Peter Huber Kältemaschinenbau CC-K12 Kälte-Badthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-K12 Kälte-Badthermostat

Peter Huber Kältemaschinenbau CC-K12 Kälte-Badthermostat — Thermostatic baths.

The CC-K12 is a refrigerated/heating bath circulator with a 10.5 L bath and a variable-speed pressure/suction pump, covering -20 to 200 °C with ±0.02 K stability. It is the compact 7 L-minimum-fill modification in the CC series, distinguished by its Pilot ONE controller and 1.8–2.1 kW heater. It serves as a temperature-control source for external circuits or direct bath immersion in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
63%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-K12 Kälte-Badthermostat Current —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Dimensions (w × D × H) 350 x 560 x 430 mm
Filling volume 7 l (minimum)
Maximum delivery flow rate 27 l/min (pressure pump; 0, 7 bar)
suction pump; 0, 4 bar: 22 l/min
Pump connection (Option, #19607) (Option)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range -20...200 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 1, 8 - 2, 1 kW
Bath opening WxD/BD mfr 290 x 152 / 150 mm (WxD/depth)
Bath volume mfr 10, 5 l
Dimensions, Weight & Mechanical Design
Weight 29 kg
Electrical, Power & Drive Specifications
Power supply 220-240V 1~/2~ 50/60Hz
Current consumption max. mfr 10 + 1, 5 A (max.)
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Cooling capacity mfr 20:0, 25; 0:0, 2; -10:0, 12; -20:0, 05 °C / kW
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Electrical
Supply must match 220–240 V 1~/2~ and be fused 10–16 A

The site circuit must provide 220–240 V single- or two-phase at 50/60 Hz and be protected by a fuse rated between 10 A and 16 A; maximum current consumption is 10 + 1.5 A, so undersized protection will trip and oversized protection will not protect the unit.

CC-K12 Circulating Bath Datasheet, page 2
Electrical
Voltage and frequency tolerances apply together

Voltage may deviate ±5% only with a simultaneous frequency deviation of ±2%, and some combinations are not allowed; supply quality outside these combined limits can cause malfunction or damage, so the site supply should be verified before commissioning.

CC-K12 Circulating Bath Datasheet, page 2
Mounting
Install indoors, dry, IP20 — no water ingress

The unit is rated IP20 and must be installed in a dry indoor location protected from water ingress; ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop above 20 °C room temperature.

CC-K12 Circulating Bath Datasheet, page 2
Process Limits
Low-temperature duty needs capacity sizing below 0 °C

Cooling capacity falls from 0.25 kW at 20 °C to 0.05 kW at -20 °C, so processes requiring sustained cooling near the lower limit must be sized against the reduced capacity to avoid inability to reach or hold setpoint.

CC-K12 Circulating Bath Datasheet, page 1
Cooling & Media
No polyglycol with the Huber pump adapter

When the Huber pump adapter is used, polyglycol must not be used as heat transfer fluid; selecting an incompatible fluid can damage the pump or compromise safety, so fluid choice must be confirmed against the adapter instructions.

CC-K12 Circulating Bath Datasheet, page 2
Electrical
Three-phase units ship without a plug

Three-phase devices with current consumption below 63 A are supplied with cable but without a plug, so the customer must provide a suitable plug and connection for commissioning.

CC-K12 Circulating Bath Datasheet, page 2
Compliance
Class A EMC may restrict residential use

The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN 55011, which has implications for residential EMC environments and may require additional measures if installed near sensitive equipment.

CC-K12 Circulating Bath Datasheet, page 2
Safety
Set adjustable overtemperature protection before unattended runs

The unit has adjustable overtemperature protection according to DIN 12876; it must be set for the application before unattended operation to prevent overheating if the control loop fails.

CC-K12 Circulating Bath Datasheet, page 1
Cooling & Media
Flammable R-290 refrigerant requires site risk assessment

The unit uses R-290 (A3, H220) flammable natural refrigerant, so the installation location must be assessed for flammable-refrigerant requirements, including ventilation and compliance approvals, before siting.

CC-K12 Circulating Bath Datasheet, page 1

Selection & Integration Q&A

When should I choose the CC-K12 over the CC-405w or CC-505wl siblings?

Choose the CC-K12 when a 10.5 L bath and 7 L minimum fill suffice and the -20 to 200 °C range with 0.05–0.25 kW cooling meets the load; the larger CC-405w/410wl/415wl/505wl siblings are for higher bath volumes or duty beyond this compact unit's capacity.

What supply and protection must the site provide for the CC-K12?

Provide 220–240 V, 1~ or 2~, 50/60 Hz with a 10–16 A fuse; maximum current is 10 + 1.5 A. Voltage ±5% and frequency ±2% apply together, and some combinations are not allowed. Three-phase units below 63 A ship without a plug.

What ambient and installation limits apply to the CC-K12?

Install indoors in a dry location: IP20, ambient 5–40 °C, with cooling capacity possibly reduced above 20 °C room temperature. The R-290 (A3) refrigerant requires a flammable-refrigerant site assessment, and Class A EMC may restrict residential use.

What maintenance and accessories are not included in the base delivery?

The base unit does not include a plug for three-phase versions below 63 A; the customer must provide it. The Huber pump adapter is an option (#19607), and polyglycol must not be used with it. Overtemperature protection must be set per DIN 12876 before unattended operation.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-K12 Circulating Bath Datasheet
Datasheet ·68,2 КБ ·2 pp.
View PDF