Peter Huber Kältemaschinenbau CC-515w Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-515w Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau CC-515w Kälte-Badumwälzthermostat — Thermostatic baths.

The CC-515w is a 18-litre refrigerated/heating bath circulator with a variable-speed pressure/suction pump and Pilot ONE controller, covering -55 to 200 °C. Its 3 kW heating and 3.3 kW cooling at 0 °C distinguish it within the CC series by higher capacity and a 400 V three-phase supply. It serves as a temperature-control source for external circuits or direct bath immersion in laboratory and pilot processes.

Direct drop-in replacements

Eppendorf
MixMate ®
50%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-515w Kälte-Badumwälzthermostat Current —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Bath volume 18 l
Filling volume 16 l
Heating capacity 3 kW
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range -55...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 455 x 515 x 1014 mm
Delivery capacity pressure pump max. mfr 25 l/min; 0, 7 bar
Delivery capacity suction pump max. mfr 18, 5 l/min; 0, 4 bar
Bath opening WxD/BD mfr 270 x 150 / 200
Dimensions, Weight & Mechanical Design
Weight 105 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~N 50Hz
Current consumption max. mfr 10, 5 A (max.)
Fuse mfr 3x16A
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 138 л/час (at 15°C, flow -20°C)
at 15°C, flow 0°C: 192 l/h
Cooling water differential pressure 3 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Cooling capacity mfr 100: 3, 3; 20: 3, 3; 0: 3, 3; -20: 1, 6; -40: 0, (at °C bath temperature)
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The CC-515w is a floor-standing refrigerated/heating circulator with an 18-litre bath (16-litre filling volume) and a 270 x 150 mm opening, 200 mm deep. It is the higher-capacity three-phase variant in the CC series; siblings such as the CC-405w, CC-410wl, CC-415wl and CC-505wl are single-phase or lower-volume configurations. The CC-515w requires a 400 V 3~N 50 Hz supply with a 3x16 A fuse and draws up to 10.5 A. The refrigerant is R-452A (A1, GWP 2141, 2.1 t CO2e).

Thermal and hydraulic behavior

Cooling capacity is 3.3 kW at 100 °C, 20 °C and 0 °C, falling to 1.6 kW at -20 °C and 0.6 kW at -40 °C; heating capacity is 3 kW. Temperature stability is ±0.02 K. The variable-speed pressure and suction pump delivers up to 25 l/min at 0.7 bar (pressure side) and 18.5 l/min at 0.4 bar (suction side), with M16x1 male connections. Maximum permissible kinematic viscosity is 50 mm²/s, which limits fluid selection at low temperatures.

Installation and control

The unit is air-cooled via a water circuit: cooling water connection is G1/2 male, requiring a differential pressure of at least 3 bar and not more than 6 bar, with intake temperature between 5 °C and 20 °C. Cooling water consumption is 192 l/h at 0 °C flow and 138 l/h at -20 °C flow. The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, and an external Pt100 sensor connection. Safety classification is III / FL with adjustable overtemperature protection to DIN 12876. Protection rating is IP20; ambient operating range is 5–40 °C. Overall dimensions are 455 x 515 x 1014 mm at 105 kg.

Integration & Operating Notes

Electrical
Three-phase supply must be within ±5% voltage and ±2% frequency

The 400 V 3~N 50 Hz supply must stay within ±5% voltage and ±2% frequency, and combinations such as -5% voltage with +2% frequency are not allowed. Verify site power quality before connection to avoid compressor or electronics damage.

CC-515w Circulating Bath Datasheet, page 2
Cooling & Media
Cooling water differential pressure must be 3–6 bar

The water circuit requires a minimum differential pressure of 3 bar and must not exceed 6 bar. Check the building supply and provide continuous flow of 192 l/h at 0 °C or 138 l/h at -20 °C; insufficient pressure prevents correct operation.

CC-515w Circulating Bath Datasheet, page 1
Process Limits
Bath fluid viscosity must not exceed 50 mm²/s

At low temperatures, fluid viscosity rises; exceeding 50 mm²/s will overload the pump and degrade control. Select a heat-transfer fluid whose kinematic viscosity at the lowest operating temperature remains below this limit.

CC-515w Circulating Bath Datasheet, page 1
Electrical
Three-phase cable supplied without plug

Devices below 63 A are delivered with a cable but no plug; the customer must provide a suitable site plug and have the connection made by a qualified electrician in accordance with the 3x16 A fuse rating.

CC-515w Circulating Bath Datasheet, page 2

Selection & Integration Q&A

When should I choose the CC-515w over the CC-505wl or CC-415wl?

Choose the CC-515w when you need 18 L bath volume, 3 kW heating and 3.3 kW cooling at 0 °C with a 400 V three-phase supply. The CC-505wl and CC-415wl are lower-capacity or single-phase variants; select them if three-phase power is unavailable or less cooling is sufficient.

What are the requirements for the cooling water circuit and external pump connections?

The cooling water circuit requires G1/2 male connections, differential pressure 3–6 bar, intake temperature 5–20 °C, and continuous flow of 192 l/h at 0 °C or 138 l/h at -20 °C. The pump uses M16x1 male connections and delivers up to 25 l/min at 0.7 bar pressure side and 18.5 l/min at 0.4 bar suction side.

What maintenance and accessories are required that are not included in base delivery?

The base delivery does not include a plug for the three-phase cable, bath fluid, or external Pt100 sensor. Regular maintenance includes checking cooling water flow and pressure, cleaning the condenser, and verifying overtemperature protection. Refrigerant R-452A handling requires certified personnel.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-515w Circulating Bath Datasheet
Datasheet ·79,6 КБ ·2 pp.
View PDF