Peter Huber Kältemaschinenbau CC-K6 Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-K6 Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau CC-K6 Kälte-Badumwälzthermostat — Thermostatic baths with Noise level 57 dB.

The CC-K6 is a compact refrigerated/heating circulator with a 4.5 l bath and a variable-speed pressure/suction pump, covering -25 to 200 °C with ±0.02 K stability. It is the small-bath member of the Cc series, distinguished by its 140×120 mm opening and 2.5 l minimum fill. It serves as a temperature-control source for external circuits and immersed samples in laboratory processes.

Noise level
57 dB

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-K6 Kälte-Badumwälzthermostat Current 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Noise level 57 dB(A)
Maximum delivery flow rate 27 l/min (pressure pump max.; 0, 7 bar)
suction pump max.; 0, 4 bar: 22 l/min
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range -25...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 210 x 400 x 546 mm
Heating capacity mfr 1, 6 - 2, 1 kW
Bath opening WxD/BD mfr 140 x 120 / 150
Bath volume mfr 4, 5 l
Filling volume min. mfr 2, 5 l (min.)
Dimensions, Weight & Mechanical Design
Weight 25 kg
Electrical, Power & Drive Specifications
Power supply 208-240V 1~/2~ 50/60Hz
Current consumption max. mfr 10 + 1, 35 A (max.)
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Cooling capacity mfr 20:0, 2; 0:0, 15; -10:0, 1; -20:0, 05 °C / kW
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior

Circulation, control and connection

Integration & Operating Notes

Electrical
Mains tolerance combination is restricted

Voltage tolerance is ±5% and frequency tolerance ±2%, but the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify the site's worst-case mains condition against this rule before commissioning, since an out-of-rule supply can prevent safe operation.

CC-K6 Datasheet, page 2
Electrical
Upstream fuse must be sized 10–16 A

The circuit must be protected with a fuse between 10 A and 16 A, matched to the maximum current draw of 10 + 1.35 A. Undersized fusing causes nuisance trips; oversized fusing leaves the unit inadequately protected.

CC-K6 Datasheet, page 2
Electrical
Country-specific plug must be ordered

Single- and two-phase units are delivered with a power cable and a country-specific plug that must be specified at ordering. Without this specification the unit may arrive without a usable mains connection for the installation site.

CC-K6 Datasheet, page 2
Mounting
Air-cooled condenser needs cool intake air

The refrigerating unit is air-cooled, so the installation site must provide sufficient clearance and cool intake air. Restricted airflow or hot intake air reduces cooling capacity and can lead to insufficient temperature control at the low end of the range.

CC-K6 Datasheet, page 1
Mounting
IP20 rules out wet or washdown locations

The degree of protection is IP20, so the unit must be kept away from water spray and conductive dust. Placement in washdown areas or near liquid handling that can splash the enclosure is not permissible.

CC-K6 Datasheet, page 2
Process Limits
Low-end cooling capacity limits heat loads near -20 °C

Cooling capacity is only 0.05 kW at -20 °C, so the unit is unsuitable for sustained high heat loads at the low end of its range. Size the process heat load against the capacity curve at the intended setpoint rather than the nominal range limit.

CC-K6 Datasheet, page 1
Process Limits
External loop pressure drop must fit the pump curve

The pressure/suction pump delivers up to 27 l/min at 0.7 bar, which sets the external circuit pressure-drop limit. Calculate the pressure drop of the external loop and immersed load against this curve to avoid insufficient circulation and poor temperature uniformity.

CC-K6 Datasheet, page 1
Safety
Flammable refrigerant imposes ignition-source rules

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, so the installation location must respect ignition-source and ventilation rules. The small 0.047 kg charge with GWP 0.02 keeps leak-reporting burden minimal but does not remove the A3 placement requirements.

CC-K6 Datasheet, page 1
Safety
Safety class III / FL governs flammable media use

The unit is classified as safety class III / FL per DIN 12876, which determines the permitted flammable-liquid use and the required protection concept. Confirm the intended bath medium against this classification before operation.

CC-K6 Datasheet, page 1
Environment
Ambient above 40 °C is outside the permissible range

The permissible ambient temperature range is 5 °C to 40 °C, and cooling capacity may drop as ambient temperature rises. Rooms that exceed this range or run hot near the upper limit degrade performance and should be assessed before installation.

CC-K6 Datasheet, page 2
Compliance
EMC class A suits industrial, not residential, sites

The equipment is EMC class A, group 1 per EN55011, so it is intended for industrial environments and may cause interference in residential areas. Account for this when siting the unit near sensitive electronics or in light-industrial locations.

CC-K6 Datasheet, page 2

Selection & Integration Q&A

When should I choose the CC-K6 rather than a larger CC-series sibling such as the CC-405w or CC-505wl?

Pick the CC-K6 when the load fits a 4.5 l bath with a 140×120 mm opening and a 2.5 l minimum fill, and when the external loop's pressure drop is within the 27 l/min at 0.7 bar pump curve. Larger siblings suit bigger immersed loads or higher heat loads near -20 °C, where the CC-K6 delivers only 0.05 kW.

What must I verify about the external circuit and its connections before integrating the CC-K6?

The pump connections are M16×1 male, so mating fittings must match that thread. Calculate the external loop pressure drop against the pump curve (27 l/min at 0.7 bar pressure side, 22 l/min at 0.4 bar suction side) to avoid insufficient circulation. An external Pt100 can be connected for control at the load.

What installation and duty-cycle limits apply to the CC-K6 from the datasheet?

Ambient must stay within 5–40 °C, with cooling capacity dropping as ambient rises. The air-cooled condenser needs clearance and cool intake air. IP20 excludes wet or washdown locations, and the A3 refrigerant (R-290) imposes ignition-source and ventilation rules at the installation site.

What is not included in the base delivery and what must be specified or planned at ordering?

The power cable is supplied with a country-specific plug that must be specified at ordering; without it the unit may arrive without a usable mains connection. Upstream fusing of 10–16 A is a site responsibility. M16×1 mating fittings for the pump connections and any external Pt100 sensor are not part of the base unit.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-K6 Refrigerated/Heating Circulator Datasheet
Datasheet ·135,3 КБ ·2 pp.
View PDF