Peter Huber Kältemaschinenbau Unistat P815 Dynamisches
Peter Huber Kältemaschinenbau Unistat P815 Dynamisches — Temperature control units with Noise level 64 dB.
The Unistat P815 Dynamisches is a free-standing, air-cooled refrigerated heating circulator with a magnetically coupled stainless-steel high-pressure pump, covering -85 to 250 °C on a 400 V 3~ 50 Hz supply. Its 'P' pump variant targets high-pressure-drop loops such as flow-through chemistry and semiconductor applications, distinguishing it from the standard-pump 815w sibling. It serves as the temperature-control source for externally closed or open circuits requiring deep-cooling capability.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat P815 Dynamisches Current | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Noise level | 64 dB(A) |
| Control output | POKO ONE |
| Heating capacity | 2 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -85...250 °C |
| Weight | 241 kg |
| Current consumption | 12 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing Unistat 8xx-class chassis, 241 kg, air-cooled CFC/HCFC-free refrigeration with R-452A (A1, H280) and R-23 (A1, H280) second stage; IP20, safety class III/FL per DIN 12876.
- The 'P' designation denotes the high-discharge-pressure pump (magnetically coupled, stainless steel, M30x1.5 male connections, DIN 3863 bore shape Y 60°), selected for high pressure-drop circuits; the 'w' sibling (Unistat 815w Dynamisches) is the water-cooled variant at 60 dB(A), and the 705w/1005w/1015w are other capacities in the same series.
- Heating capacity 2 kW; temperature range -85...250 °C; Pilot ONE controller with POKO ONE control output, 0.01 K display resolution, Pt100 external sensor input.
Behavior under load
- Cooling capacity is strongly temperature-dependent: with ethanol, 1.2 kW at -60 °C falls to 0.2 kW at -80 °C, so loads near the -85 °C limit must be small. Ratings are valid at 20 °C room temperature and derate as ambient rises.
- Pump delivers max 67 l/min and max 3 bar; flow falls to 28 l/min at 2.5 bar, so the loop pressure drop must be checked against the pump curve before committing to the 'P' variant.
- Viscosity limit 50 mm²/s constrains heat transfer fluid choice at low temperature.
Installation, control and operation
- 400 V 3~ 50 Hz, max current 12 A, 3x16 A fusing; delivered with cable but without plug (three-phase <63 A). Leakage current >3.5 mA requires considered protective earthing and RCD selection. Voltage/frequency tolerance ±5%/±2%, but -5% voltage with +2% frequency is not permitted.
- Interfaces: Ethernet, USB host and device, RS232. EMC Class A Group 1 (EN55011) — industrial environment only.
- Expansion vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications; an isolation sleeve may be needed for open applications.
Integration & Operating Notes
The unit draws up to 12 A per the datasheet and is delivered with a cable but no plug for three-phase devices below 63 A, so the customer must source and fit a compatible plug. Voltage and frequency tolerances are ±5% and ±2%, but the combination of -5% voltage with +2% frequency is explicitly not allowed; verify the supply before commissioning.
The datasheet flags leakage current >3.5 mA, which means standard 30 mA residual-current devices may nuisance-trip or be unsuitable; the protective earthing concept and any RCD selection must be engineered for this leakage level rather than assumed from a generic installation.
For externally closed applications the expansion vessel must remain unlocked — locking it causes overpressure damage to the application. For externally open (bath) applications it must be locked to prevent fluid loss. The isolation sleeve/kit state must be set accordingly before filling and venting.
The unit must not be installed or started within an ATEX zone; without an Ex px cabinet neither unit nor accessory is protected against explosion, and only the model/serial stated on the Ex px cabinet rating plate is suitable. Site selection must therefore exclude hazardous areas unless the separate ATEX annex applies.
The refrigeration machine is air-cooled and CFC/HCFC-free; adequate airflow around the condenser is required and the liquefier fins must be cleaned periodically per the maintenance section. Poor airflow or fouled fins reduce cooling capacity, which matters most near the low-temperature end of the range.
Pilot ONE can be configured to 'Off/Standby' (default, manual restart) or 'ON/Temperature control active' (automatic restart after power returns, with a brief INFO allowing cancellation). Choose according to whether unattended restart is acceptable for the process; Process Safety mode additionally requires the compressor set to Permanently ON.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors; general maintenance personnel may handle hoses, strainer, drip tray and surface cleaning only. Heat transfer fluid must be inspected/replaced and the circuit rinsed per the maintenance schedule, and hoses/connections checked periodically for leaks and quality.
Selection & Integration Q&A
Pick the P815 when the loop has high pressure drop (flow-through chemistry, semiconductor) and needs the high-discharge-pressure stainless pump; the 815w is the water-cooled variant at 60 dB(A) for sites without air-cooling headroom, and 705w/1005w/1015w differ in capacity.
M30x1.5 male pump connections per DIN 3863 bore shape Y (60°) with matching ball-socket hoses and DIN 3870 sleeve nuts; a 400 V 3~ 50 Hz supply with 3x16 A fusing and a customer-fitted plug; and a loop pressure drop checked against the pump curve (67 l/min max, 3 bar max, 28 l/min at 2.5 bar).
Install only outside ATEX zones, upright and level with feet/brakes set, ambient 5...40 °C, with a drip tray beneath; allow 2 h acclimation after temperature changes. Cooling capacity is rated at 20 °C room temperature and derates as ambient rises, so warm rooms reduce low-temperature duty.
The base unit ships with cable but no plug; an isolation sleeve/kit may be needed for externally open applications. Plan periodic condenser fin cleaning, strainer/dirt-trap cleaning, hose replacement, heat transfer fluid inspection/replacement and circuit rinsing; refrigerant work requires approved contractors.