Peter Huber Kältemaschinenbau Variostat Kältethermostat für variable Badgefäße — Thermostatic baths with Noise level 54 dB.

The Variostat is a circulating bath thermostat with a variable-speed pressure/suction pump and Pilot ONE controller, designed for use with variable bath vessels. Its -30 to 150 °C working range and 1 kW heating capacity distinguish it from fixed-bath CC siblings, while cooling capacity falls to 0.03 kW at -30 °C. It serves as a temperature-control source for external circuits requiring up to 25 l/min at 0.7 bar.

Noise level
54 dB

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
Variostat Kältethermostat für variable Badgefäße Current 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—

Full specifications

Core Performance & Functional Specifications
Noise level 54 dB(A)
Heating capacity 1 kW
Maximum delivery flow rate 25 l/min (pressure pump max.; 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
Refrigerator air-cooled, natural refrigerant
External sensor connection Pt100
Temperature range -30...150 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 183 x 493 x 443 mm
Delivery capacity suction pump max. mfr 18, 5 l/min (suction pump max.; 0, 4 bar)
Dimensions, Weight & Mechanical Design
Weight 25 kg
Electrical, Power & Drive Specifications
Power supply 220-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220)
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Cooling capacity mfr 100:0, 3; 20:0, 3; 0:0, 2; -10:0, 18; -20:0, 12;
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and operation

Integration & Operating Notes

Process Limits
Cooling capacity at -30 °C is only 0.03 kW

The published cooling capacity falls sharply at low temperatures, so the process heat load at the coldest set point must be verified against the curve. Exceeding the available capacity at the working temperature will prevent the bath from reaching or holding set point.

Datasheet, Technical Data
Electrical
Do not combine -5% voltage with +2% frequency

The supply tolerance is ±5% voltage and ±2% frequency, but the specific combination of -5% voltage and +2% frequency is not allowed. Site power quality must be checked before commissioning to avoid operating outside the permitted envelope.

Datasheet, Installation
Mounting
Air-cooled condenser needs cool intake air and clearance

Because the unit is air-cooled, sufficient clearance and cool intake air are required; cooling capacity drops as ambient temperature rises. The permitted ambient range is 5 to 40 °C, and installation must respect this for reliable operation.

Datasheet, Installation
Process Limits
Bath fluid viscosity must not exceed 50 mm²/s

The maximum permissible kinematic viscosity is 50 mm²/s. Very viscous fluids at low temperatures can overload the pump and impair circulation, so fluid selection must account for viscosity at the lowest operating temperature.

Datasheet, Technical Data
Safety
R-290 flammable refrigerant requires ignition-source review

The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant. This affects room ventilation and ignition-source assessment; installation location and safety review must account for the A3 classification.

Datasheet, Safety
Control & I/O
Remote control via USB and network port

The Pilot ONE controller supports remote control and data transmission via the USB port (with Spy software) and via the network port. This enables integration into lab data acquisition and remote operation.

Datasheet, Operation
Electrical
Country-specific plug must be specified at ordering

Single/two-phase devices are supplied with a power cable and a country-specific plug that must be specified when ordering. The correct mains connection for the installation site must be confirmed before delivery.

Datasheet, Installation
Safety
Set adjustable overtemperature protection for the application

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application to protect the bath and process from overheating if the control loop fails.

Datasheet, Safety

Selection & Integration Q&A

When should I choose the Variostat instead of a fixed-bath CC model such as the CC-405w or CC-505wl?

Choose the Variostat when the bath vessel is variable and must be connected externally; the CC-405w/410wl/415wl/505wl are fixed-bath circulating thermostats. The Variostat provides pressure and suction pump connections (M16x1) for external circuits up to 25 l/min at 0.7 bar.

What are the requirements for the external circuit plumbing and pump connections?

Pump connections are M16x1 male, so mating fittings and hoses must match this thread. The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar; external tubing and heat exchangers must be sized within this flow and pressure-drop budget.

What ambient and duty-cycle limits apply from the manual?

Permitted ambient range is 5 to 40 °C; the air-cooled condenser requires sufficient clearance and cool intake air, and capacity drops as ambient temperature rises. Supply tolerance is ±5% voltage and ±2% frequency, but -5% voltage with +2% frequency is not allowed.

What maintenance and accessories are not included in base delivery?

The power cable is supplied with a country-specific plug that must be specified at ordering. USB Spy software is referenced for PC connection but is not stated as included. No maintenance schedule or consumables are specified in the provided datasheet; overtemperature protection must be set per DIN 12876.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Variostat Circulating Bath Thermostats
Datasheet ·108,9 КБ ·2 pp.