Peter Huber Kältemaschinenbau Unistat 930w Dynamisches
Peter Huber Kältemaschinenbau Unistat 930w Dynamisches — Temperature control units.
The Unistat 930w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -90 to 200 °C. Within the Unistat w series it is the high-capacity 9xx frame, distinguished by 32 kW heating and 66 A three-phase draw. It serves closed-loop process temperature control where high heating power and deep cooling are required.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat 930w Dynamisches Current | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Control output | POKO ONE |
| Expansion volume | 29 l |
| Filling volume | 12 l |
| Heating capacity | 32 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -90...200 °C |
| Weight | 984 kg |
| Current consumption | 66 A (max.) |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 1300 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Free-standing 9xx frame (984 kg) with water-cooled refrigeration; the 'w' suffix denotes water cooling, unlike air-cooled siblings.
- Sibling models 705w, 815w, 1005w and 1015w share the 'Dynamisches' dynamic temperature control concept but differ in frame size and capacity; the 930w sits in the 9xx class with 32 kW heating capacity.
- Pump is magnetically coupled, stainless steel, with adjustable speed; pump connection M38x1.5 male, bore shape Y (60°) to DIN 3863.
- Controller: Pilot ONE with POKO ONE control output; display resolution 0.01 K; interfaces Ethernet, USB (host and device) and RS232; external Pt100 sensor connection.
Working range and media behavior
- Temperature range -90 to 200 °C; safety class III / FL with adjustable overtemperature protection to DIN 12876.
- Heat transfer fluid must not exceed 50 mm²/s kinematic viscosity over the operating range; FL-rated fluids require a fire point ≥ 65 °C, and ethanol is permitted only under supervision.
- Minimum filling capacity 12 l with a 29 l expansion tank; the expansion vessel must remain unlocked for externally closed applications and locked for externally open (bath) applications.
- Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only suitable cooling water may be used.
Installation, connection and control
- Electrical supply 460 V 3~ 60 Hz, max current 66 A, fusing 3x80 A; external branch circuit protection to UL 489 required. Because current exceeds 63 A, the unit is delivered without power cable and without plug.
- Cooling water connection G3/4 male; minimum differential pressure 1 bar, maximum pressure 6 bar; intake temperature 5–20 °C; consumption 1300 l/h at 15 °C with 0 °C flow.
- Rated output data are valid at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure; higher water temperature reduces cooling power.
- Ambient operating range 5–40 °C; protection rating IP20; leakage current exceeds 3.5 mA, which affects RCD selection.
- Pilot ONE power-interruption behavior is selectable: Off/Standby (default) or ON/Temperature control active with a brief INFO window to cancel automatic restart.
Integration & Operating Notes
With a maximum current of 66 A the unit ships without cable and without plug; the customer must procure and terminate the three-phase feeder and provide external branch circuit protection according to UL 489. Panel sizing must account for 3x80 A fusing.
The datasheet states leakage current exceeds 3.5 mA, so a standard 30 mA residual current device may be unsuitable; the earthing arrangement and protective device selection must be engineered accordingly before commissioning.
Rated cooling output is quoted at 15 °C cooling water inlet and 1 bar differential pressure; intake temperature must remain between 5 and 20 °C and pressure must not exceed 6 bar. If the water can fall below the condensation point, the cooling water pipes must be insulated.
Fluid selection must keep kinematic viscosity at or below 50 mm²/s across the whole operating range, otherwise pump delivery and cooling performance suffer. For FL-rated fluids the fire point must be at least 65 °C, and ethanol use is permitted only under supervision.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications it must be locked to prevent fluid loss. Verify the isolating sleeve or shut-off valve position before filling.
Without an Ex px cabinet the unit is not protected against explosion; installation and start-up within an ATEX zone are prohibited. Site selection must exclude hazardous areas unless the separate ATEX annex and Ex px cabinet rating plate apply.
The manual directs shutting off the cooling water supply even for shorter downtimes such as overnight, and periodic checks of hoses and hose connections for leaks and quality with replacement as required. Refrigerant-circuit work must be performed by approved refrigeration/air-conditioning contractors.
In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles; this mode requires the compressor to be set to Permanently ON. Power-interruption behavior is selectable between Off/Standby and automatic restart.
Selection & Integration Q&A
Pick the 930w when 32 kW heating and the 9xx frame capacity are needed with water cooling; the 815w is a smaller frame, while the 1005w/1015w are larger. All share the 'Dynamisches' dynamic control concept, so selection turns on heating power and frame size rather than control architecture.
A 460 V 3~ 60 Hz feeder sized for 66 A with 3x80 A fusing and UL 489 branch protection, plus a customer-supplied power cable and plug (not delivered above 63 A). Cooling water at G3/4 male with 1–6 bar differential pressure, 5–20 °C intake and 1300 l/h at 15 °C.
Ambient 5–40 °C, IP20, no ATEX zone, upright transport only, leveling feet and parking brakes set before start-up, and about 2 hours acclimation after a cold-to-warm move. A drip tray for condensation water or heat transfer fluid must be installed below the unit.
The power cable and plug are excluded above 63 A, as are heat transfer fluid, cooling-water hoses and DIN 3863/3870 fittings, and the optional gas detection sensor and processing unit. Periodic tasks include hose and connection inspection, liquefier fin cleaning, strainer cleaning and fluid replacement.