Peter Huber Kältemaschinenbau Unistat P905w Dynamisches

Peter Huber Kältemaschinenbau Unistat P905w Dynamisches — Temperature control units.

The Unistat P905w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump, covering -90 to 250 °C. Its high-discharge-pressure pump targets high-pressure-drop circuits such as flow-through chemistry and semiconductor processes. It is the water-cooled P-variant of the Unistat 905 series, requiring 460 V 3~ 60 Hz and facility cooling water.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat P905w Dynamisches Current —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 6 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -90...250 °C
Dimensions, Weight & Mechanical Design
Weight 261 kg
Electrical, Power & Drive Specifications
Current consumption 23 A (max)
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and load behavior

Installation, connection and operation

Integration & Operating Notes

Electrical
Plan branch protection for >3.5 mA leakage current

The unit's leakage current exceeds 3.5 mA, so residual-current protection and UL 489 branch-circuit protection must be coordinated by the site electrician; a standard 30 mA RCD may not be suitable without verification.

Unistat P905w Technical Data Sheet, page 2
Electrical
Site must supply the three-phase plug

Three-phase devices below 63 A are delivered with cable but without a plug; the customer must provide the mating plug and complete the connection. Hard-wired connection must be carried out by an electrician.

Unistat P905w Technical Data Sheet, page 2; Unistat Operation Manual, page 37
Cooling & Media
Verify facility cooling water before installation

The water-cooled refrigeration circuit needs 580 l/h at 15 °C with at least 1 bar differential pressure and a maximum of 6 bar; cooling water inlet must stay between 5 and 20 °C. Out-of-range water temperature degrades capacity and creates condensation risk, so pipes must be insulated where condensation can occur.

Unistat P905w Technical Data Sheet, page 2
Process Limits
Check fluid viscosity at the lowest operating temperature

Maximum permissible kinematic viscosity is 50 mm²/s. Too-viscous fluid at low temperature overloads the pump and reduces heat transfer, so the heat transfer fluid must be checked against this limit at the coldest setpoint.

Unistat P905w Technical Data Sheet, page 2
Mounting
Transport and store upright only; drain before moving

Horizontal transport damages the compressor, and only an emptied unit may be moved. Casters are not suitable for transport and carry only 25% of total mass symmetrically; use an industrial truck. Leveling feet must be activated and parking brakes engaged before commissioning.

Unistat Operation Manual, page 22
Mounting
Allow 2 hours acclimation before first power-on

When moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults.

Unistat Operation Manual, page 29
Process Limits
Expansion vessel setting depends on open vs closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss.

Unistat Operation Manual, page 33
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles; this mode requires the compressor to be set to Permanently ON to keep cooling active.

Unistat Operation Manual, page 19
Control & I/O
Decide power-failure restart behavior before commissioning

After a power outage the response can be set via Pilot ONE: 'Off/Standby' (default, manual restart) or 'ON/Temperature control active' (automatic restart with a brief INFO allowing cancellation). Choose according to whether unattended restart is acceptable for the process.

Unistat Operation Manual, page 21
Maintenance
Shut off cooling water even for overnight downtime

The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit.

Unistat Operation Manual, page 29
Maintenance
Refrigerant work requires approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. Refrigerant charges are R-452A (1.4 kg, GWP 2141) and R-23 (0.75 kg, GWP 14800), combined CO2 equivalent about 14.1 t, which matters for leak reporting and service.

Unistat P905w Technical Data Sheet, page 1; Unistat Operation Manual, page 19

Selection & Integration Q&A

When should I choose the P905w Dynamisches instead of a sibling such as the 815w or 1005w Dynamisches?

Choose the P905w when the external circuit has high pressure drop (flow-through chemistry, semiconductor) and needs the high-discharge-pressure pump; the 815w or 1005w differ in size class, so select those when cooling capacity or chamber size, not pump head, is the deciding factor.

What facility services and mating parts must be ready before the P905w can be commissioned?

A 460 V 3~ 60 Hz supply with 3x25 A fusing and a site-provided three-phase plug (cable delivered without plug), plus cooling water at 580 l/h, 15 °C, 1–6 bar on G1/2 male connections. Pump connections are M30x1.5 AG male; an external Pt100 sensor can be connected.

What mounting and duty-cycle limits apply to the P905w?

Install upright at 5–40 °C ambient, IP20, not in an ATEX zone; activate leveling feet and parking brakes. Transport only upright and emptied. Allow 2 h acclimation after temperature changes. Shut off cooling water even overnight. Use with medical devices or direct foodstuff is not permissible.

What maintenance and consumables are required, and what is not included in base delivery?

Periodically check hoses and connections for leaks and quality and replace as needed; clean liquefier fins and the hat-type strainer; inspect heat transfer fluid and replace/rinsed circuit; refrigerant work by approved contractors. The three-phase plug, heat transfer fluid, and drip tray are not included and must be provided by the site.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P905w Technical Data Sheet
Datasheet ·75,7 КБ ·3 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF