Peter Huber Kältemaschinenbau CC-K6 Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau CC-K6 Kälte-Badumwälzthermostat — Thermostatic baths with Noise level 57 dB.
The CC-K6 is a compact refrigerated/heating circulator with a 4.5 l bath and a variable-speed pressure/suction pump, covering -25 to 200 °C with ±0.02 K stability. It is the small-bath member of the Cc series, distinguished by its 140×120 mm opening and 2.5 l minimum fill. It serves as a temperature-control source for external circuits and immersed samples in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-K6 Kälte-Badumwälzthermostat Current | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| CC-410wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-510w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-515w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-520w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-525w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-905 Kälte-Badumwälzthermostat | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-905w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-906w Kälte-Badumwälzthermostat | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-K12 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-K15 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Noise level | 57 dB(A) |
| Maximum delivery flow rate | 27 l/min (pressure pump max.; 0, 7 bar) suction pump max.; 0, 4 bar: 22 l/min |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -25...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 210 x 400 x 546 mm |
| Heating capacity mfr | 1, 6 - 2, 1 kW |
| Bath opening WxD/BD mfr | 140 x 120 / 150 |
| Bath volume mfr | 4, 5 l |
| Filling volume min. mfr | 2, 5 l (min.) |
| Weight | 25 kg |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 10 + 1, 35 A (max.) |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Cooling capacity mfr | 20:0, 2; 0:0, 15; -10:0, 1; -20:0, 05 °C / kW |
| Safety class | III / FL |
Description
Construction and series positioning
- Bath volume 4.5 l, opening 140 × 120 mm, depth 150 mm; minimum filling volume 2.5 l.
- Air-cooled refrigerating unit using R-290 (A3, H220), charge 0.047 kg, GWP 0.02.
- Safety class III / FL per DIN 12876; EMC class A, group 1 per EN55011.
- Compared with the larger CC-405w/410wl/415wl/505wl siblings, the CC-K6 is the smallest-bath variant, suited to small immersed loads and short external loops rather than large bath volumes.
Thermal behavior
- Working range -25 to 200 °C; temperature stability ±0.02 K.
- Heating capacity 1.6–2.1 kW.
- Cooling capacity falls steeply at the low end: 0.2 kW at 20 °C, 0.15 kW at 0 °C, 0.1 kW at -10 °C, 0.05 kW at -20 °C. Sustained high heat loads near -20 °C are not supportable.
- Cooling capacity may drop as ambient temperature rises within the 5–40 °C permissible range.
Circulation, control and connection
- Variable-speed pressure and suction pump: max delivery 27 l/min at 0.7 bar (pressure side), 22 l/min at 0.4 bar (suction side); pump connections M16×1 male.
- Pilot ONE controller; interfaces Ethernet, USB host and device, RS232; external Pt100 sensor connection.
- Power supply 208–240 V, 1~ or 2~, 50/60 Hz; max current 10 + 1.35 A; IP20.
Integration & Operating Notes
Voltage tolerance is ±5% and frequency tolerance ±2%, but the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify the site's worst-case mains condition against this rule before commissioning, since an out-of-rule supply can prevent safe operation.
The circuit must be protected with a fuse between 10 A and 16 A, matched to the maximum current draw of 10 + 1.35 A. Undersized fusing causes nuisance trips; oversized fusing leaves the unit inadequately protected.
Single- and two-phase units are delivered with a power cable and a country-specific plug that must be specified at ordering. Without this specification the unit may arrive without a usable mains connection for the installation site.
The refrigerating unit is air-cooled, so the installation site must provide sufficient clearance and cool intake air. Restricted airflow or hot intake air reduces cooling capacity and can lead to insufficient temperature control at the low end of the range.
The degree of protection is IP20, so the unit must be kept away from water spray and conductive dust. Placement in washdown areas or near liquid handling that can splash the enclosure is not permissible.
Cooling capacity is only 0.05 kW at -20 °C, so the unit is unsuitable for sustained high heat loads at the low end of its range. Size the process heat load against the capacity curve at the intended setpoint rather than the nominal range limit.
The pressure/suction pump delivers up to 27 l/min at 0.7 bar, which sets the external circuit pressure-drop limit. Calculate the pressure drop of the external loop and immersed load against this curve to avoid insufficient circulation and poor temperature uniformity.
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, so the installation location must respect ignition-source and ventilation rules. The small 0.047 kg charge with GWP 0.02 keeps leak-reporting burden minimal but does not remove the A3 placement requirements.
The unit is classified as safety class III / FL per DIN 12876, which determines the permitted flammable-liquid use and the required protection concept. Confirm the intended bath medium against this classification before operation.
The permissible ambient temperature range is 5 °C to 40 °C, and cooling capacity may drop as ambient temperature rises. Rooms that exceed this range or run hot near the upper limit degrade performance and should be assessed before installation.
The equipment is EMC class A, group 1 per EN55011, so it is intended for industrial environments and may cause interference in residential areas. Account for this when siting the unit near sensitive electronics or in light-industrial locations.
Selection & Integration Q&A
Pick the CC-K6 when the load fits a 4.5 l bath with a 140×120 mm opening and a 2.5 l minimum fill, and when the external loop's pressure drop is within the 27 l/min at 0.7 bar pump curve. Larger siblings suit bigger immersed loads or higher heat loads near -20 °C, where the CC-K6 delivers only 0.05 kW.
The pump connections are M16×1 male, so mating fittings must match that thread. Calculate the external loop pressure drop against the pump curve (27 l/min at 0.7 bar pressure side, 22 l/min at 0.4 bar suction side) to avoid insufficient circulation. An external Pt100 can be connected for control at the load.
Ambient must stay within 5–40 °C, with cooling capacity dropping as ambient rises. The air-cooled condenser needs clearance and cool intake air. IP20 excludes wet or washdown locations, and the A3 refrigerant (R-290) imposes ignition-source and ventilation rules at the installation site.
The power cable is supplied with a country-specific plug that must be specified at ordering; without it the unit may arrive without a usable mains connection. Upstream fusing of 10–16 A is a site responsibility. M16×1 mating fittings for the pump connections and any external Pt100 sensor are not part of the base unit.