Peter Huber Kältemaschinenbau Unistat 815w Dynamisches
Peter Huber Kältemaschinenbau Unistat 815w Dynamisches — Temperature control units with Noise level 60 dB.
The Unistat 815w Dynamisches is a water-cooled, free-standing temperature control unit with a -85 to 250 °C range and 2 kW heating, positioned between the 705w and 825w siblings. Its variable-speed pressure pump and Pilot ONE controller with process-safety mode suit dynamic, exothermic reactor temperature control in closed circuits.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat 815w Dynamisches Current | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Noise level | 60 dB(A) |
| Control output | POKO ONE |
| Heating capacity | 2 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -85...250 °C |
| Weight | 222 kg |
| Current consumption | 13 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 130 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled (w) free-standing Unistat 8xx platform; 460 x 604 x 1465 mm, 222 kg, IP20, ambient 5–40 °C.
- Cascade refrigeration with R-452A (0.95 kg, GWP 2141) and R-23 (0.39 kg, GWP 14800); cooling water G1/2 male, 5–20 °C intake, 1–6 bar differential, 130 l/h at 15 °C.
- Sits between the 705w and 825w in the series; the 'Dynamisches' designation denotes the dynamic temperature-control configuration.
Behavior under load
- Cooling capacity is 1.5 kW down to -20 °C but only 0.2 kW at -80 °C, so deep-cold duty must be checked against the actual load.
- Pump delivers 40 l/min maximum but 5 l/min at 0.8 bar; external circuit pressure drop must be checked against the pump curve.
- Safety class III / FL with adjustable overtemperature protection per DIN 12876; Process Safety mode releases maximum cooling at the OT trigger instead of switching off.
Control, installation and operation
- Pilot ONE controller with POKO ONE output; internal/process control mode, 10-program/100-step programmer with linear and non-linear ramps; Ethernet, USB host/device, RS232; Pt100 external sensor input.
- 400 V 3~ 50 Hz supply, 13 A max, 3x16 A fuse; three-phase units below 63 A are delivered with cable but without plug.
- Variable-speed soft-start pump with integrated pressure control and optional external pressure sensor; pump connections M30x1.5 male.
- Must not be installed in an ATEX zone; transport upright only; allow ~2 h acclimation after cold/warm move; expansion vessel must remain unlocked for externally closed applications.
Integration & Operating Notes
The cooling water circuit requires an intake temperature between 5 and 20 °C, a minimum differential pressure of 1 bar and a maximum of 6 bar, with 130 l/h consumption at 15 °C. Pipes must be insulated if water can drop below the dew point, and the supply should be shut off even for short downtimes such as overnight.
Cooling capacity is rated at 1.5 kW down to -20 °C but only 0.2 kW at -80 °C. Low-temperature duty must be checked against the actual heat load, and the external circuit pressure drop must be checked against the pump curve (40 l/min maximum, 5 l/min at 0.8 bar) to avoid starving flow and degrading temperature control.
The unit must be transported and stored upright only; horizontal transport damages the compressor. Only an emptied unit may be moved. Allow about 2 hours acclimation when moving between cold and warm environments before switching on, to prevent condensation-induced electrical faults. Leveling feet must be activated and parking brakes engaged before commissioning.
In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. Power Failure Auto can resume temperature control automatically after a power outage if it was active, which must be assessed for unattended operation.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. The cooling water supply should be shut off even for short downtimes. The unit is safety class III / FL with adjustable overtemperature protection per DIN 12876.
Selection & Integration Q&A
Choose the 815w when the duty needs the -85 to 250 °C range with 2 kW heating and the dynamic configuration, and the load fits its 1.5 kW at -20 °C / 0.2 kW at -80 °C cooling curve. The 705w is smaller; the 825w offers more capacity if the load exceeds the 815w curve.
A 400 V 3~ 50 Hz supply with 3x16 A protection and a site-fitted plug (cable delivered without plug), cooling water at 5–20 °C with 1–6 bar differential and G1/2 male connections, and M30x1.5 male pump fittings for the external circuit. A Pt100 external sensor can be connected for process control.
Install dry, within 5–40 °C ambient, IP20, and never in an ATEX zone. Transport upright only and move only when emptied; allow ~2 h acclimation after a cold/warm move. Activate leveling feet and parking brakes before commissioning, and keep the expansion vessel unlocked for externally closed applications.
The plug for the three-phase cable is not included. Refrigerant-circuit work requires approved refrigeration contractors. Hoses and connections need periodic leak/quality checks and replacement as needed; the cooling water supply should be shut off even overnight. A condensation/fluid drip tray must be installed by the responsible body.