Peter Huber Kältemaschinenbau CC-906w Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau CC-906w Kälte-Badumwälzthermostat — Thermostatic baths.
The CC-906w is a water-cooled refrigerated bath circulator with a 30 l bath and a -90…200 °C working range, using a two-stage cascade (R-507/R-23) and a variable-speed pressure/suction pump. It is the large-bath, deep-temperature modification of the CC series, distinguished by its 460 V 3~ 60 Hz supply and 3 kW heating capacity. It serves as a precision temperature source for external circuits and bath immersion duties requiring ±0.02 K stability.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-906w Kälte-Badumwälzthermostat Current | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-410wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-510w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-515w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-520w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-525w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-905 Kälte-Badumwälzthermostat | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-905w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-K12 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-K15 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| CC-K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Bath volume | 30 l |
| Filling volume | 25 l |
| Heating capacity | 3 kW |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -90...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 605 x 706 x 1216 mm |
| Delivery capacity pressure pump max. mfr | 25 l/min; 0, 7 bar |
| Delivery capacity suction pump max. mfr | 18, 5 l/min; 0, 4 bar |
| Bath opening WxD/BD mfr | 260 x 260 / 200 |
| Weight | 234 kg |
| Current consumption | 14 A (max.) |
| Power supply | 460V 3~ 60Hz |
| Fuse mfr | 3x16A |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 180 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 3 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The CC-906w is a floor-standing refrigerated bath circulator (605 x 706 x 1216 mm, 234 kg) with a 30 l bath, 25 l minimum filling, and a 260 x 260 mm opening 200 mm deep. Within the CC series, it is the largest and coldest variant: siblings such as the CC-405w, CC-410wl, CC-415wl and CC-505wl are smaller-bath models, while this unit combines a 30 l working volume with a -90 °C lower limit via a two-stage cascade refrigerant circuit (R-507 first stage, R-23 second stage). The Pilot ONE controller provides the setpoint and safety layer, with safety classification III / FL and adjustable overtemperature protection per DIN 12876.
Working behavior and media
- Temperature range -90…200 °C with ±0.02 K stability; heating capacity 3 kW.
- Cooling is water-side dependent: 180 l/h at 15 °C inlet, with cooling power dropping and flow demand rising as inlet water warms; inlet water must stay between 5 °C and 20 °C.
- Bath fluid must have a kinematic viscosity no higher than 50 mm²/s at the lowest operating temperature.
- The variable-speed pressure/suction pump delivers up to 25 l/min at 0.7 bar (pressure side) and 18.5 l/min at 0.4 bar (suction side) through M16x1 male connections.
- A displacement rack reduces usable bath volume to 19 l, which must be accounted for when sizing sample loads.
Installation, connection and control
The unit requires a 460 V 3~ 60 Hz supply, 14 A maximum current and 3x16 A fusing, with external UL 489 branch protection; three-phase devices below 63 A are delivered with a cable but no plug. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not permitted. Cooling water connects via G1/2 male fittings at 3 bar minimum differential pressure and 6 bar maximum; the wetted circuit includes copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so water quality must be compatible. Control and data interfaces are Ethernet, USB (host and device) and RS232, with a Pt100 external sensor connection. Ambient temperature must remain 5…40 °C, and IP20 protection means the unit must be kept away from water ingress.
Integration & Operating Notes
The water-cooled condenser requires a supply within this window; warmer water reduces cooling power and raises the required flow rate, so the site supply must be verified before commissioning. Pipes must be insulated where condensation can form.
The 460 V 3~ 60 Hz supply must be protected by a UL 489 branch circuit device sized for the 14 A maximum current and 3x16 A fusing; the unit is delivered with a cable but without a plug, so the customer must provide the final connection.
A ±5% voltage tolerance is permitted only with a simultaneous ±2% frequency tolerance; the combination of -5% voltage with +2% frequency is explicitly not allowed, so site power quality must be checked against both together.
The heat transfer fluid must be selected so that its kinematic viscosity at the lowest operating temperature does not exceed 50 mm²/s; otherwise pump delivery and temperature uniformity are compromised.
With a displacement rack installed, the bath capacity falls from 30 l to 19 l, so sample loads and filling volume must be planned against the reduced figure; minimum filling without the rack is 25 l.
The cascade circuit uses R-507 (4.5 t CO2 equivalent) and R-23 (12.4 t CO2 equivalent); these charge sizes trigger periodic leak-check obligations and service access planning for the refrigerant circuit.
The unit must be installed where ambient temperature stays between 5 °C and 40 °C and where water cannot enter the IP20 enclosure; the floor-standing 234 kg frame also requires a level, load-bearing surface.
Selection & Integration Q&A
Choose the CC-906w when you need a 30 l bath and the -90 °C lower limit from the two-stage cascade; the smaller siblings suit lower volume or less deep temperature duties. The 906w also brings a 3 kW heater and 460 V 3~ supply, so confirm site power before selecting it.
Provide cooling water at 5–20 °C inlet with 3–6 bar differential pressure and 180 l/h flow at 15 °C; connect via G1/2 male fittings. The wetted circuit includes copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so water quality must be compatible, and pipes need insulation where condensation can occur.
It requires 460 V 3~ 60 Hz, 14 A max, 3x16 A fusing and external UL 489 branch protection; a cable is supplied but no plug. Voltage tolerance ±5% is allowed only with ±2% frequency tolerance, and the -5% voltage/+2% frequency combination is prohibited. Ambient must stay 5–40 °C and IP20 requires a dry location.
The R-507/R-23 cascade carries 4.5 t and 12.4 t CO2 equivalent, so periodic refrigerant leak checks are required. The unit ships with a cable but no plug, and no external Pt100 sensor or bath fluid is included; the displacement rack is an accessory that reduces bath capacity to 19 l.