Peter Huber Kältemaschinenbau Unistat TR402 Dynamisches

Peter Huber Kältemaschinenbau Unistat TR402 Dynamisches — Temperature control units.

The Unistat TR402 Dynamisches is a hydraulically closed, high-temperature thermostat with a split HT-Circulator and Pilot ONE controller, covering 80–425 °C. Its variable-speed pressure pump delivers up to 31 l/min at 1 bar, distinguishing it within the Unistat series for dynamic, externally closed processes requiring precise temperature control without oil vapour release.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat TR402 Dynamisches Current 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305 HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305w HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T320w HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330 HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 Dynamisches 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T340 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T350 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T402 Dynamisches 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Unistat TR401 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE (potential-free contact)
Filling volume 3 l (min)
Pump connection M24x1, 5 (Flow) M24x1, 5 (Return)
Pump speed adjustable Yes
Pump type Pressure pump, variable speed
External sensor connection Pt100
Temperature range 80...425 °C
Dimensions, Weight & Mechanical Design
Weight 49 kg (net)
Electrical, Power & Drive Specifications
Current consumption 16 A (max)
Fuse rating 16 A
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Cooling water connection G1/2 male
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The TR402 is a split-architecture unit: the HT-Circulator (hydraulic section) is connected to the Pilot ONE controller by an approximately 5 m cable, allowing the controller to be placed away from the hot machine. Unlike the T305 variants in the same series, this model is dedicated to the high-temperature window of 80–425 °C and is hydraulically closed, so no oil vapour escapes into the laboratory atmosphere. The pressure pump is variable-speed, with M24x1.5 flow and return connections and a maximum delivery of 31 l/min at 1 bar delivery pressure. The cooling water connection is G1/2 male. The unit requires a minimum filling capacity of 3 l and has a 2.9 l expansion volume.

Behaviour under load and media

Heating power is voltage-dependent: 3 kW at 230 V, falling to 2.2 kW at 200 V, so heat-up time calculations must use the actual site voltage. The maximum current at 230 V is just below 16 A, and the unit can be limited to 13 A or 10 A at first start-up. The expansion vessel lid can be fixed and has connections for a nitrogen blanket limited to 0.2 bar maximum pressure. Additional heaters with over-temperature protection are fitted for the expansion vessel and internal pipework. The overtemperature protection threshold can only be set after entering a randomly generated code from the Pilot ONE, preventing accidental changes to the safety limit.

Installation, control and operation

The Pilot ONE controller provides Ethernet RJ45 10/100 Mbps, USB 2.0 host and device, and RS232 interfaces. A potential-free POKO ONE contact is available for alarm signalling. The unit must be transported and stored upright only; horizontal transport damages the compressor. Casters, if present, are not suitable for transport and are loaded symmetrically with 25% of total mass. Leveling feet must be activated and parking brakes engaged before commissioning. Allow about 2 hours acclimation when moving between cold and warm environments before power-on. For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open applications the expansion vessel must be locked to prevent fluid loss. The cooling water supply should be shut off even for short downtimes such as overnight.

Integration & Operating Notes

Electrical
External branch circuit protection to UL 489 is mandatory

The panel builder must provide a UL 489 listed breaker in the supply circuit. The unit's own 16 A fuse does not replace this external protection. Size the breaker for the maximum current of just under 16 A at 230 V, or for the limited 13 A or 10 A setting if configured at first start-up.

Unistat TR402 Temperature Control System Datasheet, page 2
Process Limits
Nitrogen blanket pressure must not exceed 0.2 bar

The expansion tank lid has connections for a nitrogen blanket with a maximum pressure of 0.2 bar. Exceeding this limit risks overpressure damage to the expansion vessel. Use a regulated nitrogen supply and verify the pressure before opening the blanket valve.

Unistat TR402 Temperature Control System Datasheet, page 1
Mounting
Unit must be transported and stored upright only

Horizontal transport damages the compressor. Plan rigging and transport routes to keep the unit upright at all times. Casters, if present, are not suitable for transport and carry only 25% of total mass symmetrically; use an industrial truck or crane. Only an emptied unit may be moved.

Unistat Operation Manual, page 22
Control & I/O
Process Safety mode requires compressor set to Permanently ON

If Process Safety mode is used to protect against exothermic reactions, the compressor must be set to Permanently ON via [System Settings] > [Power/ECO Settings] > [Compressor ON/OFF/AUTO] > [Permanently ON]. Otherwise cooling may not be active at the overtemperature trigger value, defeating the safety function.

Unistat Operation Manual, page 19
Control & I/O
Power failure response must be selected for the process risk

The Pilot ONE allows the power interruption response to be set to OFF/Standby (default), ON/Temperature control active, or Power Failure Auto. Power Failure Auto resumes temperature control automatically after a power outage if it was active. Choose the setting that matches the process risk; unattended operation may require Power Failure Auto or an alarm strategy.

Unistat Operation Manual, page 20
Maintenance
Shut off cooling water even for overnight downtimes

Shut off the cooling water supply to the temperature control unit even for shorter downtimes such as overnight. This prevents water damage and unnecessary load on the cooling water circuit. Also periodically check hoses and connections for leaks and quality and replace as needed.

Unistat Operation Manual, page 29
Environment
Do not install or start up inside an ATEX zone

The unit must not be installed or started inside an ATEX zone; explosion protection requires an Ex px cabinet. Verify area classification before installation. Additionally, use with medical devices or for direct foodstuff temperature control is not permissible.

Unistat Operation Manual, page 15

Selection & Integration Q&A

When should I choose the Unistat TR402 Dynamisches instead of a T305 variant in the same series?

Choose the TR402 for processes requiring the 80–425 °C window with a hydraulically closed circuit and split controller placement up to ~5 m away. The T305 variants cover different temperature ranges; select those when your process window falls outside 80–425 °C or when a different pump/connection configuration is needed.

What are the requirements for the cooling water and power supply adjacent to this unit?

Provide a G1/2 male cooling water connection and a 200–240 V 1~/2~ 50/60 Hz supply with a 16 A fuse and external UL 489 branch protection. The unit is delivered with a country-specific plug that must be specified at ordering; hard-wired connection must be done by an electrician.

What mounting and duty-cycle limits apply from the manual?

Transport and store upright only; use an industrial truck, not casters, for transport. Activate leveling feet and parking brakes before commissioning. Avoid permanent operation with very dynamic, high temperature changes recurring in short intervals, as this voids the guarantee. Ambient temperature must be 5–40 °C.

What maintenance is required and what accessories are not included in base delivery?

Periodically check hoses and connections for leaks and replace as needed; shut off cooling water overnight. The base delivery does not include a drip tray (required below the unit), a nitrogen blanket supply, or a UL 489 breaker. The country-specific plug must be specified at ordering.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat TR402 Temperature Control System Datasheet
Datasheet ·54,2 КБ ·2 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF
PDF
Unistat® TR 7 Temperature Control Unit — Operation Manual
Manual ·3,3 МБ ·92 pp.
View PDF