Peter Huber Kältemaschinenbau Unistat P912w Dynamisches
Peter Huber Kältemaschinenbau Unistat P912w Dynamisches — Temperature control units with Noise level 65 dB.
The Unistat P912w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -90 to 250 °C. Its 6 kW heating capacity and 460 V 3~ 60 Hz supply distinguish it within the Unistat w series. It serves precise temperature control of externally closed or open applications in industrial processes.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat P912w Dynamisches Current | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Noise level | 65 dB(A) |
| Control output | POKO ONE |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -90...250 °C |
| Weight | 325 kg |
| Current consumption | 20 A (max) |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 516 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and positioning
- Water-cooled refrigeration circuit using R-452A and R-23, CFC/HCFC-free, with a combined CO2 equivalent of about 20.6 t.
- Magnetically coupled stainless-steel pump with adjustable speed; pump connection M30x1.5 male, bore shape Y (60°) per DIN 3863.
- Pilot ONE controller with POKO ONE control output; display resolution 0.01 K; interfaces Ethernet, USB (Host and Device), RS232.
- IP20 protection, net weight 325 kg, dimensions 630x704x1565 mm.
- Within the Unistat w series, the P912w is a larger free-standing unit; siblings include the 705w, 815w, 1005w and 1015w, which differ in size and capacity.
Behavior under load and media
- Temperature range -90 to 250 °C; heating capacity 6 kW.
- Cooling water consumption 516 l/h at 15 °C water and 0 °C flow; minimum differential pressure 1 bar, maximum pressure 6 bar.
- Cooling capacity decreases and required cooling water flow may increase as cooling water temperature rises above the reference 15 °C.
- Maximum permissible kinematic viscosity of heat transfer fluid is 50 mm²/s, limiting fluid choice at low temperatures.
- Safety class III / FL; for FL-rated flammable fluids, fire point must be ≥65 °C; ethanol only under supervision.
- Process Safety mode releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles; requires compressor set to Permanently ON.
Installation, connection and operation
- Requires 460 V 3~ 60 Hz supply, max current 20 A, fusing 3x25A; voltage tolerance ±5% with simultaneous frequency tolerance ±2%, but -5% voltage with +2% frequency is not allowed.
- Leakage current >3.5 mA; external branch circuit protection per UL 489 required. Three-phase devices under 63 A are supplied with cable but without plug.
- Cooling water connection G1/2 male; cooling water intake temperature 5–20 °C; insulate pipes if below condensation point.
- Ambient temperature 5–40 °C; sound pressure level 65 dB(A) ±4 dB(A).
- Minimum filling capacity 3.9 l; expansion tank filling capacity 6.5 l.
- For externally closed applications, expansion vessel must remain unlocked; for externally open applications, it must be locked.
- Allow about 2 hours acclimation when moving between cold and warm environments before power-on.
- Leveling feet must be activated and parking brakes engaged before commissioning; casters are not for transport.
- Cooling water supply should be shut off even for short downtimes (e.g., overnight).
- Periodic checks of hoses and connections for leaks and quality; replace as needed.
Integration & Operating Notes
The unit accepts ±5% voltage variation only with simultaneous ±2% frequency variation, and the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify site power quality before connection to avoid damage or warranty issues.
Leakage current exceeds 3.5 mA, so an external branch circuit protection device according to UL 489 must be installed. Three-phase devices under 63 A are supplied with cable but without plug; a suitable plug must be provided on site.
Cooling water differential pressure must be at least 1 bar and must not exceed 6 bar. Intake temperature must be between 5 °C and 20 °C; if water temperature can fall below the condensation point, insulate the cooling water pipes. Cooling capacity drops and flow demand may rise if water temperature exceeds 15 °C.
For externally closed applications, the expansion vessel must remain unlocked; locking it will cause overpressure damage. For externally open applications, it must be locked to prevent fluid loss. Verify the application type before commissioning.
Before commissioning, unscrew/activate the leveling feet and engage the parking brakes on casters (if present). Casters are not suitable for transport; use an industrial truck for moving the unit. Transport only in upright position and only when emptied.
Shut off the cooling water supply even for shorter downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Periodically check hoses and connections for leaks and quality, and replace as needed.
When Process Safety mode is enabled, the compressor must be set to Permanently ON so that maximum cooling capacity is released at the overtemperature trigger instead of switching off all poles. This mode is intended for exothermic reactions or critical processes.
Selection & Integration Q&A
Choose the P912w when its 6 kW heating capacity and -90 to 250 °C range fit your process, and when the 460 V 3~ 60 Hz supply is available. The 815w has lower noise (60 dB(A)) but likely less capacity; the 1005w and 1015w offer larger volumes. Select based on required heating/cooling power and space.
Cooling water connection is G1/2 male; supply must provide 1–6 bar differential pressure and 5–20 °C intake. Pump connection is M30x1.5 male with bore shape Y (60°) per DIN 3863; use ball socket per DIN 3863 and sleeve nut per DIN 3870 for pipework or hoses. Ensure heat transfer fluid viscosity ≤50 mm²/s.
Periodically check hoses and connections for leaks and quality; replace as needed. Shut off cooling water during downtimes. The unit is supplied with cable but without a plug for three-phase power; a suitable plug must be provided on site. External branch circuit protection per UL 489 is required and not included.