Peter Huber Kältemaschinenbau Unistat 1015w Dynamisches
Peter Huber Kältemaschinenbau Unistat 1015w Dynamisches — Temperature control units.
The Unistat 1015w Dynamisches is a water-cooled, free-standing temperature control unit with a 4 kW heater and a magnetically coupled stainless-steel pump, covering -120 to 100 °C. It is the largest of the 10x5 series, distinguished by its 28 l expansion volume and 639 kg mass. It serves externally closed or open applications requiring high heating and cooling power under industrial conditions.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat 1015w Dynamisches Current | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 28 l |
| Filling volume | 7 l (min) |
| Heating capacity | 4 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -120...100 °C |
| Weight | 639 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 560 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 3 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and positioning
- The Unistat 1015w is a water-cooled (w) free-standing unit in the 10x5 series, which also includes the 1005w, 705w, 815w and 825w. It is the largest sibling, with a 28 l expansion tank and 639 kg weight.
- The pump is magnetically coupled and made of stainless steel, with adjustable speed and an M30x1.5 male connection per DIN 3863.
- The unit uses a Pilot ONE controller with POKO ONE control output, a 0.01 K display resolution, and interfaces Ethernet, USB (host and device) and RS232.
- Safety class III / FL per DIN 12876 with adjustable overtemperature protection; suitable for both externally closed and externally open applications.
Behavior under load and media
- Temperature range is -120 to 100 °C, with a heating capacity of 4 kW. Cooling is via a water-cooled circuit using R-452A, R-23 and R-14 refrigerants.
- The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s; higher viscosity can overload the pump.
- Cooling water consumption is 560 l/h at 15 °C water and 0 °C flow, with a minimum differential pressure of 3 bar and maximum pressure of 6 bar.
- The unit is EMC class A group 1 per EN55011, intended for industrial environments.
Installation, connection and operation
- Requires a 400 V 3~ 50 Hz supply with a 3x32 A fuse and maximum current of 27.5 A. Three-phase devices under 63 A are delivered with a cable but without a plug.
- Leakage current exceeds 3.5 mA, affecting RCD selection. Voltage and frequency tolerances are ±5 % and ±2 %, but -5 % voltage with +2 % frequency is not allowed.
- Cooling water connections are G1/2 male. If cooling water temperature falls below the dew point, internal pipes must be insulated.
- The unit must not be installed in an ATEX zone. A drip tray for condensation water or heat transfer fluid must be installed below the unit.
- For externally closed applications, the expansion vessel must remain unlocked; for externally open applications, it must be locked.
Integration & Operating Notes
The unit draws up to 27.5 A and requires a 3x32 A fuse. Three-phase devices under 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site. Voltage and frequency tolerances are ±5 % and ±2 %, but a -5 % voltage combined with +2 % frequency is not allowed.
differential pressure with a flow of about 560 l/h at 15 °C, and the intake temperature must stay between 5 °C and 20 °C. If cooling water temperature drops below the dew point, internal pipes must be insulated to prevent condensation.
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s. Selecting a fluid with higher viscosity at operating temperature can overload the pump and reduce circulation.
The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet. A condensation water / heat transfer fluid drip tray must be installed below the unit; national legislation may prescribe it.
In Process Safety mode, when the set overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active.
Shut off the cooling water supply to the temperature control unit even for shorter downtimes such as overnight. Check periodically for leaks and the quality of hoses and hose connections and replace as required.
Selection & Integration Q&A
Choose the 1015w when you need the largest expansion volume (28 l) and highest heating capacity (4 kW) in the 10x5 series. The 1005w and 815w have smaller capacities; select them for lower heat loads or space constraints.
Cooling water: 3–6 bar differential pressure, 560 l/h at 15 °C, intake 5–20 °C, G1/2 male connections. Electrical: 400 V 3~ 50 Hz, 3x32 A fuse, max 27.5 A, cable without plug; leakage current >3.5 mA.
Do not install in an ATEX zone. Provide a drip tray. Ambient temperature 5–40 °C. Leveling feet must be activated and casters are not for transport. Allow 2 h acclimation after moving between cold and warm environments.
Periodically check hoses and connections for leaks and replace as needed. Shut off cooling water during downtimes. A suitable mains plug and a drip tray are not included and must be provided on site. Refrigerant work requires approved contractors.