No photo available yet

Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat P825 Dynamisches

Peter Huber Kältemaschinenbau Unistat P825 Dynamisches — Temperature control units.

The Unistat P825 Dynamisches is a free-standing refrigerated heating circulator with an air-cooled two-stage cascade refrigeration machine, covering -85 to 250 °C with a 3 kW heater and a magnetically coupled stainless-steel pump delivering up to 67 l/min at 3 bar. Its high-discharge-pressure 'P' pump distinguishes it from standard Unistat 8xx siblings, targeting externally closed loops with high pressure drop such as flow-through chemistry and semiconductor processes. It serves as a central temperature source in pilot-plant and process-scale systems requiring precise Pt100-based control via the Pilot ONE controller.

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CDisplay resolutionPower supplyPump connectionPump type
Unistat P825 Dynamisches Current —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 1005w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 1015w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 705w Dynamisches —5–400, 01230V 1~ 50HzM24x1, 5 AG malePressure pump, variable
Unistat 815w Dynamisches 605–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malePressure pump, variable
Unistat 905w Dynamisches —5–400, 01 K208V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat 912w Dynamisches 655–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat 920w Dynamisches 715–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat 930w Dynamisches —5–400, 01460V 3~ 60HzM38x1, 5 AG malemagnetically coupled, st
Unistat 950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st
Unistat P815 Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P815w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P825w Dynamisches —5–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905 Dynamisches —5–350, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P905w Dynamisches —5–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P912w Dynamisches 655–400, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st
Unistat P915w Dynamisches 645–400, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st
Unistat P920w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P925w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P930w Dynamisches —5–400, 01400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st
Unistat P950w Dynamisches —5–400, 01400V 3~ 50Hzflange DN40 PN10/40 EN10magnetically coupled, st

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Heating capacity 3 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -85...250 °C
Dimensions, Weight & Mechanical Design
Weight 238 kg
Electrical, Power & Drive Specifications
Current consumption 16 A (max.)
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Mounting
Expansion vessel lock state depends on open vs closed application

For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For externally open (bath) applications it must be locked to prevent fluid loss. Confirm the application type before commissioning and set the isolating sleeve/valve accordingly.

Unistat Operation Manual, Preparations for Operation
Control & I/O
Set power-failure restart behavior before unattended operation

Pilot ONE can be configured to remain in Off/Standby after power restoration (default, manual start required) or to resume temperature control automatically. Choose the mode according to whether an unattended restart is safe for the process; the auto-resume option shows an INFO for a few seconds allowing cancellation.

Unistat Operation Manual, Section 1.8.1 Power Interruption
Maintenance
Clean condenser fins and inspect hoses on a scheduled basis

The manual requires periodic cleaning of the liquefier fins, checking hoses and connections for leaks and quality with replacement as needed, cleaning the hat-type strainer, and inspecting/replacing the heat transfer fluid. Refrigerant-circuit work (R-452A, R-23) must be performed by approved refrigeration contractors only.

Unistat Operation Manual, Service and Maintenance

Selection & Integration Q&A

When should I choose the Unistat P825 Dynamisches over a sibling such as the Unistat 815w Dynamisches?

Choose the P825 when the external loop has high pressure drop (flow-through reactors, semiconductor tools) and needs the high-discharge-pressure pump (67 l/min, 3 bar). The 815w is water-cooled and standard-pressure; pick it when cooling water is available and loop resistance is low.

What mating fittings and hose hardware does the P825 pump connection require?

Pump connections are M30x1.5 male with bore shape Y (60°) per DIN 3863. Hoses must use a ball socket per DIN 3863 and sleeve nut per DIN 3870; mismatched fittings risk leaks at 3 bar and 250 °C.

What maintenance and accessories are not included in the base delivery?

Base delivery includes cable but no plug; the drip tray, mating plug/hard-wiring, heat transfer fluid, hoses with DIN 3863/3870 fittings, and optional gas detection sensor/processing unit are not included. Scheduled tasks: condenser fin cleaning, hose inspection, strainer cleaning, fluid replacement.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P825 Temperature Control System Datasheet
Datasheet ·112,5 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF