Peter Huber Kältemaschinenbau CC-410wl Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-410wl Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau CC-410wl Kälte-Badumwälzthermostat — Thermostatic baths.

The CC-410wl is a 22-litre refrigerated bath circulator with a variable-speed pressure/suction pump and Pilot ONE controller, covering -45 to 200 °C. Its air/water-cooled natural-refrigerant (R-1270) stage and 2.7–3 kW heater distinguish it within the CC-405/415/505/508 series. It serves as a temperature-control source for external loops and bath immersion duty in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
47%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-410wl Kälte-Badumwälzthermostat Current —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Bath volume 22 l
Filling volume 17 l (min.)
Maximum delivery flow rate 25 l/min (pressure pump max.; 0, 7 bar)
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
Refrigerator air- water-cooled, natural refrigerant
External sensor connection Pt100
Temperature range -45...200 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 2, 7 - 3 kW
Delivery capacity suction pump max. mfr 18, 5 l/min (suction pump max.; 0, 4 bar)
Bath opening WxD/BD mfr 280 x 280 / 200 mm (WxD/BD)
Dimensions, Weight & Mechanical Design
Weight 71 kg (net)
Electrical, Power & Drive Specifications
Fuse rating 16 A
Power supply 220-240V 1~/2~ 50/60Hz V, Hz
Current consumption max. mfr 15, 5 A (max.)
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-1270 (A3, H220) (1st stage)
Maximum viscosity 50 mm²/s (max. permissible)
Refrigerant quantity 1st stage mfr 0, 25 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Cooling water connection G1/2 male
Cooling water differential pressure 3 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Cooling capacity mfr 100:0, 8; 20:0, 8; 0:0, 8; -20:0, 5; -30:0, 15;
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behaviour across the working range

Installation, connection and operation

Integration & Operating Notes

Cooling & Media
Cooling water needs 3–6 bar differential pressure at 5–20 °C

The cooling water circuit requires a minimum differential pressure of 3 bar and tolerates a maximum of 6 bar, with supply water between 5 and 20 °C. Verify site water pressure before installation; insufficient pressure degrades cooling, and pipes must be insulated if water temperature is below dew point to prevent condensation.

CC-410wl Immersion Circulator Datasheet, page 1–2
Electrical
Voltage and frequency tolerances are coupled

A ±5% voltage deviation is only permissible with ±2% frequency deviation; the combination of -5% voltage and +2% frequency is explicitly not allowed. Check supply quality and generator/UPS compatibility against this rule before commissioning.

CC-410wl Immersion Circulator Datasheet, page 2
Electrical
Plug supplied only for single/two-phase units

Single/two-phase devices are delivered with a country-specific plug, while three-phase units below 63 A come with a cable but no plug. The installer must provide the appropriate termination hardware for the latter.

CC-410wl Immersion Circulator Datasheet, page 2
Process Limits
Thermofluid viscosity must stay below 50 mm²/s

Maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s. This limits fluid choice at low temperatures, where viscosity rises; select a fluid that remains within this limit across the intended operating range.

CC-410wl Immersion Circulator Datasheet, page 1
Mounting
Flammable refrigerant R-1270 requires compliant location

The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant. The installation location must respect flammable-refrigerant rules, including room classification and ventilation planning.

CC-410wl Immersion Circulator Datasheet, page 1
Mounting
IP20 enclosure must be kept dry

The unit is rated IP20, so it must be installed away from splashes and washdown areas. Ambient temperature must be between 5 and 40 °C; room climate affects cooling performance and must be controlled.

CC-410wl Immersion Circulator Datasheet, page 2
Maintenance
Use only cooling water compatible with circuit materials

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Only compatible cooling water may be used to avoid corrosion and fouling; water quality and treatment should be checked as part of maintenance.

CC-410wl Immersion Circulator Datasheet, page 2

Selection & Integration Q&A

When should I choose the CC-410wl rather than the CC-405w or CC-415wl?

Choose the CC-410wl for a 22 l bath with -45 to 200 °C range and 0.8 kW cooling at 0 °C. The CC-405w is smaller; the CC-415wl is larger. Confirm bath volume and cooling load at your coldest setpoint before deciding.

What cooling water supply and piping must be provided for the CC-410wl?

Provide cooling water at 5–20 °C with 3–6 bar differential pressure at the G1/2 male connection. Insulate pipes if water is below dew point. The circuit materials (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM) dictate water compatibility.

What are the ambient and electrical limits for installing the CC-410wl?

Ambient must be 5–40 °C, IP20 requires a dry location, and EMC is Class A Group 1. Supply is 220–240 V 1~/2~ 50/60 Hz, 15.5 A max, 16 A fuse; ±5% voltage only with ±2% frequency, and -5% voltage with +2% frequency is not allowed.

What is included in delivery and what must be ordered separately?

Single/two-phase units include a country-specific plug; three-phase units below 63 A come with cable but no plug. External Pt100 sensor, thermofluid, and cooling-water fittings are not part of the base unit and must be specified separately.