Peter Huber Kältemaschinenbau CC-905w Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-905w Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau CC-905w Kälte-Badumwälzthermostat — Thermostatic baths.

The CC-905w is a water-cooled refrigeration bath circulator with a 26 L bath and a variable-speed pressure/suction pump, covering -90 to 200 °C with ±0.02 K stability. It is the largest bath in the CC series (siblings hold 4–15 L), distinguished by its 3 kW heater and 400 V three-phase supply. It serves as a temperature-control source for external closed loops and bath immersion in process and laboratory systems.

Direct drop-in replacements

Eppendorf
MixMate ®
43%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-905w Kälte-Badumwälzthermostat Current —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Bath volume 26 l
Filling volume 19 l
Heating capacity 3 kW
Maximum delivery flow rate 25 l/min (pressure pump max.; 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range -90...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 605 x 706 x 1136 mm
Delivery capacity suction pump max. mfr 18, 5 l/min (suction pump max.; 0, 4 bar)
Bath opening WxD/BD mfr 260 x 260 / 200
Dimensions, Weight & Mechanical Design
Weight 175 kg
Electrical, Power & Drive Specifications
Current consumption 13 A (max.)
Power supply 400V 3~N 50Hz
Fuse mfr 3x16A
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 170 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 3 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Installation, connection and operation

Integration & Operating Notes

Electrical
Three-phase supply with 3×16 A fuse and no plug supplied

The unit draws up to 13 A from a 400 V 3~N 50 Hz supply and must be protected by a 3×16 A fuse. Because it is delivered with a cable but without a plug, a suitable three-phase plug and a correctly sized circuit must be provided on site before commissioning.

CC-905w Immersion Circulator Datasheet, Technical Specifications
Cooling & Media
Cooling water must stay within 5–20 °C and 3–6 bar differential

The water-cooled condenser requires an intake temperature between 5 °C and 20 °C and a differential pressure of at least 3 bar but not more than 6 bar. Insulate the cooling water pipes if the water temperature can fall below the dew point to prevent condensation.

CC-905w Immersion Circulator Datasheet, Technical Specifications
Process Limits
Cooling capacity collapses at low temperatures

Cooling power drops from 2 kW at 20 °C to only 0.13 kW at -90 °C. The process heat load must be evaluated at the actual operating temperature; a load acceptable at ambient will overrun the unit near the lower limit of the -90 to 200 °C range.

CC-905w Immersion Circulator Datasheet, Product Overview
Process Limits
Bath fluid viscosity must not exceed 50 mm²/s

The pump and heat exchanger are rated for a maximum kinematic viscosity of 50 mm²/s. Select a heat transfer fluid that stays within this limit across the entire operating temperature range, otherwise circulation and heat transfer will be inadequate.

CC-905w Immersion Circulator Datasheet, Technical Specifications
Electrical
Avoid the -5% voltage / +2% frequency combination

The supply tolerance is ±5% voltage and ±2% frequency, but the simultaneous combination of -5% voltage and +2% frequency is explicitly not allowed. Check the site power quality against this rule before connecting the unit.

CC-905w Immersion Circulator Datasheet, Technical Specifications
Maintenance
Refrigerant charge requires F-gas leak monitoring

The two refrigerant circuits contain R-452A and R-508B with CO2 equivalents of 1.9 t and 5.8 t respectively. This places the unit under refrigerant regulations, so leak checks and reporting obligations must be included in the maintenance schedule.

CC-905w Immersion Circulator Datasheet, Product Overview

Selection & Integration Q&A

When should I choose the CC-905w over the smaller CC-405w or CC-505wl siblings?

Choose the CC-905w when you need a 26 L bath and the full -90 to 200 °C range with 3 kW heating. The smaller siblings have 4–15 L baths and are preferable when bath volume and heat load are modest, avoiding the three-phase supply and water cooling.

What external circuit pressure drop can the pump overcome, and what cooling water supply must I provide?

The pressure pump delivers 25 L/min at 0.7 bar and the suction pump 18.5 L/min at 0.4 bar; size the external loop so its pressure loss stays within these heads. Cooling water must be 5–20 °C with 3–6 bar differential pressure and a G1/2 male connection.

What is supplied with the unit and what must be provided on site?

The unit is delivered with a power cable but without a plug; a 3×16 A fused 400 V 3~N 50 Hz supply and suitable plug must be provided. Bath fluid, external Pt100 sensor and cooling water fittings are not included and must be ordered separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-905w Immersion Circulator Datasheet
Datasheet ·59,9 КБ ·2 pp.