Peter Huber Kältemaschinenbau Unistat 815w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 815w Dynamisches

Peter Huber Kältemaschinenbau Unistat 815w Dynamisches — Temperature control units with Noise level 60 dB.

The Unistat 815w Dynamisches is a water-cooled, free-standing temperature control unit with a -85 to 250 °C range and 2 kW heating, positioned between the 705w and 825w siblings. Its variable-speed pressure pump and Pilot ONE controller with process-safety mode suit dynamic, exothermic reactor temperature control in closed circuits.

Noise level
60 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat 815w Dynamisches Current 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Noise level 60 dB(A)
Control output POKO ONE
Heating capacity 2 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type Pressure pump, variable speed
External sensor connection Pt100
Temperature range -85...250 °C
Dimensions, Weight & Mechanical Design
Weight 222 kg
Electrical, Power & Drive Specifications
Current consumption 13 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 130 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load

Control, installation and operation

Integration & Operating Notes

Cooling & Media
Cooling water must be 5–20 °C at 1–6 bar differential

The cooling water circuit requires an intake temperature between 5 and 20 °C, a minimum differential pressure of 1 bar and a maximum of 6 bar, with 130 l/h consumption at 15 °C. Pipes must be insulated if water can drop below the dew point, and the supply should be shut off even for short downtimes such as overnight.

Unistat 815w Technical Data, page 2; Unistat Operation Manual, page 29
Process Limits
Deep-cold cooling capacity falls to 0.2 kW at -80 °C

Cooling capacity is rated at 1.5 kW down to -20 °C but only 0.2 kW at -80 °C. Low-temperature duty must be checked against the actual heat load, and the external circuit pressure drop must be checked against the pump curve (40 l/min maximum, 5 l/min at 0.8 bar) to avoid starving flow and degrading temperature control.

Unistat 815w Technical Data, pages 1–2
Mounting
Transport upright and allow ~2 h acclimation before power-on

The unit must be transported and stored upright only; horizontal transport damages the compressor. Only an emptied unit may be moved. Allow about 2 hours acclimation when moving between cold and warm environments before switching on, to prevent condensation-induced electrical faults. Leveling feet must be activated and parking brakes engaged before commissioning.

Unistat Operation Manual, pages 22 and 29
Control & I/O
Process Safety mode requires the compressor set to Permanently ON

In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. Power Failure Auto can resume temperature control automatically after a power outage if it was active, which must be assessed for unattended operation.

Unistat Operation Manual, pages 19–20
Maintenance
Refrigerant work only by approved refrigeration contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. The cooling water supply should be shut off even for short downtimes. The unit is safety class III / FL with adjustable overtemperature protection per DIN 12876.

Unistat Operation Manual, pages 19 and 29; Unistat 815w Technical Data, page 1

Selection & Integration Q&A

When should I choose the Unistat 815w Dynamisches instead of the 705w or 825w siblings?

Choose the 815w when the duty needs the -85 to 250 °C range with 2 kW heating and the dynamic configuration, and the load fits its 1.5 kW at -20 °C / 0.2 kW at -80 °C cooling curve. The 705w is smaller; the 825w offers more capacity if the load exceeds the 815w curve.

What adjacent system parts and utilities must be provided for the 815w?

A 400 V 3~ 50 Hz supply with 3x16 A protection and a site-fitted plug (cable delivered without plug), cooling water at 5–20 °C with 1–6 bar differential and G1/2 male connections, and M30x1.5 male pump fittings for the external circuit. A Pt100 external sensor can be connected for process control.

What mounting and environment limits apply to the 815w?

Install dry, within 5–40 °C ambient, IP20, and never in an ATEX zone. Transport upright only and move only when emptied; allow ~2 h acclimation after a cold/warm move. Activate leveling feet and parking brakes before commissioning, and keep the expansion vessel unlocked for externally closed applications.

What maintenance and accessories are not included in base delivery?

The plug for the three-phase cable is not included. Refrigerant-circuit work requires approved refrigeration contractors. Hoses and connections need periodic leak/quality checks and replacement as needed; the cooling water supply should be shut off even overnight. A condensation/fluid drip tray must be installed by the responsible body.