Peter Huber Kältemaschinenbau Unistat T320w HT Dynamisches — Temperature control units.
The Unistat T320w HT is a water-cooled, dynamically controlled high-temperature circulator with 12 kW heating and a magnetically coupled stainless-steel pump, covering 15...300 °C with water cooling or 65...300 °C without active refrigeration. It is the water-cooled HT variant of the T305/T320 series, distinguished by its 24 l expansion volume and G3/4 cooling-water circuit.
Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection
| Model | Display resolution | Power supply | Pump connection | Pump type | Temperature range, °C | Expansion volume, m³ |
|---|---|---|---|---|---|---|
| Unistat T320w HT Dynamisches Current | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Chili Dynamisches | 0.01 | 200-240V 1~/2~ 50/60Hz | M24x1,5 AG male | Pressure pump, variable | 65–300 | 0.002 |
| Unistat T305 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305 HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T305w HT Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 65–300 | — |
| Unistat T320 Dynamisches | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T330w HT Dynamisches | 0, 01 K | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 65–300 | 0.024 |
| Unistat T340 Dynamisches | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T340 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T345 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T350 HT Dynamisches | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 65–300 | 0.056 |
| Unistat T402 Dynamisches | 0, 01 | 400V 3~N 50/60Hz | M24x1, 5 AG male | Pressure pump, variable | 80–425 | — |
| Unistat TR401 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 50–400 | 0.005 |
| Unistat TR402 Dynamisches | 0, 01 | 200-240V 1~/2~ 50/60Hz | M24x1, 5 (Flow) M24x1, 5 | Pressure pump, variable | 80–425 | — |
Full specifications
| Control output | POKO ONE (potential-free contact) |
| Expansion volume | 24 l |
| Filling volume | 7 l (min) |
| Heating capacity | 12 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | 65...300 °C with water cooling: 15...300 °C |
| Weight | 179 kg (net) |
| Current consumption | 20 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The T320w HT is a high-temperature (HT) dynamic temperature control unit with no active refrigeration circuit; heat removal relies on the application's own heat loss or on the water-cooling circuit.
- Within the Unistat T305/T320 family it is the water-cooled HT modification: the T305 and T305 HT are the smaller siblings, the T305w HT the water-cooled smaller unit, and the Chili Dynamisches a different series.
- Heating capacity is 12 kW; the pump is magnetically coupled in stainless steel with adjustable speed, feeding M30x1.5 male connections with Y-shaped bore per DIN 3863.
- A 24 l expansion vessel (not thermoregulated, lockable for open baths) and a 7 l minimum filling volume define the fluid inventory.
Behaviour across the working range
- Temperature range is 65...300 °C in plain operation and 15...300 °C with water cooling; the unit cannot actively refrigerate, so the lower end depends on cooling water and application heat loss.
- Cooling water intake must be 5...20 °C at 1 bar minimum differential pressure and no more than 6 bar, through a G3/4 male connection.
- The stated 10 kW cooling power is valid only at 20 °C room temperature, 15 °C cooling water and 1 bar differential pressure.
- Pump delivery is 96 l/min at zero head and 20 l/min at 3.0 bar, so loop pressure drop must be checked against this curve.
- Viscosity limit is 50 mm²/s; display resolution is 0.01 K.
Control, installation and operation
- Control is via Pilot ONE with POKO ONE potential-free contact output; interfaces are Ethernet, USB host and device, and RS232; an external Pt100 sensor can be connected.
- Power is 400 V 3~ 50 Hz with a maximum current of 20 A per phase; three-phase units below 63 A are delivered with cable but without a plug.
- Safety class III / FL with adjustable overtemperature protection per DIN 12876; protection rating IP20 requires a dry indoor site at 5...40 °C ambient.
- The unit must be transported and stored upright and emptied before moving; casters are not for transport and carry only 25% of total mass symmetrically.
Integration & Operating Notes
The 400 V 3~ 50 Hz supply must be fused at 3x20 A and wired with a cable cross-section matching that load. Three-phase units below 63 A arrive with cable but no plug, so a suitable plug must be procured and the hard-wired connection made by an electrician against the rating plate.
Because the unit has no refrigeration circuit, reaching temperatures below 65 °C requires the water-cooling option (range then 15...300 °C). Without it, the lower limit depends entirely on heat loss from the application, so the process must be assessed for adequate heat rejection before selecting this model.
Cooling water must enter between 5 °C and 20 °C at a differential pressure of at least 1 bar and no more than 6 bar through a G3/4 male connection. Pipes must be insulated if the water can drop below the dew point, and only water compatible with the copper, 1.4401 stainless steel, MS, PA, PPE, PTFE and EPDM wetted materials may be used.
The unit must be transported and stored upright only and moved only when emptied, since horizontal transport damages the compressor and a filled unit risks fluid overflow. Casters carry only 25% of total mass symmetrically and are not suitable for transport; use an industrial truck, then engage parking brakes and leveling feet before commissioning.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open bath applications it must be locked to prevent fluid loss. The vessel is not thermoregulated, so this setting must be verified at commissioning for the actual loop type.
The cooling water supply should be shut off even for short downtimes such as overnight to avoid water damage and unnecessary load on the circuit. Hoses and hose connections must be checked periodically for leaks and quality and replaced as required.
When Process Safety mode is used to protect against exothermic reactions, the compressor must be configured to Permanently ON so cooling remains active when the overtemperature trigger occurs. Power Failure Auto can also be enabled to resume temperature control automatically after an outage if it was active.
Selection & Integration Q&A
Pick the T320w HT when you need the larger 24 l expansion volume and 12 kW heating with water cooling down to 15 °C. The T305/T305 HT are smaller siblings; choose the non-w variants only if no cooling-water circuit is available and a 65 °C floor is acceptable.
Provide 5...20 °C water at 1...6 bar differential through a G3/4 male connection, compatible with copper, 1.4401, MS, PA, PPE, PTFE and EPDM. Size the loop so pressure drop leaves adequate flow: the pump gives 96 l/min at zero head but only 20 l/min at 3.0 bar.
Install indoors at 5...40 °C ambient, IP20, never in an ATEX zone. Transport and store upright and empty, allow about 2 h acclimation after a cold/warm move, and shut off cooling water even overnight. Casters are not for transport.
Three-phase units below 63 A ship with cable but no plug, so source a suitable plug. Hoses and connections must be checked periodically for leaks and quality and replaced as needed; the expansion vessel lock state must be set for open or closed loops.