Peter Huber Kältemaschinenbau Unistat P825w Dynamisches
Peter Huber Kältemaschinenbau Unistat P825w Dynamisches — Temperature control units.
The Unistat P825w Dynamisches is a water-cooled, free-standing temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -85 to 250 °C. Its 'P' designation targets applications with high pressure drop, delivering up to 3 bar and 67 l/min. It serves closed-loop process temperature control in industrial and laboratory systems.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Ambient temperature limit, °C | Display resolution | Power supply | Pump connection | Pump type |
|---|---|---|---|---|---|---|
| Unistat P825w Dynamisches Current | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 1005w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 1015w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 705w Dynamisches | — | 5–40 | 0, 01 | 230V 1~ 50Hz | M24x1, 5 AG male | Pressure pump, variable |
| Unistat 815w Dynamisches | 60 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 825w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | Pressure pump, variable |
| Unistat 905w Dynamisches | — | 5–40 | 0, 01 K | 208V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 912w Dynamisches | 65 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat 920w Dynamisches | 71 | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 930w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat 950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
| Unistat P815 Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P815w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P825 Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905 Dynamisches | — | 5–35 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P905w Dynamisches | — | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P912w Dynamisches | 65 | 5–40 | 0, 01 | 460V 3~ 60Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P915w Dynamisches | 64 | 5–40 | 0, 01 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st |
| Unistat P920w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P925w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P930w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st |
| Unistat P950w Dynamisches | — | 5–40 | 0, 01 | 400V 3~ 50Hz | flange DN40 PN10/40 EN10 | magnetically coupled, st |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 3 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -85...250 °C |
| Weight | 237 kg |
| Current consumption | 15 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled ('w') free-standing Unistat with a two-stage cascade refrigerant circuit (R-452A and R-23) and a magnetically coupled stainless-steel circulation pump.
- The 'P' suffix marks the high-pressure-drop variant; within the Unistat w series it sits alongside the 705w, 815w, 1005w and 1015w, which differ in cooling capacity and pump characteristics rather than in control architecture.
- Pump connection is M30x1.5 male; cooling water connection is G1/2 male. Safety class III / FL with adjustable overtemperature protection to DIN 12876.
Behavior under load and media
- Cooling capacity falls steeply at low temperature: 2.2 kW at 0 °C down to 0.3 kW at -80 °C, so process load must be sized against the target temperature, not the nominal range.
- The pump delivers 67 l/min maximum and 3 bar maximum pressure, with 28 l/min still available at 2.5 bar, allowing it to serve loops with significant pressure drop.
- Cooling water consumption is 180 l/h at 15 °C water and 20 °C flow; higher water temperature reduces cooling power and raises flow demand.
- Heat transfer fluid must respect the FL rating: fire point ≥ 65 °C, and ethanol only under supervised operation.
Installation, control and operation
- Powered from 400 V 3~ 50 Hz with a 3x16 A fuse; three-phase units below 63 A are delivered with cable but without a plug. Voltage tolerance is ±5% only with ±2% frequency tolerance, and the combination -5% voltage with +2% frequency is not allowed.
- Cooling water must be 5–20 °C inlet, ≥1 bar differential pressure, ≤6 bar; pipes below dew point must be insulated. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, restricting water quality.
- Pilot ONE controller with POKO ONE control output, 0.01 K display resolution, Ethernet, USB host/device and RS232 interfaces, and Pt100 external sensor input.
- IP20, ambient 5–40 °C; must not be installed in an ATEX zone. Unit must be transported and stored upright and only moved when emptied.
Integration & Operating Notes
Voltage may deviate ±5% only while frequency deviates ±2%, and the combination of -5% voltage with +2% frequency is explicitly not allowed. Site power quality must be verified before connection, since out-of-tolerance supply can damage the compressor and electronics.
Units below 63 A are delivered with a cable but no plug, so a suitable plug or hard-wired connection must be provided on site. Hard-wired connection must be carried out by an electrician, and the supply must match the rating plate.
Inlet water must be between 5 °C and 20 °C with at least 1 bar differential pressure and no more than 6 bar. Higher water temperature reduces cooling power and increases flow demand; pipes below dew point must be insulated to prevent condensation.
The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used. Aggressive or untreated water can corrode or foul these materials and degrade heat transfer.
Cooling power is 2.2 kW at 0 °C but only 0.3 kW at -80 °C. Process loads that are acceptable at mid-range temperatures may exceed capacity near the lower limit, so load sizing must be done at the actual target temperature.
For FL-rated (flammable) fluid the fire point must be at least 65 °C; ethanol is permitted only under supervised operation. Fluid selection and supervision must respect the DIN 12876 FL classification.
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage to the application. For externally open (bath) applications it must be locked to prevent fluid loss. The setting must match the application type before commissioning.
The unit must be transported and stored upright only; horizontal transport damages the compressor. Only an emptied unit may be moved, since a filled unit risks thermal fluid overflow. Casters are not for transport and carry only 25% of total mass symmetrically.
Allow about 2 hours acclimation when moving between cold and warm environments before power-on to avoid condensation-induced electrical faults. Leveling feet must be activated and parking brakes engaged before commissioning.
In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles. This mode requires the compressor to be set to Permanently ON, so the controller configuration must be checked before relying on it for exothermic protection.
Pilot ONE can be set to Off/Standby (default, manual restart) or ON/Temperature control active (automatic restart after power interruption, with a brief INFO allowing cancellation). The choice must match whether unattended restart is acceptable for the process.
The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit contains R-452A (GWP 2141, 0.95 kg) and R-23 (GWP 14800, 0.39 kg), about 7.8 t CO2 equivalent, which also triggers leak-monitoring and shipping obligations.
Selection & Integration Q&A
Pick the P825w when the application loop has high pressure drop and needs up to 3 bar / 67 l/min, or when -85 °C reach with 0.3 kW at -80 °C is sufficient. The 815w and 1005w differ in cooling capacity and pump characteristics; select them when their capacity curve better matches the load at the target temperature.
A 400 V 3~ 50 Hz supply with 3x16 A fuse, within ±5% voltage and ±2% frequency (the -5% voltage/+2% frequency combination is not allowed). The unit arrives with cable but no plug. Cooling water: 5–20 °C inlet, ≥1 bar differential, ≤6 bar, G1/2 male connection, 180 l/h at 15 °C water.
IP20, ambient 5–40 °C, not in an ATEX zone. Transport and store upright only and move only when emptied; casters are not for transport. Allow about 2 h acclimation before power-on, activate leveling feet and parking brakes, and install a drip tray below the unit.
Not included: a suitable three-phase plug, cooling-water hoses and fittings (G1/2 male), pump-loop fittings (M30x1.5 male), heat transfer fluid, and a drip tray. Periodic tasks: check hoses and connections for leaks and replace as needed, clean liquefier fins and the hat-type strainer, inspect/replace heat transfer fluid, and shut off cooling water even overnight.