Peter Huber Kältemaschinenbau CC-K15 Kälte-Badthermostat
Peter Huber Kältemaschinenbau CC-K15 Kälte-Badthermostat — Thermostatic baths.
The CC-K15 is a 15-litre refrigerated bath thermostat with a variable-speed pressure/suction pump and Pilot ONE controller, covering -20 to 200 °C at ±0.02 K stability. It is the smallest bath in the CC series, distinguished from the CC-4xx/5xx siblings by its 15 l volume and 0.25 kW cooling at 20 °C. It serves as a temperature-control source for external circuits or direct bath immersion in laboratory and pilot processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-K15 Kälte-Badthermostat Current | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-410wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-510w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-515w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-520w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-525w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-905 Kälte-Badumwälzthermostat | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-905w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-906w Kälte-Badumwälzthermostat | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-K12 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| CC-K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Bath volume | 15 l |
| Dimensions (w × D × H) | 350 x 560 x 430 mm |
| Pump connection | (Option, #19607) (Option) |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -20...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 1, 8 - 2, 1 kW |
| Delivery capacity pressure pump max. mfr | 27 l/min; 0, 7 bar |
| Delivery capacity suction pump max. mfr | 22 l/min; 0, 4 bar |
| Bath opening WxD/BD mfr | 290 x 152 / 200 |
| Filling volume min. mfr | 11, 3 l |
| Weight | 28 kg |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 10 + 1, 5 A |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Cooling capacity mfr | 20: 0, 25; 0: 0, 2; -10: 0, 12; -20: 0, 05 kW (at °C) |
| Safety class | III / FL |
Description
Construction and series positioning
The CC-K15 is a refrigerated bath thermostat built to DIN 12876 with a 15 l bath, a 290 x 152 mm opening (200 mm depth) and a minimum filling volume of 11.3 l. It uses an air-cooled refrigeration circuit with R-290 (A3, H220) natural refrigerant and a pressure/suction pump with variable speed. Within the CC series, the CC-K15 is the compact 15 l unit; the CC-405w, CC-410wl, CC-415wl and CC-505wl siblings are larger bath/circulator variants. The CC-K15 is the choice when the working volume and cooling demand fit 15 l and 0.25 kW at 20 °C, rather than the larger siblings.
Thermal behavior and load limits
- Working range: -20 to 200 °C; stability ±0.02 K.
- Cooling capacity is strongly temperature-dependent: 0.25 kW at 20 °C, 0.2 kW at 0 °C, 0.12 kW at -10 °C and 0.05 kW at -20 °C. Low-temperature loads must be sized against the coldest point, not the 20 °C figure.
- Heating capacity is 1.8–2.1 kW.
- The pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min at 0.4 bar suction, so external circuit pressure drop must be checked against these maxima.
Installation, control and operation
The unit requires a 220–240 V, 1~ or 2~, 50/60 Hz supply with a 10 A minimum and 16 A maximum fuse; current consumption is 10 + 1.5 A. Voltage and frequency tolerances are ±5 % and ±2 % but must not be combined in the unfavourable direction. A power cable with country-specific plug must be specified at order. The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, and an external Pt100 sensor connection. The unit is IP20 and Class A Group 1 for EMC, so it needs a dry indoor location; ambient temperature must stay between 5 and 40 °C, and cooling capacity drops as ambient rises. Overtemperature protection is adjustable per DIN 12876, safety class III / FL. Polyglycol must not be used as heat transfer fluid when the Huber pump adapter is fitted. A drain valve with cap (#6839) is available as an accessory.
Integration & Operating Notes
The unit draws up to 10 + 1.5 A and requires a branch protection of at least 10 A but not more than 16 A. A lower-rated breaker will trip on compressor start; a higher-rated one will not protect the internal wiring. Confirm the site breaker before commissioning.
Voltage tolerance is ±5 % and frequency tolerance ±2 %, but the unfavourable combination (-5 % voltage together with +2 % frequency) is explicitly not allowed. Check the site power quality against this rule before connecting the compressor and electronics.
The refrigeration machine is air-cooled, so the condenser must have adequate air flow and the stated space clearances. Ambient temperature must stay between 5 and 40 °C; cooling capacity drops as ambient rises, so a hot plant room erodes the published 0.25 kW at 20 °C.
Cooling capacity falls from 0.25 kW at 20 °C to 0.05 kW at -20 °C. A load that is comfortable at ambient may be unable to reach or hold -20 °C. Calculate the actual heat load at the coldest setpoint, including pump heat and external circuit gains, before committing to the lower end of the range.
When the Huber pump adapter is fitted, polyglycol must not be used as heat transfer fluid. Select a compatible fluid for the wetted parts and the application temperature range; using polyglycol will damage the pump adapter and invalidate the intended service.
A drain valve with cap is available as an accessory and simplifies bath draining during fluid changes and maintenance. Without it, draining the 15 l bath (minimum filling 11.3 l) is less practical. Specify it at order to avoid later rework.
Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be specified when ordering. Confirm the plug type and the 220–240 V 1~/2~ supply arrangement at the site before the order is placed.
Selection & Integration Q&A
Pick the CC-K15 when the bath volume fits 15 l (minimum filling 11.3 l) and the cooling demand is within 0.25 kW at 20 °C, falling to 0.05 kW at -20 °C. Choose a larger sibling when you need more bath volume or higher cooling capacity at low temperature.
The pressure/suction pump delivers up to 27 l/min at 0.7 bar and 22 l/min at 0.4 bar, so the external circuit pressure drop must fit within these limits. The supply must be 220–240 V 1~/2~ 50/60 Hz with a 10–16 A fuse, and the country-specific plug must be specified at order.
The drain valve with cap (#6839) is an accessory and is not part of the base unit. The power cable with country-specific plug must be specified at order. Heat transfer fluid is not included and must be selected to suit the application; polyglycol is not permitted with the Huber pump adapter.