Peter Huber Kältemaschinenbau CC-K20 Kälte-Badthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-K20 Kälte-Badthermostat

Peter Huber Kältemaschinenbau CC-K20 Kälte-Badthermostat — Thermostatic baths with Noise level 50 dB.

The CC-K20 is a 17-litre refrigerated/heating bath circulator with a variable-speed pressure/suction pump, covering -30 to 200 °C with ±0.02 K stability. Its air-cooled R-290 first stage and Pilot ONE controller distinguish it from larger CC-405w/410wl/415wl/505wl siblings. It serves as a precise temperature source for external circuits or direct bath immersion in laboratory processes.

Noise level
50 dB

Direct drop-in replacements

Eppendorf
MixMate ®
53%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-K20 Kälte-Badthermostat Current 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Noise level 50 dB(A)
Bath volume 17 l
Dimensions (w × D × H) 350 x 555 x 615 mm
Filling volume 12 l (min.)
Maximum delivery flow rate 27 l/min (pressure pump max.; 0, 7 bar)
suction pump max.; 0, 4 bar: 22 l/min
Pump connection (Option, #19607) (Option)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
Refrigerator air-cooled, natural refrigerant
External sensor connection Pt100
Temperature range -30...200 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 1, 8 - 2, 1 kW
Bath opening WxD/BD mfr 290 x 329 / 150 (WxD/BD)
Dimensions, Weight & Mechanical Design
Weight 41 kg (net)
Electrical, Power & Drive Specifications
Power supply 220-240V 1~/2~ 50/60Hz
Current consumption max. mfr 10 + 2 A (max.)
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Refrigerant quantity 1st stage mfr 0, 072 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Cooling capacity mfr 20:0, 4; 0:0, 35; -10:0, 27; -20:0, 16; -30:0, 0
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour under load

Installation, control and operation

Integration & Operating Notes

Process Limits
Cooling capacity at -30 °C is only 0.065 kW

The refrigeration stage must be sized against the actual heat load at the coldest set point; at -30 °C the available cooling capacity is 0.065 kW, so excessive load or poor insulation will prevent the bath from reaching or holding the set point.

CC-K20 Datasheet, Specifications
Electrical
Do not combine -5 % voltage with +2 % frequency

The supply must remain within ±5 % voltage and ±2 % frequency, but the combination of -5 % voltage and +2 % frequency is explicitly not allowed; verify grid quality at the site before connection to avoid compressor and electronics damage.

CC-K20 Datasheet, Installation
Mounting
Air-cooled condenser needs unobstructed airflow

The unit is air-cooled, so sufficient clearance and cool intake air must be provided at the installation location; ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop as ambient temperature rises.

CC-K20 Datasheet, Installation
Process Limits
Polyglycol prohibited with Huber pump adapter

When a Huber pump adapter is used, polyglycol must not be used as the heat transfer fluid; selecting the wrong fluid can damage the pump and seals.

CC-K20 Datasheet, Operation
Electrical
Country-specific plug must be specified at ordering

Single- and two-phase devices are delivered with a power cable and country-specific plug; the plug type must be stated on the order to avoid a non-matching mains connection on site.

CC-K20 Datasheet, Installation
Maintenance
Drain valve accessory simplifies fluid changes

A drain valve with cap (#6839) is available as an accessory; planning its installation at commissioning avoids awkward fluid handling during later bath draining and fluid changes.

CC-K20 Datasheet, Maintenance

Selection & Integration Q&A

When should I choose the CC-K20 rather than a larger CC-405w/410wl/415wl/505wl sibling?

Choose the CC-K20 when a 17-litre bath with 290 x 329 mm opening and -30…200 °C range covers your sample size and heat load. The larger siblings are for greater bath volume or immersion area; verify cooling capacity at your coldest set point first.

What external circuit can the built-in pump supply, and what fluid restrictions apply?

The variable-speed pressure/suction pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min at 0.4 bar suction. With a Huber pump adapter, polyglycol must not be used as heat transfer fluid.

What site conditions and duty limits must be respected for installation?

Ambient 5–40 °C, IP20 enclosure away from water spray and conductive dust, air-cooled condenser with unobstructed airflow. Supply 220–240 V 1~/2~ 50/60 Hz within ±5 %/±2 %, but not -5 % voltage with +2 % frequency.

What maintenance accessories and delivery items are not included in the base unit?

The drain valve with cap (#6839) and the pump connection (#19607) are options/accessories. The power cable and country-specific plug are included but the plug type must be specified at ordering.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-K20 Refrigerated/Heating Circulator Datasheet
Datasheet ·75,0 КБ ·2 pp.
View PDF