Peter Huber Kältemaschinenbau Unistat 705w Dynamisches
Peter Huber Kältemaschinenbau Unistat 705w Dynamisches — Temperature control units.
Hydraulically sealed refrigerated heating circulator with water-cooled refrigeration, covering -75 to 250 °C via Pilot ONE controller. The 'w' water-cooled variant in the Unistat 705 series, it requires a cooling water supply (G1/2 male, 0.5–6 bar) and delivers 11 A max at 230 V 1~ 50 Hz. Used for precise temperature control of externally closed or open applications in process engineering.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat 705w Dynamisches Current | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Control output | POKO ONE |
| Pump connection | M24x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -75...250 °C |
| Weight | 90 kg |
| Current consumption | 11 A (max.) |
| Fuse rating | 16 A |
| Power supply | 230V 1~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The Unistat 705w Dynamisches is a water-cooled ('w') hydraulically sealed refrigerated heating circulator; sibling models 815w, 825w, 1005w and 1015w differ in cooling capacity and size, while this 705w is the base water-cooled unit in the series.
- Unlike the Unistat tango (w/wl), 405(w) and 410(w), the 705(w) does not have a plate heat exchanger, so the manual's warning against very dynamic, high temperature changes recurring at short intervals does not apply in the same way.
- Refrigerant circuit uses R-1270 (A3, H220, GWP 0, 0.15 kg) and a second stage R-170 (A3, H220, 0.099 kg); work on the coolant circuit must be by approved refrigeration contractors.
Behavior and media
- Temperature range -75 to 250 °C; display resolution 0.01 K; external Pt100 sensor connection.
- Heat transfer fluid must have max. kinematic viscosity 50 mm²/s; for FL-rated fluids a fire point ≥65 °C is required, and ethanol use is supervised only.
- Minimum filling capacity 1.5 l, expansion tank 2.8 l; low-level protection via pressure sensor stops circulation if air is present (inadequate venting or low fill).
- Cooling water intake 5–20 °C; cooling power derates as water temperature rises; insulate cooling water pipes if below condensation point.
Installation, control and operation
- Cooling water connection G1/2 male, min. differential pressure 0.5 bar, max. pressure 6 bar; circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
- Pump connection M24x1.5 male; pipework must use ball socket per DIN 3863 and sleeve nut per DIN 3870.
- Electrical: 230 V 1~ 50 Hz, max. 11 A, 16 A fuse; voltage tolerance ±5% and frequency ±2% (not -5% V with +2% Hz simultaneously); Class A, Group 1 per EN55011; IP20; ambient 5–40 °C.
- Controller Pilot ONE with POKO ONE control output; interfaces Ethernet, USB (Host and Device), RS232; pump speed adjustable.
- For externally closed applications the expansion vessel must remain unlocked; for externally open (bath) applications it must be locked to prevent fluid loss.
- Power failure behavior configurable: Off/Standby (default) or ON/Temperature control active; Process Safety mode releases maximum cooling at OT trigger and requires compressor set to Permanently ON.
Integration & Operating Notes
The water-cooled refrigeration requires a cooling water supply at G1/2 male with min. differential pressure 0.5 bar and max. pressure 6 bar, and intake temperature between 5 and 20 °C. Exceeding 6 bar or falling below 0.5 bar differential can damage the unit or reduce cooling; cooling power derates as water temperature rises above 20 °C.
The unit is rated 230 V 1~ 50 Hz, max. 11 A, with a 16 A fuse. Voltage tolerance ±5% and frequency ±2% are allowed, but a -5% voltage combined with +2% frequency is not permitted. Operation outside this envelope risks malfunction or damage; hard-wired connection must be by an electrician.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. This is a mandatory configuration step before commissioning.
Maximum permissible kinematic viscosity is 50 mm²/s. For FL-rated fluids, the fire point must be ≥65 °C; ethanol use is permitted only under supervision. Fluid selection must also respect the materials in the cooling water circuit (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM).
The responsible body must install a condensation water / heat transfer fluid drip tray below the unit; national legislation may prescribe it. Transport only upright and emptied; casters are not for transport and carry only 25% of total mass symmetrically. Leveling feet must be activated and parking brakes engaged before commissioning.
In Process Safety mode, when the set overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON; otherwise the safety function is not effective.
Shut off the cooling water supply even for shorter downtimes such as overnight to prevent water damage and unnecessary load. Periodically check hoses and connections for leaks and quality, and replace as needed. Cleaning of liquefier fins, hat-type strainer and drip tray is described in the maintenance section.
Selection & Integration Q&A
Choose the 705w when your cooling demand fits the base water-cooled unit and you have a cooling water supply (G1/2, 0.5–6 bar, 5–20 °C). The 815w/1005w offer higher cooling capacities; select them if your process heat load exceeds the 705w's capacity. All are water-cooled and share the -75…250 °C range.
Connect to G1/2 male with min. differential pressure 0.5 bar and max. pressure 6 bar; intake temperature 5–20 °C. Use materials compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Insulate pipes if water temperature falls below condensation point.
Ambient temperature 5–40 °C, IP20 (indoor). Unit must not be installed in ATEX zones. Allow ~2 h acclimation after moving between cold and warm environments. For externally closed applications keep expansion vessel unlocked; for open applications lock it.
Base delivery includes power cable with country-specific plug (specify at order). Not included: drip tray (must be installed by responsible body), cooling water hoses and fittings (ball socket per DIN 3863, sleeve nut per DIN 3870), heat transfer fluid, and optional gas detection sensor/processing unit. Periodic hose replacement and strainer cleaning are required.