Peter Huber Kältemaschinenbau CC-415wl Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-415wl Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau CC-415wl Kälte-Badumwälzthermostat — Thermostatic baths.

The CC-415wl is a 5-litre refrigerated bath circulator with a variable-speed pressure/suction pump and Pilot ONE controller, covering -40 to 200 °C with ±0.02 K stability. Its water-cooled first stage (R-452A) and 1.3–1.6 kW heater distinguish it from air-cooled siblings in the Cc series. It serves as a precise temperature source for external circuits and bath immersion applications.

Direct drop-in replacements

Eppendorf
MixMate ®
45%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-415wl Kälte-Badumwälzthermostat Current —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Bath volume 5 l
Filling volume 4 l (min)
Maximum delivery flow rate 25 l/min (pressure pump, 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
Refrigerator air- and water-cooled, CFC-free
External sensor connection Pt100
Temperature range -40...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 410 x 480 x 764 mm
Heating capacity mfr 1, 3 - 1, 6 kW
Delivery capacity suction pump max. mfr 18, 5 l/min (suction pump, 0, 4 bar)
Bath opening WxD/BD mfr 120 x 110 / 150 mm
Dimensions, Weight & Mechanical Design
Weight 61 kg
Electrical, Power & Drive Specifications
Current consumption 9 A (max)
Power supply 220-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-452A (A1, H280) (1st stage)
Maximum viscosity 50 mm²/s
Refrigerant quantity 1st stage mfr 0, 38 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 66 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 3 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Cooling capacity mfr 100:1, 2; 20:1, 2; 0:1; -20:0, 6; -30:0, 2; -40:
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior

Installation and operation

Integration & Operating Notes

Cooling & Media
Verify 3 bar differential pressure before piping

The facility cooling water supply must deliver at least 3 bar differential pressure between inlet and outlet; insufficient pressure prevents proper cooling and may cause the unit to trip. Maximum permissible pressure is 6 bar, so a regulator may be needed.

CC-415wl Immersion Circulator Datasheet, page 1
Cooling & Media
Cooling water must be 5–20 °C and material-compatible

The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; aggressive water will corrode it. Water above 20 °C reduces cooling capacity and may require higher flow. Insulate internal pipes if water is below dew point.

CC-415wl Immersion Circulator Datasheet, page 2
Electrical
Do not combine -5% voltage with +2% frequency

Voltage tolerance is ±5% and frequency ±2%, but the combination of -5% voltage and +2% frequency is explicitly not allowed. Site power quality must be checked against this asymmetric rule to avoid damage or malfunction.

CC-415wl Immersion Circulator Datasheet, page 2
Electrical
Specify country-specific plug at ordering

Single- and two-phase devices are supplied with a power cable and country-specific plug; the plug type must be stated when ordering, otherwise commissioning may be delayed or field wiring required.

CC-415wl Immersion Circulator Datasheet, page 2
Mounting
IP20 requires dry indoor location

The unit has IP20 protection and no water ingress protection; it must be installed in a dry indoor area. Ambient temperature must be 5–40 °C; higher ambient reduces cooling capacity.

CC-415wl Immersion Circulator Datasheet, page 2
Process Limits
Pumped medium viscosity limited to 50 mm²/s

The maximum permissible kinematic viscosity of the circulated fluid is 50 mm²/s; higher viscosity overloads the pump and reduces circulation performance.

CC-415wl Immersion Circulator Datasheet, page 1
Compliance
Class A EMC may need extra measures in residential areas

The equipment is Class A, Group 1 to EN55011; it may cause interference in residential or sensitive environments, where additional EMC measures may be required.

CC-415wl Immersion Circulator Datasheet, page 2
Maintenance
Account for F-gas leak-check obligations

Refrigerant R-452A charge is 0.38 kg with GWP 2141, equal to 0.8 t CO2e; this determines leak-check intervals and reporting duties under F-gas regulations. Refrigerant is A1, H280 (pressurized gas hazard).

CC-415wl Immersion Circulator Datasheet, page 1

Selection & Integration Q&A

When should I choose the CC-415wl over other Cc-series models such as the CC-410wl or CC-508w?

Choose the CC-415wl when you need a 5 l bath with water-cooled condenser and -40 °C capability. Air-cooled siblings like the CC-508w avoid cooling-water plumbing but reject heat to the room; the CC-410wl offers a different bath size. Select based on bath volume, cooling-water availability, and low-temperature duty.

What are the requirements for the cooling water supply and pump circuit connections?

Cooling water: G1/2 male, 5–20 °C, 3 bar min differential, 6 bar max, 66 l/h at 15 °C. Pump: M16x1 male; pressure side 25 l/min at 0.7 bar, suction 18.5 l/min at 0.4 bar. Ensure water is compatible with copper, 1.4401, MS, PA, PPE, PTFE, EPDM.

What ambient and electrical conditions must the installation site meet?

Ambient 5–40 °C, dry indoor, IP20. Power 220–240 V 1~/2~ 50/60 Hz, max 9 A; voltage ±5%, frequency ±2%, but -5% voltage with +2% frequency is not allowed. Class A EMC may require measures in sensitive areas.

What maintenance and accessories are required beyond the base unit?

The base delivery includes power cable with country-specific plug (specify at order). Cooling-water hoses and mating fittings (G1/2, M16x1) are not included. F-gas leak checks apply (0.8 t CO2e). Insulate internal cooling-water pipes if water is below dew point.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-415wl Immersion Circulator Datasheet
Datasheet ·86,6 КБ ·2 pp.