Peter Huber Kältemaschinenbau CC-508w Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-508w Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau CC-508w Kälte-Badumwälzthermostat — Thermostatic baths with Noise level 57 dB.

The CC-508w is a water-cooled circulating bath thermostat with a 5 L bath and a variable-speed pressure/suction pump, covering -55 to 200 °C with ±0.02 K stability. It is the 5 L water-cooled variant in the CC series, distinguished from air-cooled siblings by its 5–20 °C cooling water requirement. It serves as a temperature-control source for external circuits via M16x1 connections.

Noise level
57 dB

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-508w Kälte-Badumwälzthermostat Current 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Noise level 57 dB(A)
Bath volume 5 l
Filling volume 4 l (min)
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range -55...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 410 x 480 x 764 mm
Heating capacity mfr 2, 45 kW
Delivery capacity pressure pump max. mfr 25 l/min; 0, 7 bar
Delivery capacity suction pump max. mfr 18, 5 l/min; 0, 4 bar
Bath opening WxD/BD mfr 120 x 110 / 160 mm
Electrical, Power & Drive Specifications
Power supply 208-230V 1~/2~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Cooling capacity mfr 100:1, 5; 20:1, 5; 0:1, 5; -20:0, 8; -40:0, 25 k (at °C bath temperature)
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The CC-508w is a refrigerated circulating bath (Kälte-Badumwälzthermostat) with a 5 L bath volume and a minimum filling volume of 4 L. It is water-cooled, requiring a cooling water supply at 5–20 °C, which differentiates it from air-cooled siblings such as the CC-405w, CC-410wl, CC-415wl and CC-505wl. The bath opening measures 120 x 110 mm (W x D) with a 160 mm depth. The unit uses the Pilot ONE controller and is rated IP20, safety class III/FL.

Thermal and hydraulic behavior

Installation, control and operation

Integration & Operating Notes

Electrical
Voltage and frequency tolerances are coupled and restricted

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%; the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify site supply quality against these combined limits before commissioning to avoid invalid operating conditions. The unit is supplied for 208–230 V, 1~/2~, 60 Hz.

CC-508w Circulating Bath Datasheet, page 2
Mounting
Dry, splash-free location within 5–40 °C ambient required

IP20 protection does not guard against water ingress, so the installation site must be dry and protected from splashes. The permissible ambient temperature range is 5–40 °C; operating outside this range can affect performance, reliability and safety. Plan the location with these environmental limits in mind.

CC-508w Circulating Bath Datasheet, pages 1–2
Process Limits
Pumped medium viscosity must not exceed 50 mm²/s

The maximum permissible kinematic viscosity of the circulated medium is 50 mm²/s. Higher-viscosity fluids can overload the pump and reduce circulation performance. Select the bath fluid and external circuit medium accordingly, especially at low temperatures where viscosity increases.

CC-508w Circulating Bath Datasheet, page 1
Electrical
Plug must be specified at order for single/two-phase units

Single- and two-phase devices (100–240 V) are delivered with a power cable and a country-specific plug, which must be specified when ordering. Three-phase devices below 63 A are supplied with cable but without plug, so a suitable plug must be provided on site. Confirm the plug type against the site outlet before order.

CC-508w Circulating Bath Datasheet, page 2
Safety
Flammable refrigerant R-1270 requires site safety assessment

The refrigerant is R-1270 with safety group A3 and GHS H220, meaning it is flammable. This affects ventilation requirements, handling procedures and the site safety assessment. The Global Warming Potential is 0, which supports environmental compliance but does not remove the flammability classification.

CC-508w Circulating Bath Datasheet, page 1
Control & I/O
External Pt100 and digital interfaces support system integration

The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, plus an external Pt100 sensor connection. These allow remote monitoring and control or connection to a external temperature sensor for controlling the external circuit. Plan cabling and network access accordingly.

CC-508w Circulating Bath Datasheet, page 1

Selection & Integration Q&A

When should I choose the CC-508w over the air-cooled CC-405w or CC-505wl?

Choose the CC-508w when a 5–20 °C cooling water supply is available and higher cooling capacity at low temperatures is needed; it is water-cooled and reaches -55 °C. Air-cooled siblings like the CC-405w avoid water connections but may have different capacity or temperature limits; verify their datasheets for your duty point.

What are the requirements for connecting the external circuit to the CC-508w pump?

The pump connections are M16x1 male, so hoses and fittings must match this thread. The pump is a variable-speed pressure and suction type, with max delivery 25 l/min at 0.7 bar on the pressure side and 18.5 l/min at 0.4 bar on the suction side. Ensure the medium viscosity does not exceed 50 mm²/s.

What ambient and installation limits apply to the CC-508w?

Install in a dry, splash-free location with ambient temperature between 5 and 40 °C. IP20 protection requires protection from water ingress. The unit is water-cooled and needs a cooling water supply at 5–20 °C; if water temperature falls below the condensation point, insulate internal cooling water pipes.

What maintenance and accessories are required that are not included in the base delivery?

The base delivery includes a power cable with country-specific plug for single/two-phase units (plug type must be specified at order). Cooling water connections and hoses are not included; provide M16x1 mating fittings. No maintenance schedule is given in the datasheet; follow the manual for service intervals. Refrigerant R-1270 (A3) requires appropriate handling.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-508w Circulating Bath Datasheet
Datasheet ·30,6 КБ ·2 pp.
View PDF