Peter Huber Kältemaschinenbau Unistat 705w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 705w Dynamisches

Peter Huber Kältemaschinenbau Unistat 705w Dynamisches — Temperature control units.

Hydraulically sealed refrigerated heating circulator with water-cooled refrigeration, covering -75 to 250 °C via Pilot ONE controller. The 'w' water-cooled variant in the Unistat 705 series, it requires a cooling water supply (G1/2 male, 0.5–6 bar) and delivers 11 A max at 230 V 1~ 50 Hz. Used for precise temperature control of externally closed or open applications in process engineering.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat 705w Dynamisches Current —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825 Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Pump connection M24x1, 5 AG male
Pump speed adjustable Yes
Pump type Pressure pump, variable speed
External sensor connection Pt100
Temperature range -75...250 °C
Dimensions, Weight & Mechanical Design
Weight 90 kg
Electrical, Power & Drive Specifications
Current consumption 11 A (max.)
Fuse rating 16 A
Power supply 230V 1~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior and media

Installation, control and operation

Integration & Operating Notes

Cooling & Media
Cooling water supply must meet pressure and temperature envelope

The water-cooled refrigeration requires a cooling water supply at G1/2 male with min. differential pressure 0.5 bar and max. pressure 6 bar, and intake temperature between 5 and 20 °C. Exceeding 6 bar or falling below 0.5 bar differential can damage the unit or reduce cooling; cooling power derates as water temperature rises above 20 °C.

Unistat 705w Technical Data, page 2
Electrical
Supply must stay within voltage/frequency tolerance window

The unit is rated 230 V 1~ 50 Hz, max. 11 A, with a 16 A fuse. Voltage tolerance ±5% and frequency ±2% are allowed, but a -5% voltage combined with +2% frequency is not permitted. Operation outside this envelope risks malfunction or damage; hard-wired connection must be by an electrician.

Unistat 705w Technical Data, page 2; Unistat Operation Manual, page 37
Process Limits
Expansion vessel lock state depends on application type

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications the expansion vessel must be locked to prevent fluid loss. This is a mandatory configuration step before commissioning.

Unistat Operation Manual, page 33
Process Limits
Heat transfer fluid viscosity and fire point limits

Maximum permissible kinematic viscosity is 50 mm²/s. For FL-rated fluids, the fire point must be ≥65 °C; ethanol use is permitted only under supervision. Fluid selection must also respect the materials in the cooling water circuit (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM).

Unistat 705w Technical Data, page 2; Unistat Operation Manual, Section 1.7
Mounting
Install drip tray and ensure upright transport

The responsible body must install a condensation water / heat transfer fluid drip tray below the unit; national legislation may prescribe it. Transport only upright and emptied; casters are not for transport and carry only 25% of total mass symmetrically. Leveling feet must be activated and parking brakes engaged before commissioning.

Unistat Operation Manual, Section 1.4 / page 17; page 22
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the set overtemperature trigger value is reached, the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON; otherwise the safety function is not effective.

Unistat Operation Manual, Section 1.7.2 / page 20; page 19
Maintenance
Shut off cooling water during short downtimes

Shut off the cooling water supply even for shorter downtimes such as overnight to prevent water damage and unnecessary load. Periodically check hoses and connections for leaks and quality, and replace as needed. Cleaning of liquefier fins, hat-type strainer and drip tray is described in the maintenance section.

Unistat Operation Manual, page 29; Unistat® Operation Manual — Maintenance section

Selection & Integration Q&A

When should I choose the Unistat 705w Dynamisches over the 815w or 1005w siblings?

Choose the 705w when your cooling demand fits the base water-cooled unit and you have a cooling water supply (G1/2, 0.5–6 bar, 5–20 °C). The 815w/1005w offer higher cooling capacities; select them if your process heat load exceeds the 705w's capacity. All are water-cooled and share the -75…250 °C range.

What are the requirements for the cooling water circuit and adjacent piping?

Connect to G1/2 male with min. differential pressure 0.5 bar and max. pressure 6 bar; intake temperature 5–20 °C. Use materials compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Insulate pipes if water temperature falls below condensation point.

What are the ambient and duty-cycle limits for installation?

Ambient temperature 5–40 °C, IP20 (indoor). Unit must not be installed in ATEX zones. Allow ~2 h acclimation after moving between cold and warm environments. For externally closed applications keep expansion vessel unlocked; for open applications lock it.

What maintenance items and accessories are not included in base delivery?

Base delivery includes power cable with country-specific plug (specify at order). Not included: drip tray (must be installed by responsible body), cooling water hoses and fittings (ball socket per DIN 3863, sleeve nut per DIN 3870), heat transfer fluid, and optional gas detection sensor/processing unit. Periodic hose replacement and strainer cleaning are required.