 Temperature Control System Technical Data
Datasheet ·17,8 КБ ·2 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF