Peter Huber Kältemaschinenbau Unistat 905w Dynamisches
Peter Huber Kältemaschinenbau Unistat 905w Dynamisches — Temperature control units.
The Unistat 905w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -90 to 250 °C. It is the mid-size 9xx modification in the Unistat w series, distinguished by 6 kW heating and a 34 A three-phase supply. It serves closed or open external circuits requiring deep-cold to high-temperature control.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat 905w Dynamisches Current | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -90...250 °C |
| Weight | 388 kg |
| Current consumption | 34 A (max) |
| Power supply | 208V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled refrigeration machine (suffix 'w') with R-452A/R-23 cascade, total CO2 equivalent ~14.1 t; requires on-site cooling water at 5–20 °C inlet, ≥1 bar differential pressure, 580 l/h at 15 °C.
- Free-standing 9xx chassis, 388 kg, IP20, ambient 5–40 °C; leveling feet/casters for positioning only, not transport.
- Pump: magnetically coupled stainless steel, speed-adjustable, M30x1.5 male process connection; 48 l/min at zero head, 17 l/min at 0.8 bar.
- Controller Pilot ONE with POKO ONE control output; Ethernet, USB host/device, RS232; external Pt100 sensor input; display resolution 0.01 K.
- Safety class III/FL; OT protection and Process Safety mode available.
Behavior under load and media
- Temperature range -90…250 °C; heating capacity 6 kW. Cooling power falls from 4.5 kW at -20 °C to 0.7 kW at -80 °C, so deep-cold duty must be sized against the curve, not the nominal.
- Heat transfer fluid limited to max kinematic viscosity 50 mm²/s; FL-rated fluids need fire point ≥65 °C, and ethanol only under supervision.
- Minimum filling capacity 3.2 l, expansion tank 5.3 l; external circuit volume and pressure drop must be checked against the pump curve.
- Cooling water circuit materials include copper, 1.4401 stainless, MS, PA, PPE, PTFE, EPDM; water must be compatible and pipes insulated below dew point.
Installation, connection and control
- Electrical: 208 V 3~ 60 Hz, max 34 A, 3x40 A fusing; external branch protection per UL 489; cable supplied without plug; leakage current >3.5 mA requires RCD selection review; voltage ±5% and frequency ±2% not to be combined in opposite directions.
- Cooling water G1/2 male; shut off supply even for overnight downtime.
- Expansion vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications.
- Power-failure behavior selectable: Off/Standby (default) or automatic restart; Process Safety mode requires compressor set to Permanently ON.
- Allow ~2 h acclimation after cold/warm move before power-on; transport upright and emptied only.
Integration & Operating Notes
The water-cooled refrigeration stage is designed for inlet water between 5 and 20 °C with at least 1 bar differential pressure and 580 l/h flow at 15 °C. Colder water risks condensation and requires insulated pipes; warmer water reduces capacity. Verify site supply before installation.
The unit draws up to 34 A at 208 V 3~ 60 Hz and is delivered with cable but no plug. External branch circuit protection according to UL 489 and 3x40 A fusing must be provided. Leakage current exceeds 3.5 mA, so standard 30 mA RCDs may nuisance-trip; select protective devices accordingly.
Minimum filling capacity is 3.2 l and the expansion tank holds 5.3 l. The pump delivers 48 l/min at zero head but only 17 l/min at 0.8 bar, so external circuit pressure drop must be checked. Fluid viscosity must not exceed 50 mm²/s at the operating temperature.
Horizontal transport damages the compressor, and a filled unit may overflow. Casters are not for transport and carry only 25% of mass; use an industrial truck. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to avoid condensation faults.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications it must be locked to prevent fluid loss. This is a mechanical interlock that must be verified before commissioning.
Pilot ONE can be set to Off/Standby (default, manual restart) or automatic restart after power interruption. Process Safety mode releases maximum cooling at the OT trigger and requires the compressor set to Permanently ON. Select according to whether unattended restart is acceptable for the process.
The cooling water supply should be shut off during short downtimes to avoid water damage and unnecessary load on the circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. Refrigerant-circuit work requires approved refrigeration contractors.
Selection & Integration Q&A
Pick the 905w when 6 kW heating and -90…250 °C with 34 A at 208 V 3~ match your load; the 815w is smaller, the 1005w larger. All are water-cooled 'w' variants; confirm cooling water 580 l/h at 15 °C and pump curve against your circuit.
208 V 3~ 60 Hz, max 34 A, 3x40 A fusing and UL 489 branch protection; cable supplied without plug. Cooling water G1/2 male, 5–20 °C inlet, ≥1 bar differential, 580 l/h at 15 °C. Leakage current >3.5 mA affects RCD choice.
Install upright and emptied, ambient 5–40 °C, IP20 (no wet/dusty areas), not in ATEX zones. Leveling feet or parking brakes engaged before start. Allow ~2 h acclimation after cold/warm move. Casters are not for transport.
No plug, no drip tray, no cooling-water hoses or heat transfer fluid. Periodic hose/connection checks and fluid replacement are required; cooling water shutoff for overnight stops. Refrigerant work by approved contractors; gas detection sensor optional and installed by operator.