Peter Huber Kältemaschinenbau CC-K25 Kälte-Badthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-K25 Kälte-Badthermostat

Peter Huber Kältemaschinenbau CC-K25 Kälte-Badthermostat — Thermostatic baths with Noise level 50 dB.

The CC-K25 is a refrigerated/heating circulator with a 23.5 L bath and a variable-speed pressure/suction pump, covering -30 to 200 °C with ±0.02 K stability. It is the smallest bath-volume model in the CC series, distinguished by its compact 350 mm width and natural R-290 refrigerant. It serves as a temperature-control source for external circuits or direct immersion in laboratory processes.

Noise level
50 dB

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-K25 Kälte-Badthermostat Current 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-906w Kälte-Badumwälzthermostat —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Noise level 50 dB(A)
Filling volume 17 l (min.)
Maximum delivery flow rate 27 l/min (pressure pump max.; 0, 7 bar)
suction pump max.; 0, 4 bar: 22 l/min
Pump connection (Option, #19607) (Option)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
Refrigerator air-cooled, natural refrigerant
External sensor connection Pt100
Temperature range -30...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 350 x 555 x 615 mm
Heating capacity mfr 1, 8 - 2, 1 kW
Bath opening WxD/BD mfr 290 x 329 / 200
Bath volume mfr 23, 5 l
Dimensions, Weight & Mechanical Design
Weight 41 kg
Electrical, Power & Drive Specifications
Power supply 220-240V 1~/2~ 50/60Hz
Current consumption max. mfr 10 + 2 A (max.)
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Refrigerant quantity 1st stage mfr 0, 072 kg (1st stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Cooling capacity mfr 20 | 0, 4; 0 | 0, 35; -10 | 0, 27; -20 | 0, 16;
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The CC-K25 is a bath circulator with a 23.5 L working volume and a bath opening of 290 x 329 mm (200 mm depth). It is the most compact model in the CC series, with a 350 mm width that is narrower than the 405/410/415/505 siblings. The refrigeration circuit uses R-290 (propane, A3) with a charge of only 0.072 kg, and the unit is classified safety class III / FL per DIN 12876 with adjustable overtemperature protection. The controller is the Pilot ONE, and the pump is a variable-speed pressure/suction type.

Thermal performance and load behavior

Cooling capacity is strongly temperature-dependent: 0.4 kW at 20 °C, 0.35 kW at 0 °C, 0.27 kW at -10 °C, 0.16 kW at -20 °C, and 0.065 kW at -30 °C. These figures are valid at a 20 °C room temperature; higher ambient temperatures reduce available cooling. Heating capacity is 1.8–2.1 kW. Temperature stability is ±0.02 K across the -30 to 200 °C range. The pump delivers up to 27 l/min at 0.7 bar pressure and 22 l/min at 0.4 bar suction, with adjustable speed for matching external circuit resistance.

Installation, control and operation

The unit requires a 220–240 V, 1~ or 2~, 50/60 Hz supply with a 10–16 A branch fuse. Voltage tolerance is ±5% and frequency tolerance ±2%, but the combination of -5% voltage with +2% frequency is not permitted. The enclosure is IP20, so the location must be dry and protected from water. Ambient temperature must be 5–40 °C. The Pilot ONE controller provides Ethernet, USB (host and device), and RS232 interfaces, plus a Pt100 connection for an external sensor. A drain valve with cap (#6839) is an accessory for emptying the bath. When a Huber pump adapter is used, polyglycol must not be used as the heat transfer fluid. The unit is Class A, Group 1 per EN55011 and is delivered with a country-specific plug that must be specified at order.

Integration & Operating Notes

Electrical
Supply tolerance combination -5% V with +2% Hz is prohibited

The unit accepts ±5% voltage and ±2% frequency deviation individually, but the specific combination of -5% voltage and +2% frequency is explicitly not allowed. Verify the actual supply quality at the installation point before commissioning, especially on weak laboratory grids or generator-backed supplies.

Datasheet, page 2, Power supply requirement
Mounting
IP20 enclosure requires a dry, water-protected location

With IP20 protection, the unit must not be exposed to splashing water, condensation, or washdown. Site it in a dry laboratory area, and maintain the specified 5–40 °C ambient range; cooling capacity is rated at 20 °C room temperature and drops as ambient rises.

Datasheet, page 2, Degree of Protection and ambient temperature
Process Limits
Polyglycol is incompatible with Huber pump adapters

If a Huber pump adapter is fitted, polyglycol must not be used as the heat transfer fluid. Select an alternative fluid compatible with the adapter materials and the application temperature range before filling the 23.5 L bath.

Datasheet, page 2, heat transfer fluid note
Safety
Overtemperature cut-out must be set for the actual fluid

The bath has adjustable overtemperature protection per DIN 12876. Set the cut-out below the fluid's maximum permissible temperature and above the intended setpoint to prevent overheating and fire risk, particularly with flammable fluids under safety class III / FL.

Datasheet, page 1, safety classification
Maintenance
Order drain valve #6839 for safe bath emptying

The drain valve with cap is an accessory, not part of the base delivery. Without it, emptying the 23.5 L bath for fluid changes or service requires alternative handling. Specify it at order if routine fluid replacement is planned.

Datasheet, page 2, Drain valve with cap #6839
Electrical
Country-specific plug must be specified at order

Single- and two-phase devices are delivered with a power cable and a country-specific plug. State the required plug type on the order; otherwise the unit may arrive with a plug incompatible with the site's receptacles.

Datasheet, page 2, power cable note

Selection & Integration Q&A

When should I choose the CC-K25 over the larger CC-405w or CC-505wl models?

Choose the CC-K25 when a 23.5 L bath and 350 mm width fit the bench and the cooling load stays within its capacity curve (0.4 kW at 20 °C down to 0.065 kW at -30 °C). The 405/410/415/505 siblings offer larger bath volumes and likely higher capacities for greater loads or larger immersion areas.

What supply and protection does the CC-K25 require, and can it run on a standard laboratory circuit?

It needs 220–240 V, 1~ or 2~, 50/60 Hz with a 10–16 A branch fuse. Maximum current draw is 10 + 2 A. Verify that the combination of -5% voltage and +2% frequency does not occur on the supply, and specify the country-specific plug at order.

Can the CC-K25 be installed in a wet or washdown area, and what ambient conditions apply?

No. The IP20 enclosure requires a dry location protected from water. Ambient temperature must remain between 5 and 40 °C. Cooling capacity is rated at 20 °C room temperature and may drop at higher ambient, so plan ventilation accordingly.

What accessories are needed for fluid changes and external circuit connection, and are they included?

The drain valve with cap (#6839) is an accessory for emptying the 23.5 L bath and is not included in base delivery. A Huber pump adapter is available as an option (#19607); note that polyglycol must not be used with it. The pump connection itself is an option, so specify it if an external circuit is required.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-K25 Refrigerated/Heating Circulator
Datasheet ·74,4 КБ ·2 pp.
View PDF