Peter Huber Kältemaschinenbau CC-515w Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau CC-515w Kälte-Badumwälzthermostat — Thermostatic baths.
The CC-515w is a 18-litre refrigerated/heating bath circulator with a variable-speed pressure/suction pump and Pilot ONE controller, covering -55 to 200 °C. Its 3 kW heating and 3.3 kW cooling at 0 °C distinguish it within the CC series by higher capacity and a 400 V three-phase supply. It serves as a temperature-control source for external circuits or direct bath immersion in laboratory and pilot processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-515w Kälte-Badumwälzthermostat Current | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-410wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-510w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-520w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-525w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-905 Kälte-Badumwälzthermostat | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-905w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-906w Kälte-Badumwälzthermostat | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-K12 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-K15 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| CC-K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Bath volume | 18 l |
| Filling volume | 16 l |
| Heating capacity | 3 kW |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -55...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 455 x 515 x 1014 mm |
| Delivery capacity pressure pump max. mfr | 25 l/min; 0, 7 bar |
| Delivery capacity suction pump max. mfr | 18, 5 l/min; 0, 4 bar |
| Bath opening WxD/BD mfr | 270 x 150 / 200 |
| Weight | 105 kg |
| Power supply | 400V 3~N 50Hz |
| Current consumption max. mfr | 10, 5 A (max.) |
| Fuse mfr | 3x16A |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 138 л/час (at 15°C, flow -20°C) at 15°C, flow 0°C: 192 l/h |
| Cooling water differential pressure | 3 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Cooling capacity mfr | 100: 3, 3; 20: 3, 3; 0: 3, 3; -20: 1, 6; -40: 0, (at °C bath temperature) |
| Safety class | III / FL |
Description
Construction and series positioning
The CC-515w is a floor-standing refrigerated/heating circulator with an 18-litre bath (16-litre filling volume) and a 270 x 150 mm opening, 200 mm deep. It is the higher-capacity three-phase variant in the CC series; siblings such as the CC-405w, CC-410wl, CC-415wl and CC-505wl are single-phase or lower-volume configurations. The CC-515w requires a 400 V 3~N 50 Hz supply with a 3x16 A fuse and draws up to 10.5 A. The refrigerant is R-452A (A1, GWP 2141, 2.1 t CO2e).
Thermal and hydraulic behavior
Cooling capacity is 3.3 kW at 100 °C, 20 °C and 0 °C, falling to 1.6 kW at -20 °C and 0.6 kW at -40 °C; heating capacity is 3 kW. Temperature stability is ±0.02 K. The variable-speed pressure and suction pump delivers up to 25 l/min at 0.7 bar (pressure side) and 18.5 l/min at 0.4 bar (suction side), with M16x1 male connections. Maximum permissible kinematic viscosity is 50 mm²/s, which limits fluid selection at low temperatures.
Installation and control
The unit is air-cooled via a water circuit: cooling water connection is G1/2 male, requiring a differential pressure of at least 3 bar and not more than 6 bar, with intake temperature between 5 °C and 20 °C. Cooling water consumption is 192 l/h at 0 °C flow and 138 l/h at -20 °C flow. The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, and an external Pt100 sensor connection. Safety classification is III / FL with adjustable overtemperature protection to DIN 12876. Protection rating is IP20; ambient operating range is 5–40 °C. Overall dimensions are 455 x 515 x 1014 mm at 105 kg.
Integration & Operating Notes
The 400 V 3~N 50 Hz supply must stay within ±5% voltage and ±2% frequency, and combinations such as -5% voltage with +2% frequency are not allowed. Verify site power quality before connection to avoid compressor or electronics damage.
The water circuit requires a minimum differential pressure of 3 bar and must not exceed 6 bar. Check the building supply and provide continuous flow of 192 l/h at 0 °C or 138 l/h at -20 °C; insufficient pressure prevents correct operation.
At low temperatures, fluid viscosity rises; exceeding 50 mm²/s will overload the pump and degrade control. Select a heat-transfer fluid whose kinematic viscosity at the lowest operating temperature remains below this limit.
Devices below 63 A are delivered with a cable but no plug; the customer must provide a suitable site plug and have the connection made by a qualified electrician in accordance with the 3x16 A fuse rating.
Selection & Integration Q&A
Choose the CC-515w when you need 18 L bath volume, 3 kW heating and 3.3 kW cooling at 0 °C with a 400 V three-phase supply. The CC-505wl and CC-415wl are lower-capacity or single-phase variants; select them if three-phase power is unavailable or less cooling is sufficient.
The cooling water circuit requires G1/2 male connections, differential pressure 3–6 bar, intake temperature 5–20 °C, and continuous flow of 192 l/h at 0 °C or 138 l/h at -20 °C. The pump uses M16x1 male connections and delivers up to 25 l/min at 0.7 bar pressure side and 18.5 l/min at 0.4 bar suction side.
The base delivery does not include a plug for the three-phase cable, bath fluid, or external Pt100 sensor. Regular maintenance includes checking cooling water flow and pressure, cleaning the condenser, and verifying overtemperature protection. Refrigerant R-452A handling requires certified personnel.