Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat T330 HT Dynamisches

Peter Huber Kältemaschinenbau Unistat T330 HT Dynamisches — Temperature control units.

The Unistat T330 HT Dynamisches is a high-temperature dynamic temperature control unit with a 65–300 °C working range, 24 kW heating capacity and a magnetically coupled stainless-steel pump delivering up to 96 l/min at 3.5 bar. It is the larger high-temperature dynamic modification in the Unistat T305/T330 series, distinguished by its 24 l expansion volume and 7 l minimum fill. It serves closed-loop or open-bath tempering of process equipment where rapid, controlled heat input above ambient is required.

Peter Huber Kältemaschinenbau Unistate Bis 425C Series — Model Comparison & Selection

Model Display resolutionPower supplyPump connectionPump typeTemperature range, °CExpansion volume, m³
Unistat T330 HT Dynamisches Current 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Chili Dynamisches 0.01200-240V 1~/2~ 50/60HzM24x1,5 AG malePressure pump, variable 65–3000.002
Unistat T305 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305 HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T305w HT Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable 65–300—
Unistat T320 Dynamisches 0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T320w HT Dynamisches 0, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T330w HT Dynamisches 0, 01 K400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st65–3000.024
Unistat T340 Dynamisches 0, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T340 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T345 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T350 HT Dynamisches 0, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st65–3000.056
Unistat T402 Dynamisches 0, 01400V 3~N 50/60HzM24x1, 5 AG malePressure pump, variable 80–425—
Unistat TR401 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 50–4000.005
Unistat TR402 Dynamisches 0, 01200-240V 1~/2~ 50/60HzM24x1, 5 (Flow) M24x1, 5Pressure pump, variable 80–425—

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Expansion volume 24 l
Filling volume 7 l (min)
Heating capacity 24 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range 65...300 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior and process window

Installation, control and operation

Integration & Operating Notes

Mounting
Transport and store upright only; drain before any move

Horizontal transport damages the compressor, and moving a filled unit risks thermal fluid overflow. Casters are not for transport and carry only 25% of total mass symmetrically; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning.

Unistat Operation Manual, Transport and Setup
Electrical
Leakage current above 3.5 mA affects RCD and earthing design

The unit's leakage current exceeds 3.5 mA, so protective earthing and residual-current device selection must be sized for this higher leakage. Hard-wired connection must be carried out by an electrician, and supply voltage/frequency must match the rating plate.

Unistat T330 HT Technical Data, Electrical Safety
Electrical
Voltage/frequency tolerance combination is restricted

Voltage tolerance is ±5% and frequency tolerance ±2%, but the combination of -5% voltage with +2% frequency is not allowed. Verify the site supply stays within the permitted combination before commissioning.

Unistat T330 HT Technical Data, Supply Tolerances
Process Limits
Expansion vessel lock state depends on open vs closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications it must be locked to prevent fluid loss. Set the lock state before filling and commissioning.

Unistat Operation Manual, Commissioning
Control & I/O
Process Safety mode requires compressor set to Permanently ON

When Process Safety mode is used to protect against exothermic reactions, the compressor must be configured to Permanently ON so that cooling remains active when the overtemperature trigger occurs. Configure this in the controller before enabling the mode.

Unistat Operation Manual, Normal Operation
Control & I/O
Power Failure Auto resumes tempering after outage

If temperature control is active during a power outage, Power Failure Auto resumes the process automatically when power returns. Assess whether unattended restart is acceptable for the process before enabling this function.

Unistat Operation Manual, Normal Operation
Maintenance
Shut off cooling water even for overnight downtime

The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit.

Unistat Operation Manual, Shutting Down
Maintenance
Periodically inspect hoses and connections for leaks and quality

Hoses and hose connections must be checked periodically for leaks and material quality and replaced as required. Include this in the preventive maintenance schedule for the fluid circuit.

Unistat Operation Manual, Service and Maintenance
Mounting
Do not install or start inside an ATEX zone

The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet. Site the unit outside hazardous areas and verify the area classification before installation.

Unistat Operation Manual, Safety
Mounting
Allow about 2 hours acclimation before first power-on

When moving the unit between cold and warm environments (or vice versa), allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults.

Unistat Operation Manual, Commissioning

Selection & Integration Q&A

When should I choose the T330 HT Dynamisches rather than a T305 or T305w HT sibling?

Choose the T330 HT when the process needs the larger fluid inventory (24 l expansion, 7 l minimum fill) and 24 kW heating within the 65–300 °C range. The T305/T305w HT variants are smaller high-temperature dynamic units; select them when loop volume and heat demand are lower.

What must I provide on the process side to connect the pump and ensure the loop works with this unit?

The pump port is M30x1.5 male with a DIN 3863 shape Y (60°) bore, so use ball-socket fittings to DIN 3863 and sleeve nuts to DIN 3870. Confirm the loop pressure drop fits the 96 l/min at 3.5 bar pump capability and keep fluid viscosity below 50 mm²/s.

What site and duty-cycle limits apply to installing and running this unit?

Install indoors, IP20, ambient 5–40 °C, outside any ATEX zone, upright and level with brakes/feet set. Supply is 400 V 3~ 50 Hz with ±5% voltage and ±2% frequency, but not -5% voltage with +2% frequency. Leakage current exceeds 3.5 mA.

What maintenance and delivery items should I plan for beyond the base unit?

The unit ships with a cable but no plug for three-phase supplies below 63 A, so the site must provide and fit the plug. Plan periodic hose and connection inspection and replacement, and shut off cooling water during downtimes. No accessories or consumables beyond the cable are included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat T330 HT Technical Data
Datasheet ·17,7 КБ ·2 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF