The CC-525w is a water-cooled refrigerated circulating bath with a 17 l bath, a variable-speed pressure/suction pump and a -55…200 °C working range, positioned above the smaller CC-405/410/415 and CC-505 siblings in the series. Its 3 kW heater and water-cooled condenser suit continuous external-circuit temperature control where a 400 V three-phase feed and a 3–6 bar cooling-water supply are available.

Direct drop-in replacements

Eppendorf
MixMate ®
43%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-525w Kälte-Badumwälzthermostat Current —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Bath volume 17 l
Filling volume 15 l (min.)
Heating capacity 3 kW
Maximum delivery flow rate 25 l/min (pressure pump max.; 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range -55...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 539 x 629 x 1102 mm
Delivery capacity suction pump max. mfr 18, 5 l/min (suction pump max.; 0, 4 bar)
Bath opening WxD/BD mfr 270 x 150 / 200
Dimensions, Weight & Mechanical Design
Weight 163 kg
Electrical, Power & Drive Specifications
Current consumption 13 A (max.)
Power supply 400V 3~N 50Hz
Fuse mfr 3x16A
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 205 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 3 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Cooling capacity mfr 100:7; 20:7; 0:5; -20:3; -40:1, 5 °C | kW
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and load behavior

Installation, connection and operation

Integration & Operating Notes

Electrical
Three-phase feed with 3x16 A fusing and no supplied plug

The unit draws up to 13 A on a 400 V 3~N 50 Hz supply and requires a 3x16 A fuse; because three-phase devices under 63 A ship with cable but without a plug, the mating plug and breaker must be procured and the feed sized before commissioning.

CC-525w Circulating Bath Datasheet, Technical Specifications
Cooling & Media
Cooling water must sit between 3 and 6 bar differential pressure

The water circuit needs at least 3 bar differential pressure to reach rated cooling and must never exceed 6 bar; intake temperature must stay within 5…20 °C, and pipes below the condensation point need insulation to avoid condensation on the unit and circuit.

CC-525w Circulating Bath Datasheet, Installation
Process Limits
Medium viscosity above 50 mm²/s overloads the pump

The variable-speed pressure/suction pump is rated for media up to 50 mm²/s kinematic viscosity; thicker fluids at low temperature raise circuit resistance beyond the 0.7 bar delivery / 0.4 bar suction envelope and must be avoided or pre-diluted.

CC-525w Circulating Bath Datasheet, Technical Specifications
Electrical
Voltage and frequency tolerances must occur together

A ±5% voltage deviation is only permissible with a simultaneous ±2% frequency deviation, and certain combinations are explicitly not allowed; site power quality should be checked against this before connecting the 3 kW heater and refrigeration circuit.

CC-525w Circulating Bath Datasheet, Installation
Maintenance
Refrigerant charge falls under F-gas leak-check obligations

The R-452A charge of 1.9 kg corresponds to 4.1 t CO2 equivalent (GWP 2141), so periodic leak checks and documentation are required under F-gas rules; plan access to the refrigeration circuit and keep the charge record with the unit.

CC-525w Circulating Bath Datasheet, Technical Specifications
Control & I/O
External Pt100 and digital interfaces support remote control

The Pilot ONE controller accepts an external Pt100 sensor and offers Ethernet, USB host/device and RS232, allowing the bath temperature to be controlled at the external load rather than in the bath; wiring and network provisioning must be planned with the installation.

CC-525w Circulating Bath Datasheet, Technical Specifications

Selection & Integration Q&A

When should I choose the CC-525w rather than the smaller CC-405w, CC-410wl, CC-415wl or CC-505wl siblings?

Pick the CC-525w when the 17 l bath, 3 kW heating and the -55…200 °C water-cooled envelope are needed for larger external loads; the smaller siblings suit compact bath duties where the higher cooling capacity and bath volume are not required.

What must the external circuit and cooling-water supply provide to mate with this unit?

Pump ports are M16x1 male for the external loop (up to 25 l/min at 0.7 bar, 18.5 l/min suction at 0.4 bar); the cooling-water side uses a G1/2 male connection needing 3–6 bar differential pressure and 205 l/h at 15 °C.

What is not included in base delivery and what service items should be planned?

Three-phase units under 63 A arrive with cable but no plug, so the mating plug is a separate item; the R-452A charge (4.1 t CO2e) brings F-gas leak-check obligations, and bath fluid plus displacement rack (reducing fill to 10 l) are user-supplied.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-525w Circulating Bath Datasheet
Datasheet ·71,5 КБ ·2 pp.
View PDF