The CC-905w is a water-cooled refrigeration bath circulator with a 26 L bath and a variable-speed pressure/suction pump, covering -90 to 200 °C with ±0.02 K stability. It is the largest bath in the CC series (siblings hold 4–15 L), distinguished by its 3 kW heater and 400 V three-phase supply. It serves as a temperature-control source for external closed loops and bath immersion in process and laboratory systems.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-905w Kälte-Badumwälzthermostat Current | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-410wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-510w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-515w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-520w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-525w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-905 Kälte-Badumwälzthermostat | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-906w Kälte-Badumwälzthermostat | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-K12 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-K15 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| CC-K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Bath volume | 26 l |
| Filling volume | 19 l |
| Heating capacity | 3 kW |
| Maximum delivery flow rate | 25 l/min (pressure pump max.; 0, 7 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -90...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 605 x 706 x 1136 mm |
| Delivery capacity suction pump max. mfr | 18, 5 l/min (suction pump max.; 0, 4 bar) |
| Bath opening WxD/BD mfr | 260 x 260 / 200 |
| Weight | 175 kg |
| Current consumption | 13 A (max.) |
| Power supply | 400V 3~N 50Hz |
| Fuse mfr | 3x16A |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 170 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 3 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The CC-905w is a refrigerated bath circulator with a 26 L bath volume and a minimum filling capacity of 19 L (15 L with a displacement rack).
- It is the largest-bath modification in the CC series; siblings CC-405w, CC-410wl, CC-415wl and CC-505wl are smaller-bath variants.
- The unit is water-cooled via a G1/2 male connection and uses a two-stage refrigerant circuit (R-452A and R-508B).
- Overall dimensions are 605 × 706 × 1136 mm at 175 kg net weight; the bath opening is 260 × 260 mm with 200 mm depth.
- Control is via the Pilot ONE controller with Ethernet, USB host/device and RS232 interfaces; an external Pt100 sensor can be connected.
Thermal behavior and media
- The working range is -90 to 200 °C with a temperature stability of ±0.02 K.
- Cooling capacity falls steeply at low temperatures: 2 kW at 20 °C down to 0.13 kW at -90 °C, so process heat loads must be checked against the actual setpoint.
- The bath fluid must have a kinematic viscosity no higher than 50 mm²/s to ensure circulation and heat transfer.
- Heating is provided by a 3 kW heater.
- The pressure pump delivers up to 25 L/min at 0.7 bar; the suction pump up to 18.5 L/min at 0.4 bar. Both are variable-speed.
Installation, connection and operation
- The unit requires a 400 V 3~N 50 Hz supply with a 3×16 A fuse and a maximum current consumption of 13 A. It is delivered with a cable but without a plug.
- Voltage tolerance is ±5% and frequency tolerance ±2%, but the combination of -5% voltage and +2% frequency is not permitted.
- Cooling water must be supplied at 5–20 °C with a differential pressure of at least 3 bar and no more than 6 bar; consumption is 170 L/h at 15 °C.
- Ambient operating temperature is 5–40 °C; protection rating is IP20, so a dry, ventilated location is required.
- The unit carries adjustable overtemperature protection per DIN 12876 and safety class III/FL.
- Refrigerant CO2 equivalents are 1.9 t (first stage) and 5.8 t (second stage), relevant for leak monitoring and F-gas reporting.
Integration & Operating Notes
The unit draws up to 13 A from a 400 V 3~N 50 Hz supply and must be protected by a 3×16 A fuse. Because it is delivered with a cable but without a plug, a suitable three-phase plug and a correctly sized circuit must be provided on site before commissioning.
The water-cooled condenser requires an intake temperature between 5 °C and 20 °C and a differential pressure of at least 3 bar but not more than 6 bar. Insulate the cooling water pipes if the water temperature can fall below the dew point to prevent condensation.
Cooling power drops from 2 kW at 20 °C to only 0.13 kW at -90 °C. The process heat load must be evaluated at the actual operating temperature; a load acceptable at ambient will overrun the unit near the lower limit of the -90 to 200 °C range.
The pump and heat exchanger are rated for a maximum kinematic viscosity of 50 mm²/s. Select a heat transfer fluid that stays within this limit across the entire operating temperature range, otherwise circulation and heat transfer will be inadequate.
The supply tolerance is ±5% voltage and ±2% frequency, but the simultaneous combination of -5% voltage and +2% frequency is explicitly not allowed. Check the site power quality against this rule before connecting the unit.
The two refrigerant circuits contain R-452A and R-508B with CO2 equivalents of 1.9 t and 5.8 t respectively. This places the unit under refrigerant regulations, so leak checks and reporting obligations must be included in the maintenance schedule.
Selection & Integration Q&A
Choose the CC-905w when you need a 26 L bath and the full -90 to 200 °C range with 3 kW heating. The smaller siblings have 4–15 L baths and are preferable when bath volume and heat load are modest, avoiding the three-phase supply and water cooling.
The pressure pump delivers 25 L/min at 0.7 bar and the suction pump 18.5 L/min at 0.4 bar; size the external loop so its pressure loss stays within these heads. Cooling water must be 5–20 °C with 3–6 bar differential pressure and a G1/2 male connection.
The unit is delivered with a power cable but without a plug; a 3×16 A fused 400 V 3~N 50 Hz supply and suitable plug must be provided. Bath fluid, external Pt100 sensor and cooling water fittings are not included and must be ordered separately.