Peter Huber Kältemaschinenbau Unistat T320w HT Dynamisches

Peter Huber Kältemaschinenbau Unistat T320w HT Dynamisches — Temperature control units.

The Unistat T320w HT is a water-cooled, dynamically controlled high-temperature circulator with 12 kW heating and a magnetically coupled stainless-steel pump, covering 15...300 °C with water cooling or 65...300 °C without active refrigeration. It is the water-cooled HT variant of the T305/T320 series, distinguished by its 24 l expansion volume and G3/4 cooling-water circuit.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat T320w HT Dynamisches Current 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305 HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305w HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330 HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 Dynamisches 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T340 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T350 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T402 Dynamisches 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Unistat TR401 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005
Unistat TR402 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE (potential-free contact)
Expansion volume 24 l
Filling volume 7 l (min)
Heating capacity 12 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range 65...300 °C
with water cooling: 15...300 °C
Dimensions, Weight & Mechanical Design
Weight 179 kg (net)
Electrical, Power & Drive Specifications
Current consumption 20 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behaviour across the working range

Control, installation and operation

Integration & Operating Notes

Electrical
Three-phase supply must be sized for 20 A per phase

The 400 V 3~ 50 Hz supply must be fused at 3x20 A and wired with a cable cross-section matching that load. Three-phase units below 63 A arrive with cable but no plug, so a suitable plug must be procured and the hard-wired connection made by an electrician against the rating plate.

Datasheet, Specifications; Operation Manual, Commissioning
Process Limits
No active refrigeration: 65 °C floor without water cooling

Because the unit has no refrigeration circuit, reaching temperatures below 65 °C requires the water-cooling option (range then 15...300 °C). Without it, the lower limit depends entirely on heat loss from the application, so the process must be assessed for adequate heat rejection before selecting this model.

Datasheet, Specifications and selection note
Cooling & Media
Cooling water 5...20 °C, 1...6 bar differential

Cooling water must enter between 5 °C and 20 °C at a differential pressure of at least 1 bar and no more than 6 bar through a G3/4 male connection. Pipes must be insulated if the water can drop below the dew point, and only water compatible with the copper, 1.4401 stainless steel, MS, PA, PPE, PTFE and EPDM wetted materials may be used.

Datasheet, Specifications and cooling water notes
Mounting
Transport upright and empty; casters are not for moving

The unit must be transported and stored upright only and moved only when emptied, since horizontal transport damages the compressor and a filled unit risks fluid overflow. Casters carry only 25% of total mass symmetrically and are not suitable for transport; use an industrial truck, then engage parking brakes and leveling feet before commissioning.

Operation Manual, Transport and Installation
Process Limits
Expansion vessel lock state depends on open or closed loop

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open bath applications it must be locked to prevent fluid loss. The vessel is not thermoregulated, so this setting must be verified at commissioning for the actual loop type.

Operation Manual, Setup Mode; Datasheet, expansion tank note
Maintenance
Shut off cooling water during overnight downtime

The cooling water supply should be shut off even for short downtimes such as overnight to avoid water damage and unnecessary load on the circuit. Hoses and hose connections must be checked periodically for leaks and quality and replaced as required.

Operation Manual, Service and Maintenance
Control & I/O
Process Safety mode needs compressor set to Permanently ON

When Process Safety mode is used to protect against exothermic reactions, the compressor must be configured to Permanently ON so cooling remains active when the overtemperature trigger occurs. Power Failure Auto can also be enabled to resume temperature control automatically after an outage if it was active.

Operation Manual, Normal Operation

Selection & Integration Q&A

When should I choose the T320w HT instead of the T305 HT or T305w HT?

Pick the T320w HT when you need the larger 24 l expansion volume and 12 kW heating with water cooling down to 15 °C. The T305/T305 HT are smaller siblings; choose the non-w variants only if no cooling-water circuit is available and a 65 °C floor is acceptable.

What must the cooling-water supply and pump loop look like for this unit?

Provide 5...20 °C water at 1...6 bar differential through a G3/4 male connection, compatible with copper, 1.4401, MS, PA, PPE, PTFE and EPDM. Size the loop so pressure drop leaves adequate flow: the pump gives 96 l/min at zero head but only 20 l/min at 3.0 bar.

What site and duty-cycle limits apply from the manual?

Install indoors at 5...40 °C ambient, IP20, never in an ATEX zone. Transport and store upright and empty, allow about 2 h acclimation after a cold/warm move, and shut off cooling water even overnight. Casters are not for transport.

What is not included in base delivery and what maintenance is required?

Three-phase units below 63 A ship with cable but no plug, so source a suitable plug. Hoses and connections must be checked periodically for leaks and quality and replaced as needed; the expansion vessel lock state must be set for open or closed loops.