Peter Huber Kältemaschinenbau Unistat P815 Dynamisches

Peter Huber Kältemaschinenbau Unistat P815 Dynamisches — Temperature control units with Noise level 64 dB.

The Unistat P815 Dynamisches is a free-standing, air-cooled refrigerated heating circulator with a magnetically coupled stainless-steel high-pressure pump, covering -85 to 250 °C on a 400 V 3~ 50 Hz supply. Its 'P' pump variant targets high-pressure-drop loops such as flow-through chemistry and semiconductor applications, distinguishing it from the standard-pump 815w sibling. It serves as the temperature-control source for externally closed or open circuits requiring deep-cooling capability.

Noise level
64 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat P815 Dynamisches Current 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Noise level 64 dB(A)
Control output POKO ONE
Heating capacity 2 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -85...250 °C
Dimensions, Weight & Mechanical Design
Weight 241 kg
Electrical, Power & Drive Specifications
Current consumption 12 A (max)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load

Installation, control and operation

Integration & Operating Notes

Electrical
Supply must be 400 V 3~ with 3x16 A fusing and a fitted plug

The unit draws up to 12 A per the datasheet and is delivered with a cable but no plug for three-phase devices below 63 A, so the customer must source and fit a compatible plug. Voltage and frequency tolerances are ±5% and ±2%, but the combination of -5% voltage with +2% frequency is explicitly not allowed; verify the supply before commissioning.

Datasheet, Specifications; Operation Manual, Section on power connection
Electrical
Leakage current above 3.5 mA dictates RCD and earthing design

The datasheet flags leakage current >3.5 mA, which means standard 30 mA residual-current devices may nuisance-trip or be unsuitable; the protective earthing concept and any RCD selection must be engineered for this leakage level rather than assumed from a generic installation.

Datasheet, Specifications (safety note
Process Limits
Expansion vessel lock state must match open vs closed application

For externally closed applications the expansion vessel must remain unlocked — locking it causes overpressure damage to the application. For externally open (bath) applications it must be locked to prevent fluid loss. The isolation sleeve/kit state must be set accordingly before filling and venting.

Operation Manual, Section 2 Preparations for Operation
Mounting
ATEX zones are excluded; Ex px cabinet is the only route

The unit must not be installed or started within an ATEX zone; without an Ex px cabinet neither unit nor accessory is protected against explosion, and only the model/serial stated on the Ex px cabinet rating plate is suitable. Site selection must therefore exclude hazardous areas unless the separate ATEX annex applies.

Operation Manual, Section 1.3.3
Cooling & Media
Air-cooled condenser needs airflow and periodic fin cleaning

The refrigeration machine is air-cooled and CFC/HCFC-free; adequate airflow around the condenser is required and the liquefier fins must be cleaned periodically per the maintenance section. Poor airflow or fouled fins reduce cooling capacity, which matters most near the low-temperature end of the range.

Datasheet, Overview; Operation Manual, Section Service and Maintenance
Control & I/O
Set power-interruption response before unattended operation

Pilot ONE can be configured to 'Off/Standby' (default, manual restart) or 'ON/Temperature control active' (automatic restart after power returns, with a brief INFO allowing cancellation). Choose according to whether unattended restart is acceptable for the process; Process Safety mode additionally requires the compressor set to Permanently ON.

Operation Manual, Section 1.8.1 and Setup Mode
Maintenance
Refrigerant-circuit work is restricted to approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors; general maintenance personnel may handle hoses, strainer, drip tray and surface cleaning only. Heat transfer fluid must be inspected/replaced and the circuit rinsed per the maintenance schedule, and hoses/connections checked periodically for leaks and quality.

Operation Manual, Section 1.4.2 and Service and Maintenance

Selection & Integration Q&A

When should I choose the P815 Dynamisches rather than the 815w Dynamisches or another 8xx sibling?

Pick the P815 when the loop has high pressure drop (flow-through chemistry, semiconductor) and needs the high-discharge-pressure stainless pump; the 815w is the water-cooled variant at 60 dB(A) for sites without air-cooling headroom, and 705w/1005w/1015w differ in capacity.

What must the adjacent loop and power installation provide for this unit?

M30x1.5 male pump connections per DIN 3863 bore shape Y (60°) with matching ball-socket hoses and DIN 3870 sleeve nuts; a 400 V 3~ 50 Hz supply with 3x16 A fusing and a customer-fitted plug; and a loop pressure drop checked against the pump curve (67 l/min max, 3 bar max, 28 l/min at 2.5 bar).

What site and duty-cycle limits apply from the manual?

Install only outside ATEX zones, upright and level with feet/brakes set, ambient 5...40 °C, with a drip tray beneath; allow 2 h acclimation after temperature changes. Cooling capacity is rated at 20 °C room temperature and derates as ambient rises, so warm rooms reduce low-temperature duty.

What maintenance and accessories are needed beyond the base delivery?

The base unit ships with cable but no plug; an isolation sleeve/kit may be needed for externally open applications. Plan periodic condenser fin cleaning, strainer/dirt-trap cleaning, hose replacement, heat transfer fluid inspection/replacement and circuit rinsing; refrigerant work requires approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P815 Temperature Control System Datasheet
Datasheet ·111,7 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF