Peter Huber Kältemaschinenbau Unistat 825w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 825w Dynamisches

Peter Huber Kältemaschinenbau Unistat 825w Dynamisches — Temperature control units.

The Unistat 825w Dynamisches is a water-cooled, free-standing temperature control unit with a variable-speed pressure pump and Pilot ONE controller, covering -85 to 250 °C. Within the Unistat 8xx series it is distinguished by its water cooling and dynamic performance, requiring a 400 V 3-phase supply and cooling water at 1–6 bar. It serves precise temperature control of externally closed or open applications in industrial processes.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat 825w Dynamisches Current —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Filling volume 3 l
Heating capacity 3 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type Pressure pump, variable speed
External sensor connection Pt100
Temperature range -85...250 °C
Dimensions, Weight & Mechanical Design
Weight 223 kg
Electrical, Power & Drive Specifications
Current consumption 15 A (max.)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 231 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal performance and media

Installation and operation

Safety and compliance

Integration & Operating Notes

Electrical
Supply must be 400 V 3-phase with 3x16 A fusing

The unit draws up to 15 A per phase and requires a 400 V 3~ 50 Hz supply protected by a 3x16 A fuse. Voltage and frequency tolerances are ±5% and ±2% but must not combine unfavorably (e.g., -5% voltage with +2% frequency is not allowed). A plug is not included and must be fitted by an electrician.

Datasheet, Technical Specifications; Operation Manual, Section 1.8
Cooling & Media
Cooling water must be 5–20 °C at 1–6 bar differential

The cooling water circuit requires a minimum differential pressure of 1 bar and must not exceed 6 bar. Inlet temperature must be between 5 °C and 20 °C; if pipes can drop below dew point they must be insulated. Consumption is 231 l/h at 15 °C water and 0 °C flow. Shut off cooling water even for short downtimes.

Datasheet, Technical Specifications; Operation Manual, Section 1.4
Mounting
Do not install in ATEX zone; ensure level mounting

The unit must not be installed or started within an ATEX zone; explosion protection requires an Ex px cabinet. It must be transported and stored upright only, and leveling feet or casters with parking brakes must be secured before commissioning. Allow about 2 hours acclimation when moving between cold and warm environments before power-on.

Operation Manual, Section 1.3.3 and Section 1.4
Process Limits
Expansion vessel must be unlocked for closed applications

For externally closed applications the expansion vessel must remain unlocked; locking it will damage the application due to overpressure. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. The heat transfer fluid must have a fire point of ≥65 °C if ethanol is used, and only supervised operation is permitted.

Operation Manual, Section 1.7 and Section 2.3
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. Power failure behavior can be configured to Off/Standby or automatic restart; choose according to whether unattended restart is acceptable.

Operation Manual, Section 1.7.2 and Section 1.8.1
Maintenance
Refrigerant work only by approved contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. The hat-type strainer/dirt trap requires cleaning with draining of the cooling circuit. A drip tray must be installed below the unit.

Operation Manual, Section 1.4.2 and Section 7

Selection & Integration Q&A

When should I choose the Unistat 825w Dynamisches over the 815w or 705w models?

Choose the 825w for its water cooling and higher cooling capacity at low temperatures; the 815w is air-cooled and the 705w lacks a plate heat exchanger. The 825w is suited for deep cooling to -85 °C with water cooling available, while the 705w may suffice for less demanding applications.

What are the requirements for the cooling water supply and electrical connection?

Cooling water must be 5–20 °C, with 1–6 bar differential pressure and G1/2 male connections; consumption is 231 l/h at 15 °C water and 0 °C flow. Electrical supply is 400 V 3~ 50 Hz, max. 15 A, fused at 3x16 A; a plug is not included and must be fitted by an electrician.

What are the ambient and duty-cycle limitations for installation?

Ambient temperature must be 5–40 °C; the unit is IP20 and must not be installed in an ATEX zone. It must be transported and stored upright, and leveling feet or casters with brakes must be secured. Allow 2 hours acclimation after temperature changes before power-on.

What maintenance is required and what accessories are not included?

Refrigerant circuit work must be by approved contractors; periodically check hoses and clean the strainer. A drip tray must be installed. A power plug is not included; heat transfer fluid and cooling water hoses are not included and must be selected per viscosity and temperature limits.