Peter Huber Kältemaschinenbau CC-905 Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-905 Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau CC-905 Kälte-Badumwälzthermostat — Thermostatic baths with Noise level 64 dB.

The CC-905 is a refrigerated/heating circulator with a 26-litre bath and a working range of -90 to 200 °C, driven by a two-stage cascade using R-452A and R-508B. It is the deep-cold modification of the Cc series, distinguished by its -90 °C lower limit, and serves as a temperature-control source for external circuits and bath duties.

Noise level
64 dB

Direct drop-in replacements

Eppendorf
MixMate ®
50%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-905 Kälte-Badumwälzthermostat Current 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Noise level 64 dB(A)
Bath volume 26 l
Dimensions (w × D × H) 605 x 706 x 1136 mm
Filling volume 19 l (min)
Heating capacity 3 kW
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range -90...200 °C
Temperature stability mfr ±0, 02 K
Delivery capacity pressure pump max. mfr 25 l/min; 0, 7 bar
Delivery capacity suction pump max. mfr 18, 5 l/min; 0, 4 bar (suction pump)
Bath opening WxD/BD mfr 260 x 260 / 200
Dimensions, Weight & Mechanical Design
Weight 171 kg
Electrical, Power & Drive Specifications
Current consumption 14 A (max)
Power supply 400V 3~N 50Hz
Fuse mfr 3x16A
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behaviour across the working range

Installation, control and operation

Integration & Operating Notes

Electrical
Site must supply 400 V 3~N with 3x16 A fusing

The unit draws up to 14 A per the datasheet and requires a 400 V three-phase neutral supply at 50 Hz with 3x16 A fusing. Cable sizing, breaker selection and phase balance must be planned before delivery; the unit arrives with cable but no plug, so the mating plug must be sourced on site.

Datasheet, Technical Data (power supply, current consumption, fuse
Electrical
Do not combine -5 % voltage with +2 % frequency

Voltage and frequency tolerances are ±5 % and ±2 % respectively, but the unfavourable combination (-5 % voltage together with +2 % frequency) is explicitly not allowed. Verify site power quality against this rule before commissioning the compressor-driven unit.

Datasheet, Technical Data (voltage/frequency tolerance
Process Limits
Check low-temperature load against the cooling curve

Cooling capacity drops from 2 kW at 20 °C to 0.13 kW at -90 °C. Any process heat load at the cold end must be compared with the curve at the intended setpoint, otherwise pull-down time or hold temperature will not be met.

Datasheet, Specifications (cooling capacity
Process Limits
External loop pressure drop must fit 0.7 bar pump head

The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar. Hoses, external application and fittings must be sized so the circuit pressure drop stays within the available head at the required flow.

Datasheet, Specifications (delivery capacity
Process Limits
Heat-transfer fluid viscosity must stay below 50 mm²/s

The maximum permissible kinematic viscosity is 50 mm²/s. Fluid selection must account for viscosity rise at low temperature, and pump connections are M16x1 male, so hoses and adapters must match this thread.

Datasheet, Technical Data (viscosity, pump connection
Mounting
Install within 5–40 °C ambient and respect IP20

The unit is IP20 and rated for ambient temperatures of 5 to 40 °C. Cooling capacity data are valid at 20 °C room temperature and may drop as ambient rises, so a hot or cramped location erodes published duty and may trip the machine.

Datasheet, Technical Data (ambient temperature, IP rating
Maintenance
Refrigerant charge carries F-gas reporting obligations

R-452A (1.5 kg, GWP 2141, 3.2 t CO2e) and R-508B (0.42 kg, GWP 13396, 5.6 t CO2e) are A1, H280 pressurised gases under PED Art. 4.3. Leak-check intervals, labelling and refrigerant handling obligations follow from these charges.

Datasheet, Specifications (refrigerant data
Control & I/O
Pilot ONE offers Ethernet, USB and RS232 plus Pt100 input

The Pilot ONE controller provides Ethernet, USB host and device, and RS232 interfaces, and accepts an external Pt100 sensor. These interfaces allow remote monitoring and external-temperature control of the application.

Datasheet, Technical Data (controller, interface, sensor connection

Selection & Integration Q&A

When should I choose the CC-905 rather than a smaller Cc-series sibling such as the CC-505wl?

Choose the CC-905 when the process needs setpoints down to -90 °C or a 26 l bath; its cascade refrigeration is what reaches the deep-cold end. Siblings such as CC-505wl cover less extreme ranges and smaller duties, so pick them when -90 °C is not required.

What must the external circuit and heat-transfer fluid satisfy to work with the CC-905 pump?

The loop must fit within 0.7 bar pressure-side head (25 l/min max) and 0.4 bar suction-side (18.5 l/min), with M16x1 male connections. Fluid kinematic viscosity must stay at or below 50 mm²/s at operating temperature.

What site conditions and power quality does the CC-905 require for reliable operation?

Install at 5–40 °C ambient, IP20, with 400 V 3~N 50 Hz supply, 14 A max and 3x16 A fusing. Voltage/frequency tolerances are ±5 %/±2 % but must not combine unfavourably (-5 % V with +2 % Hz is not allowed).

What is not included in the base delivery that I must plan for?

Three-phase units below 63 A ship with cable but without a plug, so the site plug must be provided. Heat-transfer fluid, external Pt100 sensor and hoses are not part of the base unit and must be specified separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-905 Refrigerated/Heating Circulator
Datasheet ·69,2 КБ ·2 pp.
View PDF