Peter Huber Kältemaschinenbau CC-510w Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-510w Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau CC-510w Kälte-Badumwälzthermostat — Thermostatic baths.

The CC-510w is a water-cooled refrigerated bath circulator with an 18 l bath, a variable-speed pressure/suction pump and the Pilot ONE controller, covering -50 to 200 °C at ±0.02 K stability. Its 3 kW heater and 2.4 kW cooling at 0 °C (0.4 kW at -40 °C) position it above the smaller CC-405/410/415/505 siblings for higher heat loads. It serves as a temperature-control source for external circuits and bath immersion applications requiring three-phase power and building cooling water.

Direct drop-in replacements

Eppendorf
MixMate ®
50%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-510w Kälte-Badumwälzthermostat Current —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Bath volume 18 l
Filling volume 16 l (min)
Heating capacity 3 kW
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range -50...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 455 x 515 x 1014 mm
Delivery capacity pressure pump max. mfr 25 l/min; 0, 7 bar
Delivery capacity suction pump max. mfr 18, 5 l/min; 0, 4 bar
Bath opening WxD/BD mfr 270 x 150 / 200 mm
Dimensions, Weight & Mechanical Design
Weight 106 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~N 50Hz
Current consumption max. mfr 10, 5 A (max)
Fuse mfr 3x16A
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water differential pressure 3 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Cooling capacity mfr 100:2, 4; 20:2, 4; 0:2, 4; -20:1; -40:0, 4 °C / (multiple temperature points)
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Cooling & Media
Cooling water must be 5–20 °C at 3–6 bar differential pressure

The water-cooled refrigeration circuit requires a differential pressure between 3 and 6 bar and an intake temperature between 5 °C and 20 °C. Verify these conditions in the building water system before commissioning; insulate pipework where condensation can form.

CC-510w Immersion Circulator Datasheet, pages 1–2
Process Limits
Medium viscosity above 50 mm²/s exceeds pump capability

The maximum permissible kinematic viscosity of the bath medium is 50 mm²/s. Selecting a fluid that is more viscous, particularly at low temperatures, will prevent proper circulation through the external circuit and bath.

CC-510w Immersion Circulator Datasheet, page 1
Process Limits
Cooling capacity at -40 °C is only 0.4 kW

Cooling power falls from 2.4 kW at 0 °C to 1 kW at -20 °C and 0.4 kW at -40 °C. The heat load of the application at the lowest operating temperature must be checked against this reduced capacity to avoid losing temperature control.

CC-510w Immersion Circulator Datasheet, page 1
Mounting
Install in 5–40 °C ambient, IP20, Class A environment

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature. It is classified Class A, Group 1 for electromagnetic disturbance to EN55011, so it is intended for non-domestic environments and may require EMC measures.

CC-510w Immersion Circulator Datasheet, page 2
Maintenance
Refrigerant R-452A charge of 1 kg carries F-gas obligations

The refrigerant R-452A has a GWP of 2141 and a charge of 1 kg, giving a CO2 equivalent of 2.1 t. This is relevant for leak checks and F-gas handling requirements at the installation site.

CC-510w Immersion Circulator Datasheet, page 1

Selection & Integration Q&A

When should I choose the CC-510w rather than the smaller CC-405w, CC-410wl, CC-415wl or CC-505wl?

Choose the CC-510w when the heat load at low temperature requires its 2.4 kW cooling at 0 °C and 0.4 kW at -40 °C, and when the 18 l bath and 3 kW heater suit the application. The smaller siblings offer less capacity; select them only if their lower cooling and bath volume meet the duty.

What must be provided on site for the electrical and cooling-water connections?

A 400 V 3~N 50 Hz supply fused at 3x16 A with a suitable plug (cable supplied, plug not included), and cooling water at 5–20 °C with 3–6 bar differential pressure on G1/2 male connections. Voltage tolerance is ±5% only with ±2% frequency tolerance.

What are the ambient and installation limits for the CC-510w?

Operate between 5 °C and 40 °C ambient at IP20. The unit is Class A, Group 1 (EN55011) for electromagnetic disturbance, so it is intended for non-domestic environments and may require EMC measures. Cooling water pipes must be insulated where condensation can occur.

What is not included in the base delivery that I need to plan for?

The three-phase unit is supplied with cable but without a plug, so a site plug and 3x16 A protection must be provided. External circuit connections use M16x1 male pump fittings and G1/2 male cooling-water fittings; an external Pt100 sensor is supported but not stated as included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-510w Immersion Circulator Datasheet
Datasheet ·60,9 КБ ·2 pp.
View PDF