Peter Huber Kältemaschinenbau CC-525w Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau CC-525w Kälte-Badumwälzthermostat — Thermostatic baths.
The CC-525w is a water-cooled refrigerated circulating bath with a 17 l bath, a variable-speed pressure/suction pump and a -55…200 °C working range, positioned above the smaller CC-405/410/415 and CC-505 siblings in the series. Its 3 kW heater and water-cooled condenser suit continuous external-circuit temperature control where a 400 V three-phase feed and a 3–6 bar cooling-water supply are available.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-525w Kälte-Badumwälzthermostat Current | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-410wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-510w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-515w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-520w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-905 Kälte-Badumwälzthermostat | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-905w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-906w Kälte-Badumwälzthermostat | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-K12 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-K15 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| CC-K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Bath volume | 17 l |
| Filling volume | 15 l (min.) |
| Heating capacity | 3 kW |
| Maximum delivery flow rate | 25 l/min (pressure pump max.; 0, 7 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -55...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 539 x 629 x 1102 mm |
| Delivery capacity suction pump max. mfr | 18, 5 l/min (suction pump max.; 0, 4 bar) |
| Bath opening WxD/BD mfr | 270 x 150 / 200 |
| Weight | 163 kg |
| Current consumption | 13 A (max.) |
| Power supply | 400V 3~N 50Hz |
| Fuse mfr | 3x16A |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 205 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 3 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Cooling capacity mfr | 100:7; 20:7; 0:5; -20:3; -40:1, 5 °C | kW |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled refrigerated bath with 17 l bath volume (15 l minimum fill; 10 l with displacement rack) and a 270 x 150 mm bath opening, 200 mm deep.
- Pump is a variable-speed pressure and suction type: up to 25 l/min at 0.7 bar delivery, 18.5 l/min at 0.4 bar suction, with M16x1 male pump connections.
- Within the Cc series, the CC-525w is the larger-bath, water-cooled modification; the CC-405w, CC-410wl, CC-415wl and CC-505wl are the smaller siblings, so choose this model when the 17 l bath and the higher cooling/heating envelope are needed rather than the compact variants.
- Control is via the Pilot ONE controller with Ethernet, USB (host and device) and RS232 interfaces and a Pt100 external sensor connection.
Working range and load behavior
- Temperature range -55…200 °C with ±0.02 K stability; heating capacity 3 kW.
- Cooling capacity is quoted at 100:7, 20:7, 0:5, -20:3 and -40:1.5 °C in kW, i.e. capacity falls steeply toward the low end of the range.
- Cooling is water-side dependent: the datasheet states cooling power drops as cooling-water temperature rises, so rated capacities assume the specified water conditions.
- Medium viscosity must not exceed 50 mm²/s; the pump is speed-adjustable to match external circuit resistance.
Installation, connection and operation
- Requires a 400 V 3~N 50 Hz supply, 13 A maximum current, 3x16 A fuse; three-phase units under 63 A are delivered with cable but without a plug.
- Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%, and certain combinations are not permitted.
- Cooling-water circuit: G1/2 male connection, 205 l/h consumption at 15 °C, minimum differential pressure 3 bar, maximum pressure 6 bar; intake temperature 5…20 °C, with pipe insulation required below the condensation point.
- Wetted cooling-water materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, which constrains water quality.
- Safety: adjustable overtemperature protection to DIN 12876, safety class III / FL; IP20 enclosure, 5…40 °C ambient.
- Refrigerant R-452A, 1.9 kg, GWP 2141, 4.1 t CO2 equivalent, subject to F-gas rules; EMC classification Class A, Group 1 to EN55011.
Integration & Operating Notes
The unit draws up to 13 A on a 400 V 3~N 50 Hz supply and requires a 3x16 A fuse; because three-phase devices under 63 A ship with cable but without a plug, the mating plug and breaker must be procured and the feed sized before commissioning.
The water circuit needs at least 3 bar differential pressure to reach rated cooling and must never exceed 6 bar; intake temperature must stay within 5…20 °C, and pipes below the condensation point need insulation to avoid condensation on the unit and circuit.
The variable-speed pressure/suction pump is rated for media up to 50 mm²/s kinematic viscosity; thicker fluids at low temperature raise circuit resistance beyond the 0.7 bar delivery / 0.4 bar suction envelope and must be avoided or pre-diluted.
A ±5% voltage deviation is only permissible with a simultaneous ±2% frequency deviation, and certain combinations are explicitly not allowed; site power quality should be checked against this before connecting the 3 kW heater and refrigeration circuit.
The R-452A charge of 1.9 kg corresponds to 4.1 t CO2 equivalent (GWP 2141), so periodic leak checks and documentation are required under F-gas rules; plan access to the refrigeration circuit and keep the charge record with the unit.
The Pilot ONE controller accepts an external Pt100 sensor and offers Ethernet, USB host/device and RS232, allowing the bath temperature to be controlled at the external load rather than in the bath; wiring and network provisioning must be planned with the installation.
Selection & Integration Q&A
Pick the CC-525w when the 17 l bath, 3 kW heating and the -55…200 °C water-cooled envelope are needed for larger external loads; the smaller siblings suit compact bath duties where the higher cooling capacity and bath volume are not required.
Pump ports are M16x1 male for the external loop (up to 25 l/min at 0.7 bar, 18.5 l/min suction at 0.4 bar); the cooling-water side uses a G1/2 male connection needing 3–6 bar differential pressure and 205 l/h at 15 °C.
Three-phase units under 63 A arrive with cable but no plug, so the mating plug is a separate item; the R-452A charge (4.1 t CO2e) brings F-gas leak-check obligations, and bath fluid plus displacement rack (reducing fill to 10 l) are user-supplied.