Peter Huber Kältemaschinenbau Unistat P815w Dynamisches
Peter Huber Kältemaschinenbau Unistat P815w Dynamisches — Temperature control units.
The Unistat P815w Dynamisches is a water-cooled, magnetically coupled stainless-steel pump temperature control unit spanning -85 to 250 °C with 2 kW heating, driven by the Pilot ONE controller. Its 'P' designation targets applications with high pressure drop, distinguishing it from the 815w Dynamisches sibling. It serves precise dynamic temperature control of externally closed circuits in process and laboratory systems.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat P815w Dynamisches Current | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 2 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -85...250 °C |
| Weight | 231 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 130 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled ('w') free-standing Unistat of the 8xx class; magnetically coupled stainless-steel circulation pump with M30x1.5 male pump connection and G1/2 male cooling-water connection.
- The 'P' suffix marks the modification for applications with high pressure drop; the plain 815w Dynamisches is the neighbor when external circuit resistance is low. The 705w, 1005w and 1015w Dynamisches differ in class/size, not in this pressure-oriented configuration.
- Heating capacity 2 kW; temperature range -85...250 °C; display resolution 0.01 K; external Pt100 sensor input; safety class III/FL.
Behavior under load
- Cooling capacity falls steeply at low temperature: 1.5 kW at -20 °C versus 0.2 kW at -80 °C, so permissible heat load at the cold end must be sized against the 0.2 kW figure.
- Pump delivers up to 67 l/min and 3 bar maximum; 28 l/min remains available at 2.5 bar, so the external circuit pressure drop must be checked against this curve. Pump speed is adjustable.
- Cooling water at 15 °C with 0 °C flow consumes 130 l/h; higher intake water temperatures reduce cooling power and raise water flow.
Control, installation and operation
- Pilot ONE controller with POKO ONE control output; interfaces Ethernet, USB (host and device) and RS232; touchscreen with LED temperature, flow-chart and status indicators.
- Setup mode covers overtemperature protection (OT limit heating, OT expansion vessel, Process Safety), Delta T limiter, Pt100 monitoring and auto/expert control structure; filling, venting, degassing and draining differ for externally closed versus externally open applications.
- Power: 400 V 3~ 50 Hz with 3x16 A fuse; three-phase units below 63 A ship with cable but without plug. Leakage current exceeds 3.5 mA, requiring suitable RCD and protective earthing. Voltage tolerance ±5 % only with ±2 % frequency tolerance; -5 % voltage with +2 % frequency is not allowed.
Integration & Operating Notes
The ±5 % voltage and ±2 % frequency tolerances may not be combined freely: -5 % voltage with +2 % frequency is explicitly not allowed. Verify the site supply against this matrix before connection, because a compliant-looking voltage alone does not guarantee permissible operation.
Three-phase units below 63 A are delivered with a cable but without a plug, so the customer must supply and install it. Because leakage current exceeds 3.5 mA, a suitable residual current device and protective earthing must be provided; the electrician must confirm the rating-plate voltage and frequency match.
The cooling water circuit requires a minimum differential pressure of 1 bar and tolerates a maximum of 6 bar, with intake temperature between 5 °C and 20 °C. Higher intake temperatures reduce cooling power and increase water flow, so facility water conditions directly cap achievable performance.
Without an Ex px cabinet the unit is not protected against explosion, so site selection must exclude ATEX zones. Ex px operation requires the separate ATEX annex and only the model/serial number stated on the Ex px cabinet rating plate is suitable.
Horizontal transport damages the compressor, and only an emptied unit may be moved to avoid thermal fluid overflow. Casters are not for transport and carry only 25 % of total mass symmetrically; use an industrial truck. After moving between cold and warm environments, allow about 2 hours acclimation before switching on.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage. For externally open (bath) applications it must be locked to prevent fluid loss. The wrong state in either direction produces damage or fluid loss, so the application type must be fixed before commissioning.
In Process Safety mode the controller releases maximum cooling capacity at the OT trigger instead of switching off all poles, which requires the compressor to be set to Permanently ON. Configure this deliberately when protecting exothermic processes such as glass reactors.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors; general maintenance personnel cannot service the refrigerant circuit. The refrigerant charge carries a reportable CO2 equivalent (about 7.8 t total for R-452A and R-23), which must be managed under F-gas rules.
Selection & Integration Q&A
Pick the P815w when the external circuit has high pressure drop: the 'P' modification is intended for applications with high pressure drop, and the pump still delivers 28 l/min at 2.5 bar. The plain 815w suits low-resistance circuits; 705w/1005w differ in class and capacity.
Cooling water at 5–20 °C with 1–6 bar differential pressure through a G1/2 male connection (130 l/h at 15 °C water, 0 °C flow), and a 400 V 3~ 50 Hz supply fused at 3x16 A. The unit ships with cable but no plug; leakage current >3.5 mA requires a suitable RCD and earthing.
Ambient must stay 5–40 °C, ATEX zones are excluded, and the unit must stand upright on activated leveling feet with casters braked. Allow 2 h acclimation after cold/warm moves. Shut off cooling water even overnight; the expansion vessel lock state must match open vs closed application.
The customer supplies the three-phase plug and a drip tray below the unit. Periodic checks cover hoses and connections, liquefier fins, the hat-type strainer/dirt trap, emergency-stop function test and heat transfer fluid inspection/replacement. Refrigerant-circuit work requires approved contractors.