Peter Huber Kältemaschinenbau Unistat 905w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 905w Dynamisches

Peter Huber Kältemaschinenbau Unistat 905w Dynamisches — Temperature control units.

The Unistat 905w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -90 to 250 °C. It is the mid-size 9xx modification in the Unistat w series, distinguished by 6 kW heating and a 34 A three-phase supply. It serves closed or open external circuits requiring deep-cold to high-temperature control.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat 905w Dynamisches Current —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 6 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -90...250 °C
Dimensions, Weight & Mechanical Design
Weight 388 kg
Electrical, Power & Drive Specifications
Current consumption 34 A (max)
Power supply 208V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Cooling & Media
Cooling water must stay within 5–20 °C and ≥1 bar differential

The water-cooled refrigeration stage is designed for inlet water between 5 and 20 °C with at least 1 bar differential pressure and 580 l/h flow at 15 °C. Colder water risks condensation and requires insulated pipes; warmer water reduces capacity. Verify site supply before installation.

Unistat 905w Technical Data, page 2
Electrical
Provide UL 489 branch protection and a suitable plug on site

The unit draws up to 34 A at 208 V 3~ 60 Hz and is delivered with cable but no plug. External branch circuit protection according to UL 489 and 3x40 A fusing must be provided. Leakage current exceeds 3.5 mA, so standard 30 mA RCDs may nuisance-trip; select protective devices accordingly.

Unistat 905w Technical Data, page 2
Process Limits
Match external circuit volume and pressure drop to pump curve

Minimum filling capacity is 3.2 l and the expansion tank holds 5.3 l. The pump delivers 48 l/min at zero head but only 17 l/min at 0.8 bar, so external circuit pressure drop must be checked. Fluid viscosity must not exceed 50 mm²/s at the operating temperature.

Unistat 905w Technical Data, page 2
Mounting
Transport upright and emptied; acclimate 2 h before power-on

Horizontal transport damages the compressor, and a filled unit may overflow. Casters are not for transport and carry only 25% of mass; use an industrial truck. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to avoid condensation faults.

Unistat Operation Manual, page 22 and 29
Control & I/O
Set expansion vessel lock state per open or closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications it must be locked to prevent fluid loss. This is a mechanical interlock that must be verified before commissioning.

Unistat Operation Manual, page 33
Control & I/O
Choose power-failure restart behavior before unattended operation

Pilot ONE can be set to Off/Standby (default, manual restart) or automatic restart after power interruption. Process Safety mode releases maximum cooling at the OT trigger and requires the compressor set to Permanently ON. Select according to whether unattended restart is acceptable for the process.

Unistat Operation Manual, page 19–21
Maintenance
Shut off cooling water even for overnight downtime

The cooling water supply should be shut off during short downtimes to avoid water damage and unnecessary load on the circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. Refrigerant-circuit work requires approved refrigeration contractors.

Unistat Operation Manual, page 29

Selection & Integration Q&A

When should I choose the Unistat 905w over the 815w or 1005w siblings?

Pick the 905w when 6 kW heating and -90…250 °C with 34 A at 208 V 3~ match your load; the 815w is smaller, the 1005w larger. All are water-cooled 'w' variants; confirm cooling water 580 l/h at 15 °C and pump curve against your circuit.

What upstream electrical and cooling-water provisions does the 905w require?

208 V 3~ 60 Hz, max 34 A, 3x40 A fusing and UL 489 branch protection; cable supplied without plug. Cooling water G1/2 male, 5–20 °C inlet, ≥1 bar differential, 580 l/h at 15 °C. Leakage current >3.5 mA affects RCD choice.

What mounting and ambient limits apply from the manual?

Install upright and emptied, ambient 5–40 °C, IP20 (no wet/dusty areas), not in ATEX zones. Leveling feet or parking brakes engaged before start. Allow ~2 h acclimation after cold/warm move. Casters are not for transport.

What maintenance and accessories are not included in base delivery?

No plug, no drip tray, no cooling-water hoses or heat transfer fluid. Periodic hose/connection checks and fluid replacement are required; cooling water shutoff for overnight stops. Refrigerant work by approved contractors; gas detection sensor optional and installed by operator.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 905w Technical Data
Datasheet ·74,7 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF