Peter Huber Kältemaschinenbau CC-K25 Kälte-Badthermostat
Peter Huber Kältemaschinenbau CC-K25 Kälte-Badthermostat — Thermostatic baths with Noise level 50 dB.
The CC-K25 is a refrigerated/heating circulator with a 23.5 L bath and a variable-speed pressure/suction pump, covering -30 to 200 °C with ±0.02 K stability. It is the smallest bath-volume model in the CC series, distinguished by its compact 350 mm width and natural R-290 refrigerant. It serves as a temperature-control source for external circuits or direct immersion in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-K25 Kälte-Badthermostat Current | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| CC-410wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-510w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-515w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-520w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-525w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-905 Kälte-Badumwälzthermostat | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-905w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-906w Kälte-Badumwälzthermostat | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-K12 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-K15 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Noise level | 50 dB(A) |
| Filling volume | 17 l (min.) |
| Maximum delivery flow rate | 27 l/min (pressure pump max.; 0, 7 bar) suction pump max.; 0, 4 bar: 22 l/min |
| Pump connection | (Option, #19607) (Option) |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| Refrigerator | air-cooled, natural refrigerant |
| External sensor connection | Pt100 |
| Temperature range | -30...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 350 x 555 x 615 mm |
| Heating capacity mfr | 1, 8 - 2, 1 kW |
| Bath opening WxD/BD mfr | 290 x 329 / 200 |
| Bath volume mfr | 23, 5 l |
| Weight | 41 kg |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 10 + 2 A (max.) |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Refrigerant quantity 1st stage mfr | 0, 072 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Cooling capacity mfr | 20 | 0, 4; 0 | 0, 35; -10 | 0, 27; -20 | 0, 16; |
| Safety class | III / FL |
Description
Construction and series positioning
The CC-K25 is a bath circulator with a 23.5 L working volume and a bath opening of 290 x 329 mm (200 mm depth). It is the most compact model in the CC series, with a 350 mm width that is narrower than the 405/410/415/505 siblings. The refrigeration circuit uses R-290 (propane, A3) with a charge of only 0.072 kg, and the unit is classified safety class III / FL per DIN 12876 with adjustable overtemperature protection. The controller is the Pilot ONE, and the pump is a variable-speed pressure/suction type.
Thermal performance and load behavior
Cooling capacity is strongly temperature-dependent: 0.4 kW at 20 °C, 0.35 kW at 0 °C, 0.27 kW at -10 °C, 0.16 kW at -20 °C, and 0.065 kW at -30 °C. These figures are valid at a 20 °C room temperature; higher ambient temperatures reduce available cooling. Heating capacity is 1.8–2.1 kW. Temperature stability is ±0.02 K across the -30 to 200 °C range. The pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min at 0.4 bar suction, with adjustable speed for matching external circuit resistance.
Installation, control and operation
The unit requires a 220–240 V, 1~ or 2~, 50/60 Hz supply with a 10–16 A branch fuse. Voltage tolerance is ±5% and frequency tolerance ±2%, but the combination of -5% voltage with +2% frequency is not permitted. The enclosure is IP20, so the location must be dry and protected from water. Ambient temperature must be 5–40 °C. The Pilot ONE controller provides Ethernet, USB (host and device), and RS232 interfaces, plus a Pt100 connection for an external sensor. A drain valve with cap (#6839) is an accessory for emptying the bath. When a Huber pump adapter is used, polyglycol must not be used as the heat transfer fluid. The unit is Class A, Group 1 per EN55011 and is delivered with a country-specific plug that must be specified at order.
Integration & Operating Notes
The unit accepts ±5% voltage and ±2% frequency deviation individually, but the specific combination of -5% voltage and +2% frequency is explicitly not allowed. Verify the actual supply quality at the installation point before commissioning, especially on weak laboratory grids or generator-backed supplies.
With IP20 protection, the unit must not be exposed to splashing water, condensation, or washdown. Site it in a dry laboratory area, and maintain the specified 5–40 °C ambient range; cooling capacity is rated at 20 °C room temperature and drops as ambient rises.
If a Huber pump adapter is fitted, polyglycol must not be used as the heat transfer fluid. Select an alternative fluid compatible with the adapter materials and the application temperature range before filling the 23.5 L bath.
The bath has adjustable overtemperature protection per DIN 12876. Set the cut-out below the fluid's maximum permissible temperature and above the intended setpoint to prevent overheating and fire risk, particularly with flammable fluids under safety class III / FL.
The drain valve with cap is an accessory, not part of the base delivery. Without it, emptying the 23.5 L bath for fluid changes or service requires alternative handling. Specify it at order if routine fluid replacement is planned.
Single- and two-phase devices are delivered with a power cable and a country-specific plug. State the required plug type on the order; otherwise the unit may arrive with a plug incompatible with the site's receptacles.
Selection & Integration Q&A
Choose the CC-K25 when a 23.5 L bath and 350 mm width fit the bench and the cooling load stays within its capacity curve (0.4 kW at 20 °C down to 0.065 kW at -30 °C). The 405/410/415/505 siblings offer larger bath volumes and likely higher capacities for greater loads or larger immersion areas.
It needs 220–240 V, 1~ or 2~, 50/60 Hz with a 10–16 A branch fuse. Maximum current draw is 10 + 2 A. Verify that the combination of -5% voltage and +2% frequency does not occur on the supply, and specify the country-specific plug at order.
No. The IP20 enclosure requires a dry location protected from water. Ambient temperature must remain between 5 and 40 °C. Cooling capacity is rated at 20 °C room temperature and may drop at higher ambient, so plan ventilation accordingly.
The drain valve with cap (#6839) is an accessory for emptying the 23.5 L bath and is not included in base delivery. A Huber pump adapter is available as an option (#19607); note that polyglycol must not be used with it. The pump connection itself is an option, so specify it if an external circuit is required.