Peter Huber Kältemaschinenbau CC-K15 Kälte-Badthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-K15 Kälte-Badthermostat

Peter Huber Kältemaschinenbau CC-K15 Kälte-Badthermostat — Thermostatic baths.

The CC-K15 is a 15-litre refrigerated bath thermostat with a variable-speed pressure/suction pump and Pilot ONE controller, covering -20 to 200 °C at ±0.02 K stability. It is the smallest bath in the CC series, distinguished from the CC-4xx/5xx siblings by its 15 l volume and 0.25 kW cooling at 20 °C. It serves as a temperature-control source for external circuits or direct bath immersion in laboratory and pilot processes.

Direct drop-in replacements

Eppendorf
MixMate ®
71%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-K15 Kälte-Badthermostat Current —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Bath volume 15 l
Dimensions (w × D × H) 350 x 560 x 430 mm
Pump connection (Option, #19607) (Option)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range -20...200 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 1, 8 - 2, 1 kW
Delivery capacity pressure pump max. mfr 27 l/min; 0, 7 bar
Delivery capacity suction pump max. mfr 22 l/min; 0, 4 bar
Bath opening WxD/BD mfr 290 x 152 / 200
Filling volume min. mfr 11, 3 l
Dimensions, Weight & Mechanical Design
Weight 28 kg
Electrical, Power & Drive Specifications
Power supply 220-240V 1~/2~ 50/60Hz
Current consumption max. mfr 10 + 1, 5 A
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Cooling capacity mfr 20: 0, 25; 0: 0, 2; -10: 0, 12; -20: 0, 05 kW (at °C)
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The CC-K15 is a refrigerated bath thermostat built to DIN 12876 with a 15 l bath, a 290 x 152 mm opening (200 mm depth) and a minimum filling volume of 11.3 l. It uses an air-cooled refrigeration circuit with R-290 (A3, H220) natural refrigerant and a pressure/suction pump with variable speed. Within the CC series, the CC-K15 is the compact 15 l unit; the CC-405w, CC-410wl, CC-415wl and CC-505wl siblings are larger bath/circulator variants. The CC-K15 is the choice when the working volume and cooling demand fit 15 l and 0.25 kW at 20 °C, rather than the larger siblings.

Thermal behavior and load limits

Installation, control and operation

The unit requires a 220–240 V, 1~ or 2~, 50/60 Hz supply with a 10 A minimum and 16 A maximum fuse; current consumption is 10 + 1.5 A. Voltage and frequency tolerances are ±5 % and ±2 % but must not be combined in the unfavourable direction. A power cable with country-specific plug must be specified at order. The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, and an external Pt100 sensor connection. The unit is IP20 and Class A Group 1 for EMC, so it needs a dry indoor location; ambient temperature must stay between 5 and 40 °C, and cooling capacity drops as ambient rises. Overtemperature protection is adjustable per DIN 12876, safety class III / FL. Polyglycol must not be used as heat transfer fluid when the Huber pump adapter is fitted. A drain valve with cap (#6839) is available as an accessory.

Integration & Operating Notes

Electrical
Supply fuse must be between 10 A and 16 A

The unit draws up to 10 + 1.5 A and requires a branch protection of at least 10 A but not more than 16 A. A lower-rated breaker will trip on compressor start; a higher-rated one will not protect the internal wiring. Confirm the site breaker before commissioning.

Datasheet, Technical Data
Electrical
Do not combine -5 % voltage with +2 % frequency

Voltage tolerance is ±5 % and frequency tolerance ±2 %, but the unfavourable combination (-5 % voltage together with +2 % frequency) is explicitly not allowed. Check the site power quality against this rule before connecting the compressor and electronics.

Datasheet, Technical Data
Mounting
Air-cooled condenser needs clear space and cool room air

The refrigeration machine is air-cooled, so the condenser must have adequate air flow and the stated space clearances. Ambient temperature must stay between 5 and 40 °C; cooling capacity drops as ambient rises, so a hot plant room erodes the published 0.25 kW at 20 °C.

Datasheet, Technical Data
Process Limits
Size low-temperature loads against 0.05 kW at -20 °C

Cooling capacity falls from 0.25 kW at 20 °C to 0.05 kW at -20 °C. A load that is comfortable at ambient may be unable to reach or hold -20 °C. Calculate the actual heat load at the coldest setpoint, including pump heat and external circuit gains, before committing to the lower end of the range.

Datasheet, Technical Data
Process Limits
Polyglycol is not permitted with the Huber pump adapter

When the Huber pump adapter is fitted, polyglycol must not be used as heat transfer fluid. Select a compatible fluid for the wetted parts and the application temperature range; using polyglycol will damage the pump adapter and invalidate the intended service.

Datasheet, Technical Data
Maintenance
Order the drain valve with cap (#6839) for service

A drain valve with cap is available as an accessory and simplifies bath draining during fluid changes and maintenance. Without it, draining the 15 l bath (minimum filling 11.3 l) is less practical. Specify it at order to avoid later rework.

Datasheet, Technical Data
Electrical
Specify country-specific plug at order

Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be specified when ordering. Confirm the plug type and the 220–240 V 1~/2~ supply arrangement at the site before the order is placed.

Datasheet, Technical Data

Selection & Integration Q&A

When should I choose the CC-K15 rather than a larger CC-4xx or CC-5xx sibling?

Pick the CC-K15 when the bath volume fits 15 l (minimum filling 11.3 l) and the cooling demand is within 0.25 kW at 20 °C, falling to 0.05 kW at -20 °C. Choose a larger sibling when you need more bath volume or higher cooling capacity at low temperature.

What must I check on the external circuit and power supply before integrating the CC-K15?

The pressure/suction pump delivers up to 27 l/min at 0.7 bar and 22 l/min at 0.4 bar, so the external circuit pressure drop must fit within these limits. The supply must be 220–240 V 1~/2~ 50/60 Hz with a 10–16 A fuse, and the country-specific plug must be specified at order.

What maintenance accessories and consumables are not included in the base delivery?

The drain valve with cap (#6839) is an accessory and is not part of the base unit. The power cable with country-specific plug must be specified at order. Heat transfer fluid is not included and must be selected to suit the application; polyglycol is not permitted with the Huber pump adapter.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-K15 Immersion Circulator
Datasheet ·69,4 КБ ·2 pp.
View PDF