Peter Huber Kältemaschinenbau CC-410wl Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau CC-410wl Kälte-Badumwälzthermostat — Thermostatic baths.
The CC-410wl is a 22-litre refrigerated bath circulator with a variable-speed pressure/suction pump and Pilot ONE controller, covering -45 to 200 °C. Its air/water-cooled natural-refrigerant (R-1270) stage and 2.7–3 kW heater distinguish it within the CC-405/415/505/508 series. It serves as a temperature-control source for external loops and bath immersion duty in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-410wl Kälte-Badumwälzthermostat Current | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-510w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-515w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-520w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-525w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-905 Kälte-Badumwälzthermostat | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-905w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-906w Kälte-Badumwälzthermostat | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-K12 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-K15 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| CC-K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Bath volume | 22 l |
| Filling volume | 17 l (min.) |
| Maximum delivery flow rate | 25 l/min (pressure pump max.; 0, 7 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| Refrigerator | air- water-cooled, natural refrigerant |
| External sensor connection | Pt100 |
| Temperature range | -45...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Heating capacity mfr | 2, 7 - 3 kW |
| Delivery capacity suction pump max. mfr | 18, 5 l/min (suction pump max.; 0, 4 bar) |
| Bath opening WxD/BD mfr | 280 x 280 / 200 mm (WxD/BD) |
| Weight | 71 kg (net) |
| Fuse rating | 16 A |
| Power supply | 220-240V 1~/2~ 50/60Hz V, Hz |
| Current consumption max. mfr | 15, 5 A (max.) |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-1270 (A3, H220) (1st stage) |
| Maximum viscosity | 50 mm²/s (max. permissible) |
| Refrigerant quantity 1st stage mfr | 0, 25 kg (1st stage) |
| Ambient temperature limit | 5...40 °C (permissible) |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 3 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Cooling capacity mfr | 100:0, 8; 20:0, 8; 0:0, 8; -20:0, 5; -30:0, 15; |
| Safety class | III / FL |
Description
Construction and series positioning
- The CC-410wl is a refrigerated bath circulator with a 22 l bath (minimum filling 17 l; 8.5 l with displacement rack) and a 280 x 280 mm opening, 200 mm deep.
- It sits between the smaller CC-405w and the larger CC-415wl in the series; the CC-505wl/CC-508w siblings differ in configuration and are not interchangeable references for this unit's ratings.
- Refrigeration is a single-stage air/water-cooled circuit using R-1270 (A3, H220), 0.25 kg charge; heating capacity is 2.7–3 kW.
- The pump is a variable-speed pressure and suction type: up to 25 l/min at 0.7 bar in pressure mode, up to 18.5 l/min at 0.4 bar in suction mode.
Behaviour across the working range
- Temperature range is -45 to 200 °C with ±0.02 K stability; the controller is Pilot ONE with an external Pt100 sensor connection.
- Cooling capacity is 0.8 kW at 100, 20 and 0 °C, falling to 0.5 kW at -20 °C, 0.15 kW at -30 °C and 0.1 kW at -40 °C — deep-cold heat loads must be sized against these low-end figures.
- Maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s, which constrains fluid selection at the cold end.
Installation, connection and operation
- Electrical supply is 220–240 V, 1~ or 2~, 50/60 Hz; maximum current consumption 15.5 A with a 16 A fuse. Voltage and frequency tolerances are coupled: ±5% voltage only with ±2% frequency, and -5% voltage with +2% frequency is not permitted.
- Cooling water connection is G1/2 male; the circuit requires at least 3 bar differential pressure and tolerates up to 6 bar, with water temperature 5–20 °C. Insulate pipes if water is below dew point.
- Pump connection is M16x1 AG male. Interfaces include Ethernet, USB host and device, and RS232.
- Ambient temperature must be 5–40 °C; protection rating is IP20, so the unit must be kept dry. EMC classification is Class A, Group 1 (EN55011), i.e. industrial-grade equipment.
- Single/two-phase units are delivered with a country-specific plug; three-phase units below 63 A come with cable but no plug.
Integration & Operating Notes
The cooling water circuit requires a minimum differential pressure of 3 bar and tolerates a maximum of 6 bar, with supply water between 5 and 20 °C. Verify site water pressure before installation; insufficient pressure degrades cooling, and pipes must be insulated if water temperature is below dew point to prevent condensation.
A ±5% voltage deviation is only permissible with ±2% frequency deviation; the combination of -5% voltage and +2% frequency is explicitly not allowed. Check supply quality and generator/UPS compatibility against this rule before commissioning.
Single/two-phase devices are delivered with a country-specific plug, while three-phase units below 63 A come with a cable but no plug. The installer must provide the appropriate termination hardware for the latter.
Maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s. This limits fluid choice at low temperatures, where viscosity rises; select a fluid that remains within this limit across the intended operating range.
The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant. The installation location must respect flammable-refrigerant rules, including room classification and ventilation planning.
The unit is rated IP20, so it must be installed away from splashes and washdown areas. Ambient temperature must be between 5 and 40 °C; room climate affects cooling performance and must be controlled.
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Only compatible cooling water may be used to avoid corrosion and fouling; water quality and treatment should be checked as part of maintenance.
Selection & Integration Q&A
Choose the CC-410wl for a 22 l bath with -45 to 200 °C range and 0.8 kW cooling at 0 °C. The CC-405w is smaller; the CC-415wl is larger. Confirm bath volume and cooling load at your coldest setpoint before deciding.
Provide cooling water at 5–20 °C with 3–6 bar differential pressure at the G1/2 male connection. Insulate pipes if water is below dew point. The circuit materials (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM) dictate water compatibility.
Ambient must be 5–40 °C, IP20 requires a dry location, and EMC is Class A Group 1. Supply is 220–240 V 1~/2~ 50/60 Hz, 15.5 A max, 16 A fuse; ±5% voltage only with ±2% frequency, and -5% voltage with +2% frequency is not allowed.
Single/two-phase units include a country-specific plug; three-phase units below 63 A come with cable but no plug. External Pt100 sensor, thermofluid, and cooling-water fittings are not part of the base unit and must be specified separately.